Skip to content

SAPHYRE 7209685Recalls & reported problems

Smith & Nephew, Inc. · PROBE SAPHYRE 60 DEG 3MM · Class 2

All models of SAPHYRE (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010470287), received between Dec 29, 2017 and Nov 4, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device2
  • Detachment of Device or Device Component1
  • Material Separation1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A SHOULDER SURGERY, THE TEMPERATURE OF THE SAPHYRE PROBE WAS TOO HIGH AFTER USING IT FOR MORE THAN 10 MINUTES, THE SURGEON WAS AFRAID TO USE IT ANYMORE. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20598640 ↗
  • InjuryJun 3, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING A KNEE ARTHROSCOPIC EXPLORATION AND MENISCECTOMY, THE FRONT END OF A PROBE SAPHYRE WAND FELL OFF INTO THE BODY CAVITY. THE DOCTOR CAREFULLY LOOKED FOR THE FALLEN PART TO MAKE SURE THERE WAS NO RESIDUE IN THE PATIENT, THE RESIDUES WERE REMOVED USING TWEEZER. SURGERY WAS RESUMED WITH A BACK-UP DEVICE, AFTER A DELAY GREATER THAN 30 MIN. NO FURTHER COMPLICATIONS WERE REPORTED. THE CURRENT HEALTH STATUS OF THE PATIENT IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14594476 ↗
  • InjuryAug 8, 2021Material Separation; Material Separation

    IT WAS REPORTED THAT DURING A BILATERAL ARTHROSCOPY AND BILATERAL HIGH TIBIAL OSTEOTOMY, THE ELECTRODE OF THE PROBE FELL INTO THE PATIENT. ALTHOUGH THERE WAS NO SIGNIFICANT DELAY, IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12292235 ↗
  • MalfunctionDec 29, 2017Overheating of Device; Smoking

    IT WAS REPORTED THAT DURING A HIP LABIAL REPAIR SURGERY IT WAS FOUND THAT THE HANDLE OF THE PROBE WAS OVERHEATED AND SCALDING, AS WELL AS SLIGHT SMOKE/STEAM WAS OBSERVED. FINALLY, A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7153752 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page