Sertera Biopsy DeviceRecalls & reported problems across all models
Hologic, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one Sertera Biopsy Device model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 2
We found 2 models of Sertera Biopsy Device by Hologic, Inc. in the FDA's device registry (GUDID). Together they are named in 16 adverse event reports in the FDA's MAUDE database, received between Jan 14, 2022 and Sep 30, 2026 — classified as 2 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (8 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Sertera Biopsy Device recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jan 15, 2025Open, Classified
Devices presented a condition in which the inner needle separated from the hub and produced a longer throw of the inner needle or dislodgement.
Z-1170-2025 · Root cause: Process control · Recall details · FDA record ↗
- May 2, 2023Open, Classified
The inner needle of the Sertera device is either detaching or discharging and may cause delay in surgery
Z-2035-2023 · Root cause: Device Design · Recall details · FDA record ↗
Most reported Sertera Biopsy Device problems
Device problems
- Mechanical Problem8
- Appropriate Device Problem Term/Code Not Available2
- Detachment of Device or Device Component2
- Failure to Obtain Sample2
- Material Separation2
Patient problems
- No Clinical Signs, Symptoms or Conditions16
Sertera Biopsy Device models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Sertera Biopsy Device Sertera-14INSTRUMENT, BIOPSY2 recalls14 reports
- Sertera Biopsy Device Sertera-12INSTRUMENT, BIOPSY2 reports
Recent Sertera Biopsy Device adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Failure to Obtain Sample
It was reported by the user site that during a breast biopsy procedure on (b)(6)2026, "The radiologist was doing an u/s "[ultrasound]" guided core biopsy and stated that he took 3 samples and all three samples came back "shredded". He then switched needles and the new needle samples were good." No patient and/or user harm was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942525 ↗InjuryJan 29, 2025Failure to Obtain Sample; Failure to Obtain Sample
IT WAS REPORTED THAT DURING A SERTERA PROCEDURE ON (B)(6) 2025, WHEN THE RADIOLOGIST TOOK THE SAMPLE THE DEVICE WAS STUCK AND WOULDN´T LET THEM UNSHEATHE TO GET THE CORE. THE CUSTOMER RESPONDED THAT THE TISSUE WAS LOST AND COULDN´T BE USED. NO OTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21261425 ↗MalfunctionMay 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT ON (B)(6), A SERTERA PROCEDURE WAS PERFORMED AND WHEN THE RADIOLOGIST FIRED THE NEEDLE THE FIRST TIME OUTSIDE THE PATIENT NOTICED THE NEEDLE LOOKED TOO LONG AND FULLY FIRED SO THEY DECIDED TO COCKED IT AGAIN TO TEST IT. THIS TIME THE RADIOLOGIST NOTICED THE TROUGH SEEMED TO NOT BE FULLY RETRACTED IN THE NEEDLE WHEN COCKED. WHEN THE RADIOLOGIST FIRED THE NEEDLE CAME OUT AND HIT THE WALL AND BOUNCED OFF THE WALL AND HIT THE FLOOR. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19372201 ↗MalfunctionMar 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A SERTERA PROCEDURE ON MARCH 7TH, THE PHYSICIAN PRE-TESTED THE NEEDLE BEFORE USING IT ON THE PATIENT AND THE NEEDLE DISCONNECTED AND DISCHARGED WHILE IN THE PHYSICIAN´S HAND. NO PATIENT OR STAFF INJURY REPORTED. NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16611840 ↗MalfunctionFeb 17, 2023Material Separation; Material Separation
IT WAS REPORTED THAT ON JANUARY 25TH, A SERTERA PROCEDURE WAS PERFORMED AND THE NEEDLE DISLODGED WHILE RADIOLOGIST PERFORMED A TEST FIRE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16400831 ↗MalfunctionFeb 17, 2023Material Separation; Material Separation
IT WAS REPORTED THAT ON JANUARY 25TH, A SERTERA PROCEDURE WAS PERFORMED AND UPON FIRING, TROUGH DISLODGED FROM THE DEVICE INTO THE PATIENT. WHEN PULLING THE DEVICE OUT THE TROUGH REMAINED IN THE PATIENT. HAD TO BE REMOVED MANUALLY. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16397624 ↗MalfunctionDec 21, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2022, A SERTERA PROCEDURE WAS PERFORMED AND THE NEEDLE UNIT IS FIRING ON IT'S OWN AFTER BEING TESTED AND WITHOUT BEING IN HANDS OF PHYSICIAN. THE DOCTOR DID NOT NOTICE ANY ISSUE UNTIL THE ISSUE OCCURRED DURING TESTING. THE INNER CANNULA DETACHED AFTER FIRING. THE DEVICE WAS COCKED FOR TESTING PURPOSES PRIOR TO USE ON THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16029249 ↗MalfunctionDec 21, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) "2022", A SERTERA PROCEDURE WAS PERFORMED AND THE NEEDLE UNIT IS FIRING ON IT'S OWN AFTER BEING TESTED AND WITHOUT BEING IN HANDS OF PHYSICIAN. THE DOCTOR DID NOT NOTICE ANY ISSUE UNTIL THE ISSUE OCCURRED DURING TESTING. THE INNER CANNULA DETACHED AFTER FIRING. THE DEVICE WAS COCKED FOR TESTING PURPOSES PRIOR TO USE ON THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16029051 ↗MalfunctionDec 21, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2022, A SERTERA PROCEDURE WAS PERFORMED AND THE NEEDLE UNIT IS FIRING ON IT'S OWN AFTER BEING TESTED AND WITHOUT BEING IN HANDS OF PHYSICIAN. THE DOCTOR DID NOT NOTICE ANY ISSUE UNTIL THE ISSUE OCCURRED DURING TESTING. THE INNER CANNULA DETACHED AFTER FIRING. THE DEVICE WAS COCKED FOR TESTING PURPOSES PRIOR TO USE ON THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16028994 ↗MalfunctionDec 21, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2022, A SERTERA PROCEDURE WAS PERFORMED AND THE NEEDLE UNIT IS FIRING ON IT'S OWN AFTER BEING TESTED AND WITHOUT BEING IN HANDS OF PHYSICIAN. THE DOCTOR DID NOT NOTICE ANY ISSUE UNTIL THE ISSUE OCCURRED DURING TESTING. THE INNER CANNULA DETACHED AFTER FIRING. THE DEVICE WAS COCKED FOR TESTING PURPOSES PRIOR TO USE ON THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16028975 ↗MalfunctionDec 21, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2022, A SERTERA PROCEDURE WAS PERFORMED AND THE NEEDLE UNIT IS FIRING ON IT'S OWN AFTER BEING TESTED AND WITHOUT BEING IN HANDS OF PHYSICIAN. THE DOCTOR DID NOT NOTICE ANY ISSUE UNTIL THE ISSUE OCCURRED DURING TESTING. THE INNER CANNULA DETACHED AFTER FIRING. THE DEVICE WAS COCKED FOR TESTING PURPOSES PRIOR TO USE ON THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16028962 ↗MalfunctionDec 15, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2022 A SERTERA PROCEDURE WAS PERFORMED AND THE NEEDLE UNIT IS FIRING ON ITS OWN AFTER BEING TESTED AND WITHOUT BEING IN HANDS OF PHYSICIAN. THE DOCTOR DID NOT NOTICE ANY ISSUE UNTIL THE ISSUE OCCURRED DURING TESTING. THE INNER CANNULA DETACHED AFTER FIRING. THE DEVICE WAS COCKED FOR TESTING PURPOSES PRIOR TO USE ON THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15992060 ↗MalfunctionNov 30, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2022, A SERTERA PROCEDURE WAS PERFORMED, AND THE DEVICE MADE A RATTLING NOISE. THE DEVICE WAS NOT ABLE TO GET A CORE WHEN INSERTED INTO THE PATIENT'S SKIN AND THEN A PIECE GOT DISLODGED AND DISCHARGED FROM THE DEVICE . THE PROCEDURE WAS COMPLETED WITH A SECOND DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15889847 ↗MalfunctionNov 30, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2022 A SERTERA PROCEDURE WAS PERFORMED, AND THE FIRST NEEDLE, THE ONE THAT THE DOCTOR PRE-FIRED GOT DISLODGED AND WAS DISCHARGED FROM THE HANDPIECE. NO HARM OR INJURY WAS REPORTED TO THE PATIENT OR USER. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15889734 ↗MalfunctionJan 14, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT ON (B)(6) 2021, DURING A SERTERA BIOPSY PROCEDURE , TWO NEEDLES WERE USED INSIDE THE BREAST AND COULD NO COLLECT SAMPLES. THE PROCEDURE HAD TO BE ABORTED AND RESCHEDULED. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13258689 ↗
Other Hologic, Inc. products
- BioZorb9 models · 326 reports · 3 recalls
- Selenia Dimensions9 models · 245 reports · 3 recalls
- NovaSure5 models · 344 reports
- Brevera Breast Biopsy System5 models · 300 reports
- 3Dimensions9 models · 67 reports · 3 recalls
- Aptima15 models · 82 reports · 2 recalls
- NovaSure SureSound2 models · 182 reports
- Panther Fusion3 models · 9 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page