Secura VRRecalls & reported problems across all models
Medtronic, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 115
- Deaths reported
- 1
- Injuries reported
- 78
- Recalls
- 0
We found 2 models of Secura VR by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 115 adverse event reports in the FDA's MAUDE database, received between Apr 6, 2017 and Apr 1, 2025 — classified as 1 death, 78 injuries, 36 malfunctions.
The most frequently reported specific device problem is “Failure to Charge” (28 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Secura VR problems
Device problems
- Failure to Charge28
- Inappropriate/Inadequate Shock/Stimulation18
- Battery Problem15
- Delayed Charge Time15
- Electrical /Electronic Property Problem15
- Premature Discharge of Battery15
- Adverse Event Without Identified Device or Use Problem13
- Failure to Interrogate12
Patient problems
- No Known Impact Or Consequence To Patient45
- No Clinical Signs, Symptoms or Conditions29
- Shock from Patient Lead(s)20
- Unspecified Infection10
- No Consequences Or Impact To Patient4
- Endocarditis3
- Arrhythmia1
- Bacteremia1
Secura VR models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Secura® VR D224VRCDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 death101 reports
- Secura® VR D204VRMDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)14 reports
Recent Secura VR adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2025Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE REMOTE MONITOR WAS UNABLE TO ESTABLISH TELEMETRY. TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE PATIENT WAS ADVISED TO GET IN CONTACT WITH CARDIOLOGIST FOR A DEVICE CHECK. NO FURTHER INFORMATION COULD BE OBTAINED THROUGH FOLLOW-UP. THE MOBILE MONITORING APPLICATION AND IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21748775 ↗InjuryFeb 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED DUE TO A POCKET INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 21473790 ↗InjuryAug 9, 2024Failure to Charge; Electrical /Electronic Property Problem; Delayed Charge Time; Failure to Charge; Electrical /Electronic Property Problem; Delayed Charge Time
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD REACHED RECOMMENDED REPLACEMENT TIME (RRT) TWO YEARS AGO. THE ICD WAS NOW AT END OF SERVICE (EOS) AND CHARGE CIRCUIT INACTIVE OCCURRED. ALERTS WERE TRIGGERED FOR CHARGE CIRCUIT TIMEOUT AND FOR A LONG CHARGE TIME GREATER THAN 30 SECONDS. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD CAPTURE THRESHOLD WAS UNAVAILABLE, AND CAPTURE COULD NOT BE CONFIRMED. THE RV LEAD REMAINS IN USE, AND THE ICD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19954136 ↗InjuryJul 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT C OMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 19680269 ↗MalfunctionApr 10, 2024Failure to Charge; Electrical /Electronic Property Problem; Delayed Charge Time; Failure to Charge; Electrical /Electronic Property Problem; Delayed Charge Time
IT WAS REPORTED THAT THE PATIENT HAD FELL AND THEIR DEVICE REACHED END OF SERVICE (EOS) AND TRIGGERED ALERTS FOR LONGER THAN EXPECTED CHARGE TIME AT ELECTIVE REPLACEMENT INDICATOR (ERI) AND CHARGE CIRCUIT TIMEOUT. ADDITIONALLY, THE DEVICE EXHIBITED CHARGE CIRCUIT INACTIVE. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Fall; No Clinical Signs, Symptoms or Conditions
MAUDE report 19081934 ↗InjuryMar 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 18995724 ↗InjuryJan 12, 2024Failure to Charge; Delayed Charge Time; Failure to Charge; Delayed Charge Time
IT WAS FURTHER REPORTED THAT THE DEVICE WAS EXPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18504394 ↗MalfunctionOct 27, 2023Battery Problem; Communication or Transmission Problem; Reset Problem; Battery Problem; Communication or Transmission Problem; Reset Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI)/ RECOMMENDED REPLACEMENT TIME (RRT) SEVERAL YEARS AGO AND WAS ELECTIVELY LEFT IN USE WELL AFTER ERI/RRT. THE DEVICE HAD REACH END OF SERVICE (EOS). THE DEVICE WAS NOTED TO HAVE AN ELECTRICAL RESET AND WIRELESS TELEMETRY WAS NO LONGER AVAILABLE. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18025211 ↗InjuryOct 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO ENDOCARDITIS, METHICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS AND BACTEREMIA. IT WAS NOTED THAT THE INFECTION SOURCE WAS SUSPECTED TO BE A CHEMOTHERAPY PORT; HOWEVER, VEGETATION WAS NOTED ON THE ICD SYSTEM. A NEW ICD SYSTEM WAS IMPLANTED A FEW DAYS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Drug Resistant Bacterial Infection; Endocarditis
MAUDE report 17977545 ↗MalfunctionJul 31, 2023Failure to Charge; Protective Measures Problem; Failure to Charge; Protective Measures Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) ALERTED DUE TO A CHARGE CIRCUIT TIMEOUT. IT WAS NOTED THAT THE DEVICE HAD REACHED END OF SERVICE (EOS) THREE YEARS PRIOR. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17425372 ↗InjuryJun 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD DEVELOPED AN INFECTION OF UNKNOWN ORIGIN WITHIN THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) POCKET. THE ICD SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 17218339 ↗MalfunctionMay 22, 2023Failure to Charge; Failure to Charge
IT WAS REPORTED THAT THE PATIENT CALLED TO REPORT HEARING A BEEPING NOISE FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THE PATIENT MENTIONED THAT THE DEVICE WAS DUE FOR A CHANGE OUT SEVERAL YEARS AGO AND THE PATIENT CHOSE TO LEAVE THE DEVICE IMPLANTED AND JUST TURN OFF THE DEVICE. THE DEVICE ALERTED FOR A CHARGE CIRCUIT TIMEOUT AND HAD REACHED RECOMMENDED REPLACEMENT TIME (RRT) WITH ALL DETECTIONS PROGRAMMED OFF. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16974550 ↗MalfunctionMay 15, 2023Failure to Charge; Failure to Charge
IT WAS REPORTED, THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAS BEEN AT END OF SERVICE (EOS). FOR A CONSIDERABLE AMOUNT OF TIME SINCE THE PATIENT AND CARDIOLOGIST DECIDED, THAT THE ICD SHOULD BE TURNED OFF, AND NOT REPLACED/REMOVED. AS A RESULT OF THE DEVICE AGE, THE CHARGE CIRCUIT TIMEOUT IS NOW OCCURRING. THE DEVICE REMAINS INACTIVATED. AND IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16932054 ↗InjuryMay 3, 2023Failure to Charge; Electrical /Electronic Property Problem; Delayed Charge Time; Failure to Charge; Electrical /Electronic Property Problem; Delayed Charge Time
IT WAS REPORTED THAT AFTER THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD REACHED END OF SERVICE (EOS) A CHARGE CIRCUIT INACTIVE AND CHARGE CIRCUIT OCCURRED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16863026 ↗MalfunctionApr 13, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT APPROXIMATELY THREE YEARS AFTER THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) TRIGGERED RECOMMENDED REPLACEMENT TIME (RRT), IT TRIGGERED AN ALERT DUE TO THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NOT HAVING ENOUGH ENERGY TO FORM THE CAPACITORS. THE PATIENT OPTED TO HAVE THE DEVICE REMOVED AND THE DEVICE WAS LATER EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16732847 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page