Single Use Electrosurgical KnifeRecalls & reported problems across all models

Olympus Medical Systems Corp. · 9 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one Single Use Electrosurgical Knife model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
509
Deaths reported
4
Injuries reported
295
Recalls
2

We found 9 models of Single Use Electrosurgical Knife by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 509 adverse event reports in the FDA's MAUDE database, received between Jun 21, 2018 and Aug 7, 2025 — classified as 4 deaths, 295 injuries, 210 malfunctions.

The most frequently reported specific device problem is “Break” (255 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 9 reportsDec 2023: 2 reportsJan 2024: 21 reportsFeb 2024: 89 reportsMar 2024: 54 reportsApr 2024: 37 reportsMay 2024: 15 reportsJun 2024: 11 reportsJul 2024: 19 reportsAug 2024: 16 reportsSep 2024: 6 reportsOct 2024: 8 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 9 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202689
Last 36 months. The most recent month may be incomplete.

Single Use Electrosurgical Knife recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 1, 2024Open, Classified

    Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury

    Z-1547-2024 · Root cause: Use error · Recall details · FDA record ↗

  2. Feb 17, 2024Open, Classified

    A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.

    Z-1222-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported Single Use Electrosurgical Knife problems

Device problems

  • Break255
  • Adverse Event Without Identified Device or Use Problem185
  • Detachment of Device or Device Component130
  • Insufficient Device Problem Information7
  • Crack5
  • Material Split, Cut or Torn3
  • Melted3
  • Material Fragmentation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions293
  • Hemorrhage/Blood Loss/Bleeding90
  • Perforation88
  • Foreign Body In Patient20
  • Fever14
  • No Consequences Or Impact To Patient12
  • Unspecified Infection10
  • Pleural Effusion9

Single Use Electrosurgical Knife models

9 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Single Use Electrosurgical Knife adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Olympus Medical Systems Corp. products

All Olympus Medical Systems Corp. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page