SAPHYRERecalls & reported problems across all models
Smith & Nephew, Inc. · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls
- 0
We found 4 models of SAPHYRE by Smith & Nephew, Inc. in the FDA's device registry (GUDID). Together they are named in 22 adverse event reports in the FDA's MAUDE database, received between Dec 29, 2017 and Nov 4, 2024 — classified as 7 injuries, 15 malfunctions.
The most frequently reported specific device problem is “Material Separation” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported SAPHYRE problems
Device problems
- Material Separation7
- Overheating of Device5
- Break4
- Adverse Event Without Identified Device or Use Problem1
- Arcing1
- Crack1
- Detachment of Device or Device Component1
- Electrical Shorting1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- No Known Impact Or Consequence To Patient8
- Foreign Body In Patient2
SAPHYRE models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- SAPHYRE 7209686ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES14 reports
- SAPHYRE 7209685ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES4 reports
- SAPHYRE 7210113ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES3 reports
- SAPHYRE 7209684ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 report
Recent SAPHYRE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionNov 4, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING A SHOULDER SURGERY, THE TEMPERATURE OF THE SAPHYRE PROBE WAS TOO HIGH AFTER USING IT FOR MORE THAN 10 MINUTES, THE SURGEON WAS AFRAID TO USE IT ANYMORE. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20598640 ↗MalfunctionSep 24, 2024Break; Break
IT WAS REPORTED THAT DURING A JOINT CAVITY DEBRIDEMENT WHEN THE PROBE SAPHYRE WAND WAS USED FOR THE FIRST TIME, THE INSULATING LAYER WAS SPLIT INSIDE THE ARTICULAR CAVITY, THE WAND WAS QUICKLY REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING AN S+N BACKUP DEVICE. THERE WAS NO SURGICAL DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20296525 ↗InjurySep 26, 2023Material Separation; Material Separation
IT WAS REPORTED THAT ON (B)(6) 2023, A LEFT SHOULDER ARTHROSCOPIC EXPLORATION AND CLEANING, ADHESION RELEASE, SUBACROMIAL BURSA RESECTION AND ACROMIOPLASTY WERE PERFORMED IN THE OPERATING ROOM UNDER GENERAL ANESTHESIA. AFTER THE SURGERY WAS COMPLETED, AND THE INCISION HAD BEEN CLOSED. WHEN THE NURSE COUNTED THE SURGICAL ITEMS, IT WAS NOTICED, THAT THE ELECTRODE BLADE AT THE FRONT END FROM THE SAPHYRE PROBE WAS MISSING. AN X-RAY WAS PERFORMED AND REVIEWED ON THE SAME DAY. WHICH SUGGESTED, THERE WAS A METAL FOREIGN BODY UNDER THE PATIENT'S LEFT ACROMION, WITH A DIAMETER OF ABOUT 1.8 MM. ON THE MORNING OF (B)(6), THE PATIENT WAS SATISFIED WITH A RELIEF OF HIS SYMPTOMS. NO DISCOMFORT WAS REPORTED. SINCE THE REPORTED REMAINING FRAGMENTS WERE VERY SMALL, THE DOCTORS DID NOT CONSIDER THEM TO BE OF MUCH CONSEQUENCE. IT WAS THEN DECIDED, WITH THE PATIENT'S CONSENT, THAT AN ADDITIONAL OPERATION WOULD NOT BE SCHEDULED TO REMOVE THE REMAINING FRAGMENTS.
Patient problems: Foreign Body In Patient
MAUDE report 17821624 ↗InjuryNov 14, 2022Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE TIP AND THE SHAFT OF THE PROBE SAPHYRE WAND, GOT SEPARATED WHEN USED FOR 5 MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15791903 ↗InjuryNov 10, 2022Break; Break
IT WAS REPORTED THAT DURING A MENISCUS REPAIR PROCEDURE, THE SAPHYRE PROBE ELECTRODE WAS FOUND BROKEN 5 MINUTES AFTER ENTERING THE JOINT CAVITY. THE BROKEN PARTS WERE TAKEN OUT USING TWEEZERS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NON-SIGNIFICANT DELAY USING A BACK-UP DEVICE. THE PATIENT IS REPORTED TO BE FINE AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15770439 ↗MalfunctionSep 27, 2022Crack; Crack
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, IN THE JOINT CAVITY, THE SKIN OF THE PROBE SAPHYRE WAND CRACK AT THE BEGINNING OF THE ABLATION. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A NON-SIGNIFICANT SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15491797 ↗MalfunctionAug 19, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT, AFTER AN UBE OPERATION OF SPINE, THE TEMPERATURE OF THE PROBE SAPHYRE BLADE WAS TOO HIGH. SURGERY WAS FINISHED WITHOUT DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15259473 ↗InjuryJun 3, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING A KNEE ARTHROSCOPIC EXPLORATION AND MENISCECTOMY, THE FRONT END OF A PROBE SAPHYRE WAND FELL OFF INTO THE BODY CAVITY. THE DOCTOR CAREFULLY LOOKED FOR THE FALLEN PART TO MAKE SURE THERE WAS NO RESIDUE IN THE PATIENT, THE RESIDUES WERE REMOVED USING TWEEZER. SURGERY WAS RESUMED WITH A BACK-UP DEVICE, AFTER A DELAY GREATER THAN 30 MIN. NO FURTHER COMPLICATIONS WERE REPORTED. THE CURRENT HEALTH STATUS OF THE PATIENT IS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14594476 ↗InjuryDec 10, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC TREATMENT USING A SAPHYRE 90° PROBE; THE LENS OF THE ARTHROSCOPE GOT DAMAGED DUE TO IMPROPER OPERATION. IN CONSEQUENCE, THE LENS APERTURE BECAME SMALLER AFFECTING THE VISUAL FIELD MAKING IT INOPERABLE. THE SAPHYRE DID NOT BREAK INTO FRAGMENTS INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED BY CHANGING THE SURGICAL METHOD. THE SURGICAL METHOD CHANGED FROM ARTHROSCOPIC TO THE TRADITIONAL WHICH THE INCISION WAS LARGER. NO DELAYS NOR FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12976912 ↗InjuryAug 8, 2021Material Separation; Material Separation
IT WAS REPORTED THAT DURING A BILATERAL ARTHROSCOPY AND BILATERAL HIGH TIBIAL OSTEOTOMY, THE ELECTRODE OF THE PROBE FELL INTO THE PATIENT. ALTHOUGH THERE WAS NO SIGNIFICANT DELAY, IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12292235 ↗MalfunctionJul 7, 2021Material Separation; Material Deformation; Material Separation; Material Deformation
IT WAS REPORTED THAT DURING USE FOR 20 MINUTES, USING THE "PROBE SAPHYRE 90DEG", THE HEAD METAL HAD A DEFORMATION FRACTURE AND A RUBBER DEFORMATION CONTRACTURE. IT IS UNKNOWN IF THERE WAS A DELAY. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12129703 ↗InjuryJan 30, 2021Break; Break
IT WAS REPORTED THAT, DURING A SHOULDER ROTATOR CUFF REPAIR SURGERY, THE PROBE ELECTRODE FALL OFF. DELAY WAS GREATER THAN 30 MINUTES AND A BACK UP WAS AVAILABLE TO COMPLETE THE PROCEDURE. THE PIECES WERE SUCCESSFULLY REMOVED FROM THE PATIENT. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11256330 ↗MalfunctionJan 12, 2021Failure to Deliver Energy; Failure to Deliver Energy
IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY, THE PROBE HEAD HAS NO POSITIVE ENERGY OUTPUT AND CANNOT BE USED WHILE BEING USED IN DEFAULT SETTINGS. THE PROBE HEAD IS FOR SINGLE USE. THE HOST SHUTDOWN AND RESTART STILL CAN NOT BE USED, FINALLY A NEW PROBE HEAD WAS USED TO COMPLETE THE SURGERY WITH A DELAY GREATER THAN 30 MINUTES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11158139 ↗MalfunctionNov 20, 2020Overheating of Device; Electrical Shorting; Overheating of Device; Electrical Shorting
IT WAS REPORTED THAT DURING AN ACL AFTER THE WAND WAS CONNECTED, THE TIP SPARK AND DEVICE OVERHEATED. THE PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE AND NO SIGNIFICANT DELAY OR FURTHER COMPLICATIONS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10879962 ↗MalfunctionApr 21, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING ARTHROSCOPY OF THE LEFT SHOULDER, JOINT CAVITY CLEANING, ACROMIOPLASTY, AND ROTATOR CUFF AVULSION SURGERY THE PROBE SAPHYRE TIP SHAFT SEPARATED WHEN IT WAS USED FOR 2 MINUTES. THE DEVICE BREAKS INSIDE THE PATIENT AND WAS RETRIEVED USING SUCTION. THERE WAS AN AVAILABLE BACKUP DEVICE. NO DELAY WAS REPORTED. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 9984389 ↗
Other Smith & Nephew, Inc. products
- FAST-FIX18 models · 9,554 reports
- JOURNEY566 models · 6,250 reports · 4 recalls
- GENESIS II462 models · 4,599 reports · 2 recalls
- DYONICS, POWERMAX, ELITE2 models · 4,308 reports
- TRIGEN381 models · 3,599 reports · 5 recalls
- LEGION446 models · 2,039 reports · 4 recalls
- REAL INTELLIGENCE32 models · 2,017 reports · 3 recalls
- Coblation62 models · 2,126 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page