Skip to content

Medtronic LINQ Heart Monitor recallZ-2102-2021 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated Jun 1, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.

FDA root cause category: Software design

What the company is doing

The firm began notifying their consignees on 06/02/2021 by letter. The letter explained the issue, provided patient management recommendations, and reported that a software update was anticipated in late 2021. For questions, contact your local Medtronic Representative or Medtronic Technical Services at 800-929-4043.

Recall details

Recall number
Z-2102-2021
Status
Open, Classified
Initiated by the company
Jun 1, 2021
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Quantity
666210 devices
Product code
MXD
FDA submissions
K162855
Distribution
Worldwide Distribution

Product description

Medtronic Reveal LINQ LNQ11 / PA96000

Code information (models, lots, serial numbers)

All serial numbers distributed on or after February 23, 2017    GTIN: 643169845763, 643169845770, 643169845787, 643169845794, 643169845800, 643169845817, 643169845824, 00643169845831, 00643169845848, 00643169845855, 00643169845862, 00643169845879, 00643169845893, 00763000188382, 00763000188399, 00763000188405, 00763000188412, 00763000188429, 00763000188436, 00763000188443, 00763000188450, 00763000188467, 643169845688, 643169845695, 643169845701, 643169845718, 643169845732, 643169845749, 643169845756

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page