Pruitt Irrigation Occlusion Catheter recallZ-1751-2025 · LeMaitre Vascular, Inc. · initiated Apr 11, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from LeMaitre Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
FDA root cause category: Packaging process control
What the company is doing
LeMaitre notified OUS consignees beginning on 04/11/2025, and US consignees on 04/23/2025 via email and letter. Consignees were instructed to not use the devices, immediately quarantine any on hand, and complete and return the response form. Once received a LeMaitre Vascular representative will contact customers to provide return instructions. If the products were transferred or further distributed, customers were instructed to notify those customers and provide a copy of the recall letter.
Recall details
- Recall number
- Z-1751-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 11, 2025
- Posted by the FDA
- May 12, 2025
- Recalling firm
- LeMaitre Vascular, Inc.
- Quantity
- 65,977 units
- Product code
- DYG
- FDA submissions
- K811276
- Distribution
- Nationwide distribution. International distribution to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, BULGARIA, CAMBODIA, CANADA, CHILE, COLOMBIA, COSTA RICA, CYPRUS, CZECHIA, DENMARK, DOMINICAN REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONGOLIA, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORTH MACEDONIA, NORWAY, OMAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKIYE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.
Product description
Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56
Code information (models, lots, serial numbers)
UDI-DI: 00840663101535, 00840663101542. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56. Lots not expired with the following prefix: POC, QPO, XPO.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Pruitt Irrigation Occlusion Catheter 2102-09Lemaitre Vascular, Inc. · Occluder, Catheter Tip1 recall4 reports
- Pruitt Irrigation Occlusion Catheter e2102-09Lemaitre Vascular, Inc. · Occluder, Catheter Tip1 recall0 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K811276 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Pruitt Occlusion Catheter 2103-36Lemaitre Vascular, Inc. · Catheter, Flow Directed1 recall4 reports
- Pruitt Occlusion Catheter 2103-56Lemaitre Vascular, Inc. · Catheter, Flow Directed1 recall1 report
- Pruitt Occlusion Catheter 2103-46Lemaitre Vascular, Inc. · Catheter, Flow Directed1 recall0 reports
- Pruitt Occlusion Catheter e2103-36Lemaitre Vascular, Inc. · Catheter, Flow Directed1 recall0 reports
- Pruitt Occlusion Catheter e2103-46Lemaitre Vascular, Inc. · Catheter, Flow Directed1 recall0 reports
- Pruitt Occlusion Catheter e2103-56Lemaitre Vascular, Inc. · Catheter, Flow Directed1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96716).
- Z-1750-2025 · Open, ClassifiedTufTex Over-the-Wire Embolectomy Catheter. All Catalog numbers: 1651-34, 1651-38, 1651-44, 1651-48, 1651-58, 1651-64, 1651-68, 1651-78, 1651-84, 1651-88, E1651-34, E1651-38, E1651-…
- Z-1752-2025 · Open, ClassifiedPruitt Irrigation Occlusion Catheter. Catalog Numbers: 2102-09, E2102-09.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page