RistTMRecalls & reported problems across all models
Micro Therapeutics, Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls
- 0
We found 3 models of RistTM by Micro Therapeutics, Inc. in the FDA's device registry (GUDID). Together they are named in 26 adverse event reports in the FDA's MAUDE database, received between Aug 1, 2022 and Aug 13, 2026 — classified as 10 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Break” (20 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported RistTM problems
Device problems
- Break20
- Material Deformation11
- Packaging Problem5
- Difficult to Remove3
- Stretched3
- Adverse Event Without Identified Device or Use Problem2
- Difficult to Advance2
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- Vasoconstriction7
- Foreign Body In Patient6
- Device Embedded In Tissue or Plaque3
- Headache2
- Bacterial Infection1
- Blurred Vision1
- Ischemia1
RistTM models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- RistTM 106F-071-95Catheter, percutaneous10 reports
- RistTM 106F-071-100Catheter, percutaneous10 reports
- RistTM 106F-071-105Catheter, percutaneous6 reports
Recent RistTM adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50240422 ↗InjuryAug 4, 2026Break; Difficult to Remove
Medtronic received a report that when the Hospital / lab was done with stroke treatment and trying to remove the RIST catheter, it fractured. A Vascular Surgeon tried pulling it out from the wrist but it did not work. The interventional team went through the groin and tried to snare the RIST but this also did not work. The distal tip of the RIST broke off from the catheter. The RIST catheter remains in the patient.
Patient problems: Foreign Body In Patient; Vasoconstriction
MAUDE report 50126702 ↗MalfunctionMay 22, 2026Material Deformation; Packaging Problem
MEDTRONIC RECEIVED A REPORT THAT A RIST CATHETER WAS DAMAGED OUT OF PACKAGING. A KINK WAS OBSERVED. NO PATIENT COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE DAMAGE WAS ON THE PACKAGED INTRODUCER GUIDE. THE CAUSE OF THE EVENT WAS DETERMINED TO BE KINKED UPON DELIVERY. ALL INFORMATION HAD BEEN CONFIRMED/PROVIDED BY THE PHYSICIAN. ADDITIONAL INFORMATION RECEIVED. FACILITY NAME AND ADDRESS WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25257696 ↗InjuryDec 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT A PATIENT TREATED WITH A PED2 PIPELINE STENT HAD A HEADACHE AND LIGHT SENSITIVITY AND VISION BECOMING BLURRY WITH BRIGHT LIGHT. INTERMITTENT FLASHES OF LIGHTS IN THE PERIPHERAL VISION-MAINLY ON THE LEFT SIDE. STATES SHE HAS INTERMITTENT, DULL, ACHING FRONTAL HEADACHE. THE PATIENT WAS HOSPITALIZED. THE EVENT RESOLVED ON 2023-FEB-14. POSSIBLY DEVICE AND PROCEDURE RELATED. MRI WAS POSITIVE FOR T2 FLAIR AS WELL AS MULTIPLE SMALL SUBACUTE INFARCT POSSIBLE SUBACUTE ISCHEMIA. THE PATIENT WAS BEING TREATED FOR AN ANEURYSM IN THE INTERNAL CAROTID ARTERY (ICA), C6 SEGMENT. LOCATION OF ANEURYSM IN VESSEL: SIDEWALL. ANEURYSM MORPHOLOGY: SACCULAR. MAXIMUM ANEURYSM DIAMETER: 8.5MM. ANEURYSM DOME HEIGHT: 6MM. ANEURYSM DOME WIDTH: 5MM. ANEURYSM NECK SIZE: 3.8MM. PARENT ARTERY DIAMETER DISTAL TO ANEURYSM: 3.2MM. PARENT ARTERY DIAMETER PROXIMAL TO ANEURYSM: 3.7MM. NO STASIS AT THE END OF THE PROCEDURE. COMPLETE NECK COVERAGE AT THE END OF THE PROCEDURE. RAYMOND AND ROY OCCLUSION CLASS 3 AT THE END OF THE PROCEDURE. ACCESS VIA RADIAL RIGHT. ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS HOSPITALIZED FROM FEBRUARY (B)(6) 2023. THE PATIENT WAS GIVEN ATORVAS
Patient problems: Blurred Vision; Headache; Ischemia; Stroke/CVA; Visual Disturbances
MAUDE report 23795282 ↗MalfunctionOct 20, 2025Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING A RIST CATHETER, IT WAS NOTICED THE HUB WAS BREAKING WHEN THE TECHNICAL WAS ABOUT TO LOAD THE NON-MEDTRONIC CATHETER BACK INTO THE RIST. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THE EVENT, AND THE PATIENT OUTCOME WAS REPORTED AS ALIVE WITH NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THE BREAKING WAS COMPLETELY AT THE HUB WHERE THERE WAS SEPARATION. THE PHYSICIAN JUST TRIED TO USED THE CATHETER IT JUST BROKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23342270 ↗MalfunctionJun 2, 2025Break; Material Deformation; Packaging Problem; Break; Material Deformation; Packaging Problem
MEDTRONIC RECEIVED A REPORT THAT THERE WAS "PRODUCT BREAKAGE" WITH A RIST CATHETER. AFTER OPENING THE PRODUCT, A KINK WAS DISCOVERED IN THE MIDDLE SECTION. THE PROBLEM WAS CONFIRMED AFTER OPENING AND BEFORE SOAKING IN THE SALINE SOLUTION. THE CAUSE WAS UNKNOWN, THOUGH IT WAS NOTED THERE WERE NO PARTICULAR PROBLEMS WITH THE HANDLING. A DIFFERENT CATHETER WAS USED INSTEAD. THE PATIENT DID NOT EXPERIENCE ANY HEALTH DAMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22124094 ↗InjuryApr 3, 2025Break; Stretched; Break; Stretched
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE PATIENT RECEIVED THE ANTISPASMODIC MEDICATION PRIOR TO INSERTION OF THE RIST CATHETER.
Patient problems: Foreign Body In Patient; Vasoconstriction
MAUDE report 21765506 ↗MalfunctionApr 2, 2025Break; Fracture; Material Deformation; Break; Fracture; Material Deformation
IT WAS REPORTED THAT DURING A NEUROVASCULAR PROCEDURE INVOLVING THE CATHETER RADIAL GUIDE, SEVERAL ISSUES WERE OBSERVED. THE HUB BROKE OFF FROM THE CATHETER WHERE THE WHITE HUB ENDS ON THE BACK TABLE. ADDITIONALLY, THE DEVICE HAD A KINK ON THE PROXIMAL END NEAR THE HUB, AND BLOOD WAS NOTICED COMING FROM THE DISTAL END OF THE DEVICE. THE DEVICE WAS ALSO CRACKED ON THE PROXIMAL END NEAR THE HUB. DESPITE THESE ISSUES, THE PROCEDURE WAS COMPLETED SAFELY WITHOUT INJURY TO THE PATIENT, AND THE CATHETER WAS REMOVED AS SOON AS POSSIBLE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21753801 ↗MalfunctionApr 2, 2025Break; Fracture; Material Deformation; Break; Fracture; Material Deformation
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE PHYSICIAN PLACED AN EXCHANGE WIRE IN THE PATIENT AND SWITCHED OUT THE CATHETER WITH ANOTHER RIST CATHETER TO RESOLVE THE CATHETER/HUB BREAK. THE DEVICE HAD A CRACK ON THE PROXIMAL END SO BLOOD WAS LEAKING OUT OF THE SIDE OF THE CATHETER. THE ACTUAL CAUSE WAS NOT DETERMINED. TWO OTHER RIST CATHETERS THAT WERE BROKEN OR KINKED GETTING PULLED OUT OF THE PACKAGE WERE ALSO NOTED. IT WAS MOST LIKELY THE CAUSE OF THE BREAK IN THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21753788 ↗MalfunctionMar 27, 2025Break; Packaging Problem; Break; Packaging Problem
IT WAS REPORTED THAT THE HUB WAS NOT ATTACHED TO THE CATHETER FOR THE CATHETER RADIAL GUIDE 106F-071-95. THIS ISSUE OCCURRED TWICE WHEN THE PHYSICIAN REMOVED THE 6F RIST CATHETER FROM THE PACKAGING AND NOTICED THE HUB WAS NOT ATTACHED. THE PHYSICIAN DISCARDED THE FIRST UNIT, OPENED ANOTHER FROM THE SAME LOT, AND ENCOUNTERED THE SAME ISSUE. NEITHER DEVICE WAS USED IN THE PROCEDURE, AND THE PHYSICIAN OPTED TO UTILIZE A COMPETITIVE GUIDE SHEATH TO COMPLETE THE CASE. THERE WAS NO PATIENT INVOLVED. ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WAS NO DAMAGE OR EVIDENCE OF TAMPERING TO DEVICE PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21713469 ↗MalfunctionMar 27, 2025Break; Packaging Problem; Break; Packaging Problem
IT WAS REPORTED THAT THE HUB WAS NOT ATTACHED TO THE CATHETER FOR THE CATHETER RADIAL GUIDE 106F-071-95. THIS ISSUE OCCURRED TWICE WHEN THE PHYSICIAN REMOVED THE 6F RIST CATHETER FROM THE PACKAGING AND NOTICED THE HUB WAS NOT ATTACHED. THE PHYSICIAN DISCARDED THE FIRST UNIT, OPENED ANOTHER FROM THE SAME LOT, AND ENCOUNTERED THE SAME ISSUE. NEITHER DEVICE WAS USED IN THE PROCEDURE, AND THE PHYSICIAN OPTED TO UTILIZE A COMPETITIVE GUIDE SHEATH TO COMPLETE THE CASE. THERE WAS NO PATIENT INVOLVED. ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WAS NO DAMAGE OR EVIDENCE OF TAMPERING TO DEVICE PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21713453 ↗InjuryJan 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THE CAUSE OF THE EVENT WAS NOT DETERMINED. THIS INFORMATION HAS BEEN CONFIRMED WITH THE PHYSICIAN. ADDITIONAL INFORMATION WAS RECEIVED REPORTING THE ADVERSE EVENT (AE) WAS NOT RELATED TO THE DISEASE UNDER STUDY. THE SITE ASSESSED THIS EVENT AS NOT RELATED TO ANTIPLATELET MEDICATION OR DEVICE AND POSSIBLY RELATED TO THE STUDY PROCEDURE.
Patient problems: Bacterial Infection; Headache; Pain; Vasoconstriction
MAUDE report 21273193 ↗MalfunctionNov 8, 2024Break; Leak/Splash; Break; Leak/Splash
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE CAUSE OF THE CATHETER SEPARATION/BREAK AND LEAK WAS NOT DETERMINED. IT WAS CLARIFIED THAT THE BREAK AND LEAK OCCUR AT THE HUB THAT CONNECTS TO THE CATHETER. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EVENT/INJURY STOPPING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20640899 ↗InjuryApr 11, 2024Device Dislodged or Dislocated; Material Deformation; Physical Resistance/Sticking; Device Dislodged or Dislocated; Material Deformation; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THE PATIENT IS CURRENTLY RECOVERING WITHOUT ISSUE POST PROCEDURE PER PHYSICIAN. THE CAUSE OF THE RESISTANCE/SEPARATION WAS NOT DETERMINED.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 19088198 ↗MalfunctionApr 5, 2024Break; Packaging Problem; Break; Packaging Problem
ADDITIONAL INFORMATION RECEIVED REPORTED THERE WAS NO EVIDENCE OF OF TAMPERING AND BOX AND PACKAGING LOOKED UNDAMAGED AS WELL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19052732 ↗
Other Micro Therapeutics, Inc. products
- PipelineTM Flex Embolization Device with Shield TechnologyTM79 models · 1,020 reports
- AxiumTM139 models · 832 reports · 1 recall
- PIPELINE114 models · 622 reports · 1 recall
- Concerto Detachable Coil System34 models · 595 reports
- Axium Prime Detachable Coil System70 models · 497 reports
- ConcertoTM70 models · 436 reports
- EchelonTM5 models · 225 reports · 2 recalls
- Axium Detachable Coil System69 models · 288 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page