Single Use Electrosurgical KnifeRecalls & reported problems across all models
Olympus Medical Systems Corp. · 9 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one Single Use Electrosurgical Knife model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 509
- Deaths reported
- 4
- Injuries reported
- 295
- Recalls
- 2
We found 9 models of Single Use Electrosurgical Knife by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 509 adverse event reports in the FDA's MAUDE database, received between Jun 21, 2018 and Aug 7, 2025 — classified as 4 deaths, 295 injuries, 210 malfunctions.
The most frequently reported specific device problem is “Break” (255 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Single Use Electrosurgical Knife recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Mar 1, 2024Open, Classified
Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury
Z-1547-2024 · Root cause: Use error · Recall details · FDA record ↗
- Feb 17, 2024Open, Classified
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
Z-1222-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported Single Use Electrosurgical Knife problems
Device problems
- Break255
- Adverse Event Without Identified Device or Use Problem185
- Detachment of Device or Device Component130
- Insufficient Device Problem Information7
- Crack5
- Material Split, Cut or Torn3
- Melted3
- Material Fragmentation2
Patient problems
- No Clinical Signs, Symptoms or Conditions293
- Hemorrhage/Blood Loss/Bleeding90
- Perforation88
- Foreign Body In Patient20
- Fever14
- No Consequences Or Impact To Patient12
- Unspecified Infection10
- Pleural Effusion9
Single Use Electrosurgical Knife models
9 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Single Use Electrosurgical Knife KD-611LUnit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Single Use Electrosurgical Knife KD-640LUnit, electrosurgical, endoscopic (with or without accessories)2 recalls19 reports
- Single Use Electrosurgical Knife KD-620LRUnit, electrosurgical, endoscopic (with or without accessories)1 death48 reports
- Single Use Electrosurgical Knife KD-610LUnit, electrosurgical, endoscopic (with or without accessories)2 deaths42 reports
- Single Use Electrosurgical Knife KD-630LUnit, electrosurgical, endoscopic (with or without accessories)17 reports
- Single Use Electrosurgical Knife KD-620QRUnit, Electrosurgical, Endoscopic (With Or Without Accessories)11 reports
- Single Use Electrosurgical Knife KD-620LRUnit, Electrosurgical, Endoscopic (With Or Without Accessories)10 reports
- Single Use Electrosurgical Knife KD-620QRUnit, electrosurgical, endoscopic (with or without accessories)4 reports
- Single Use Electrosurgical Knife KD-620URUnit, electrosurgical, endoscopic (with or without accessories)1 report
Recent Single Use Electrosurgical Knife adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryAug 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC SEPTOTOMY PROCEDURE, THE TIP OF THE SINGLE USE ELECTROSURGICAL KNIFE FELL OFF INTO THE PATIENT'S STOMACH. IT WAS OBSERVED THAT THE OPERATOR WAS USING THE DEVICE WITH A LOT OF ENERGY IN THE CUT MODE. THE OPERATOR LOOKED FOR THE TIP THAT FELL OFF INSIDE THE PATIENT BUT COULD NOT FIND IT. THE FALLEN TIP WAS NOT RETRIEVED. THE USERS REPORTEDLY WERE NOT WORRIED ABOUT IT TOO MUCH. THE PROCEDURE WAS PROLONGED BY LESS THAN 30 MINUTES AND COMPLETED SUCCESSFULLY WITH A DIFFERENT KNIFE. THE PATIENT WAS STABLE AND THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: Foreign Body In Patient
MAUDE report 22742969 ↗InjuryAug 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC SEPTOTOMY PROCEDURE, THE TIP OF THE SINGLE USE ELECTROSURGICAL KNIFE FELL OFF INTO THE PATIENT'S STOMACH. IT WAS OBSERVED THAT THE OPERATOR WAS USING THE DEVICE WITH A LOT OF ENERGY IN THE CUT MODE. THE OPERATOR LOOKED FOR THE TIP THAT FELL OFF INSIDE THE PATIENT BUT COULD NOT FIND IT. THE FALLEN TIP WAS NOT RETRIEVED. THE USERS REPORTEDLY WERE NOT WORRIED ABOUT IT TOO MUCH. THE PROCEDURE WAS PROLONGED BY LESS THAN 30 MINUTES AND COMPLETED SUCCESSFULLY WITH A DIFFERENT KNIFE. THE PATIENT WAS STABLE AND THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: Foreign Body In Patient
MAUDE report 22742955 ↗InjuryJun 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER =2 CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER <2 CM (2.6%) (P< 0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%) (P
Patient problems: Bowel Perforation; Hemorrhage/Blood Loss/Bleeding; Unspecified Infection
MAUDE report 22133183 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER = 2 CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER < 2 CM (2.6%) (P < 0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%)
Patient problems: Unspecified Infection
MAUDE report 22127712 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY: BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER =2CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER <2CM (2.6%) (P<0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%) (P=0.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 22127675 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER = 2 CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER < 2 CM (2.6%) (P < 0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%)
Patient problems: Unspecified Infection
MAUDE report 22127660 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER =2 CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER < 2 CM (2.6%) (P <0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%) (P
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 22127654 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER = 2 CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER < 2CM (2.6%) (P< 0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%) (P
Patient problems: Fistula; Unspecified Infection
MAUDE report 22127547 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER = 2 CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER < 2 CM (2.6%) (P < 0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%)
Patient problems: Bowel Perforation; Pleural Effusion; Unspecified Infection
MAUDE report 22127507 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER = 2 CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER < 2 CM (2.6%) (P < 0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%)
Patient problems: Bowel Perforation; Hemorrhage/Blood Loss/Bleeding; Pleural Effusion
MAUDE report 22127400 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION FOR DUODENAL SUBEPITHELIAL LESIONS." LITERATURE SUMMARY BACKGROUND AND AIM: THIS WORK AIMS TO EVALUATE THE EFFICACY AND SAFETY OF FREE-HAND ENDOSCOPIC FULL-THICKNESS RESECTION (EFTR) FOR DUODENAL SUBEPITHELIAL LESIONS (SELS). METHODS: WE PERFORMED A RETROSPECTIVE REVIEW OF 105 PATIENTS WITH DUODENAL SELS WHO UNDERWENT FREE-HAND EFTR. FREE-HAND EFTR MEANS NO OTHER DEVICES (OVER-THE-SCOPE CLIP OR FULL-THICKNESS RESECTION DEVICE) ARE REQUIRED. THE PREOPERATIVE BASELINE DATA, PROCEDURE-RELATED CHARACTERISTICS, AND POSTOPERATIVE OUTCOMES WERE ANALYZED. RESULTS: THE TECHNICAL SUCCESS RATE WAS 99.0%, AND THE EN BLOC RESECTION RATE WAS 94.2%. A TOTAL OF NINE (8.7%) PATIENTS EXPERIENCED MAJOR POSTOPERATIVE ADVERSE EVENTS (AES). THE INCIDENCE OF MAJOR AES WAS SIGNIFICANTLY HIGHER FOR LESIONS WITH A MAXIMUM DIAMETER =2CM (30.4%) THAN FOR LESIONS WITH A MAXIMUM DIAMETER <2CM (2.6%) (P<0.001). THERE WERE ALSO SIGNIFICANT DIFFERENCES IN THE INCIDENCE OF MAJOR AES FOR PERI-AMPULLARY LESIONS (37.5%), BULB LESIONS (4.8%), BULB-DESCENDING JUNCTION LESIONS (6.7%), AND DESCENDING PART LESIONS (12.5%) (P=0.0
Patient problems: Hemorrhage/Blood Loss/Bleeding; Swelling/ Edema
MAUDE report 22127252 ↗InjuryMay 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.
Patient problems: Airway Obstruction; Hemorrhage/Blood Loss/Bleeding; Loss of consciousness; Pain; Pleural Effusion; Pneumoperitoneum; Pneumothorax; Pulmonary Emphysema; Unspecified Infection
MAUDE report 22013756 ↗InjuryDec 11, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: Foreign Body In Patient
MAUDE report 20909724 ↗InjuryDec 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE CERAMIC TIP OF THE KNIFE FELL OFF AND INTO THE PATIENT¿S STOMACH DURING A THERAPEUTIC ENDOSCOPIC SUBMUCOSAL DISSECTION PROCEDURE. THE PROCEDURE WAS DELAYED FOR 15 MINUTES TO RETRIEVE THE DEVICE. ALTHOUGH REQUESTED, IT IS UNKNOWN HOW THE DEVICE WAS RETRIEVED, AND IF ANY ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20859655 ↗InjuryNov 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AFTER AN UNSPECIFIED SURGERY, THE TIP OF THE KNIFE WIRE FELL OFF INTO THE PATIENT'S BODY AND WAS RETRIEVED. THE CASE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO REPORTS OF ANY PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20745778 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page