Pruitt Irrigation Occlusion CatheterRecalls & reported problems across all models
Lemaitre Vascular, Inc. · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names at least one Pruitt Irrigation Occlusion Catheter model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 2 models of Pruitt Irrigation Occlusion Catheter by Lemaitre Vascular, Inc. in the FDA's device registry (GUDID). Together they are named in 4 adverse event reports in the FDA's MAUDE database, received between Oct 29, 2021 and Feb 27, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Failure to Deflate” (2 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Pruitt Irrigation Occlusion Catheter recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Apr 11, 2025Class IIOpen, Classified
During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
Z-1751-2025 · Root cause: Packaging process control · names 2 of these models · Recall details · FDA record ↗
Most reported Pruitt Irrigation Occlusion Catheter problems
Device problems
- Failure to Deflate2
- Material Rupture2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Pruitt Irrigation Occlusion Catheter models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Pruitt Irrigation Occlusion Catheter 2102-09Occluder, Catheter Tip1 recall4 reports
- Pruitt Irrigation Occlusion Catheter e2102-09Occluder, Catheter Tip1 recall0 reports
Recent Pruitt Irrigation Occlusion Catheter adverse event reports
The 4 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PRUITT IRRIGATION OCCLUSION CATHETER BALLOON RUPTURED WHILE ATTEMPTING TO USE IT TO OCCLUDE VESSEL. IT DID NOT HOLD FLUID TO OCCLUDE VESSEL, EVEN ON FIRST PASS. IT WAS UNSURE IF THE PRODUCT WAS TESTED PRIOR TO USE. NO INJURY WAS REPORTED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18790728 ↗MalfunctionJul 20, 2023Failure to Deflate; Failure to Deflate
IT WAS REPORTED THE PRUITT IRRIGATION OCCLUSION CATHETER BALLOON DIDN'T DEFLATE. NO INJURY REPORTED TO PATIENT AS THE DEVICE WAS NOT DIRECTLY USED ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17362430 ↗MalfunctionApr 14, 2023Failure to Deflate; Failure to Deflate
IT WAS REPORTED THAT THE PRUITT IRRIGATION OCCLUSION CATHETER BALLOON FAILED TO DEFLATE DURING THE PRE-USE CHECK. THE DEVICE WAS NOT USED FOR THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16742964 ↗MalfunctionOct 29, 2021Material Rupture; Material Rupture
THE CATHETER WAS USED DURING THE REPLACEMENT OF THE ARTIFICIAL BLOOD VESSEL IN THE THORACOABDOMINAL REGION. THE BALLOON OF THE CATHETER RUPTURED DURING THE OPERATION. THE PHYSICIAN REPLACED THAT CATHETER WITH ANOTHER PRUITT IRRIGATION OCCLUSION CATHETER AND CONTINUED THE OPERATION. THERE WAS NO IMPACT ON THE PATIENT BECAUSE OF THE CATHETER MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12719744 ↗
Other Lemaitre Vascular, Inc. products
- LeMaitre Embolectomy Catheter47 models · 438 reports · 2 recalls
- Pruitt F3 Carotid Shunt10 models · 305 reports
- Artegraft Vascular Graft15 models · 25 reports · 3 recalls
- Trivex System3 models · 69 reports
- Syntel Silicone OTW Embolectomy Catheter8 models · 66 reports
- Syntel Silicone Thrombectomy Catheter5 models · 6 reports · 1 recall
- Pruitt Occlusion Catheter6 models · 5 reports · 1 recall
- LifeSpan ePTFE Vascular Graft10 models · 27 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.