Single Use Mechanical Lithotriptor VRecalls & reported problems across all models
Olympus Medical Systems Corp. · 6 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one Single Use Mechanical Lithotriptor V model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 359
- Deaths reported
- 3
- Injuries reported
- 83
- Recalls
- 1
We found 6 models of Single Use Mechanical Lithotriptor V by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 359 adverse event reports in the FDA's MAUDE database, received between Jul 25, 2018 and Aug 13, 2026 — classified as 3 deaths, 83 injuries, 273 malfunctions.
The most frequently reported specific device problem is “Break” (192 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Single Use Mechanical Lithotriptor V recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Dec 16, 2024Open, Classified
Potential for distal tip tearing of the Mechanical Lithotriptor V.
Z-0962-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported Single Use Mechanical Lithotriptor V problems
Device problems
- Break192
- Material Split, Cut or Torn80
- Detachment of Device or Device Component29
- Difficult to Remove14
- Adverse Event Without Identified Device or Use Problem11
- Mechanical Problem11
- Crack9
- Material Separation7
Patient problems
- No Clinical Signs, Symptoms or Conditions278
- No Consequences Or Impact To Patient31
- Foreign Body In Patient16
- Device Embedded In Tissue or Plaque8
- Pancreatitis8
- No Code Available7
- Perforation7
- Hemorrhage/Blood Loss/Bleeding5
Single Use Mechanical Lithotriptor V models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Single Use Mechanical Lithotriptor V BML-V442QR-30Lithotriptor, biliary mechanical1 recall1 death269 reports
- Single Use Mechanical Lithotriptor V BML-V437QR-30Lithotriptor, biliary mechanical33 reports
- Single Use Mechanical Lithotriptor V BML-V242QR-30Lithotriptor, biliary mechanical21 reports
- Single Use Mechanical Lithotriptor V BML-V437QR-30Lithotriptor, Biliary Mechanical21 reports
- Single Use Mechanical Lithotriptor V BML-V232QR-30Lithotriptor, biliary mechanical2 deaths12 reports
- Single Use Mechanical Lithotriptor V BML-V232QR-26Lithotriptor, biliary mechanical3 reports
Recent Single Use Mechanical Lithotriptor V adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50235853 ↗InjuryJul 30, 2026Break
It was reported that during a therapeutic Endoscopic Retrograde Cholangiopancreatography while removing a stone, the basket cable broke in the proximal area. The doctor used the emergency Lithotriptor handle an attempt to remove the single use mechanical Lithotriptor and the cable broke again, this time in the distal area, next to the basket, leaving the stone and the basket inside the patient. As such the procedure was delayed for equal to or greater then 30 minutes under sedation, completed with a competitors device. No reports of patient harm.
Patient problems: Foreign Body In Patient
MAUDE report 50090758 ↗InjuryJun 26, 2026Break
IT WAS REPORTED THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE, THE SINGLE USE MECHANICAL LITHOTRIPTOR V, WHICH WAS BEING USED TO CRUSH STONES, BROKE WHILE INSIDE THE PATIENT. WHEN UTILIZING THE EMERGENCY HANDLE, THE FOUR WIRES INSIDE THE DEVICE ALSO BROKE. THE PATIENT WAS SENT FOR SURGERY. SURGERY WAS SUCCESSFUL. THE PATIENT REMAINED HOSPITALIZED. THERE WERE NO FURTHER REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25638966 ↗InjuryMay 15, 2026Failure to Cut; Mechanical Jam
IT WAS REPORTED THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THIS SINGLE USE MECHANICAL LITHOTRIPTOR BASKET FAILED TO CRUSH A STONE. THE PHYSICIAN THEN USED THE EMERGENCY HANDLE. THE STONE WAS CRUSHED AND THE BASKET WAS RETRIEVED FROM THE PATIENT. THERE WERE NO FURTHER REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25191988 ↗InjuryMar 5, 2026Break; Difficult to Remove; Mechanical Jam
IT WAS REPORTED THAT DURING AN ELECTIVE AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE UNDER SEDATION WITH PROPOFOL, THE LOCKING ROD OF THE SINGLE USE MECHANICAL LITHOTRIPTOR BROKE DIRECTLY AT THE END OF THE BML HANDLE, AND REST REMAINED STUCK IN THE HANDLE DURING MECHANICAL LITHOTRIPSY. THE BASKET BROKE OFF FROM THE DEVICE AND REMAINED IN THE PATIENT, WITH THE WIRES STILL ATTACHED AND ACCESSIBLE IN CONTACT WITH THE BASKET. THE BASKET ITSELF HAD ENCLOSED A LARGE INTRADUCTAL CALCULUS AND COULD NOT INITIALLY BE REMOVED FROM THE DHC. AN EMERGENCY LITHOTRIPSY SET FROM A THIRD-PARTY COMPANY WAS THEN USED. THE WIRES WERE INSERTED INTO THIS DEVICE AND THE STONE WAS LITHOTRIPSIED, AND THE DEFECTIVE BASKET WAS SUCCESSFULLY RETRIEVED FROM THE PATIENT BY PULLING IT OUT BY WIRES OF THE OTHER DEVICE BUT THE CONCRETIONS REMAINED INSIDE THE BILE DUCT. THE PATIENT THEN UNDERWENT FURTHER TREATMENT, INCLUDING STENT PLACEMENT AND PLAN A CHOLANGIOSCOPY WITH EHL. THE PROCEDURE WAS COMPLETED WITH A SIGNIFICANT DELAY OF APPROXIMATELY 40 MINUTES THE PATIENT REMAINED HEMODYNAMICALLY STABLE. THERE WERE NO REPORTS OF PATIENT HARM OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24525833 ↗InjuryFeb 18, 2026Break; Detachment of Device or Device Component
NO RELEVANT ADDITIONAL INFORMATION SUBMITTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378347 ↗MalfunctionDec 26, 2025Break; Break
IT WAS REPORTED THE SINGLE USE MECHANICAL LITHOTRIPTOR HAD BASKET BREAK. THE ISSUE OCCURRED DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23897792 ↗MalfunctionSep 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE LITHOCRUSH WIRES DETACHED FROM THE SHEATH OF THE SUBJECT DEVICE, CAUSING A SLIGHT DELAY. THE BASKET AND STONES WERE SUCCESSFULLY RETRIEVED WITH THE SAME DEVICE. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23071955 ↗MalfunctionAug 15, 2025Break; Break
IT WAS REPORTED THE SUBJECT DEVICE THE TAB THROUGH WHICH THE GUIDE IS INSERTED WAS BROKEN AND IT CAME OUT. THIS ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22818429 ↗InjuryAug 6, 2025Difficult to Remove; Physical Resistance/Sticking; Difficult to Remove; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE METAL COIL SHEATH WAS PUSHED ON THE PLASTIC SHEATH OF THE SUBJECT DEVICE TO CRUSH THE STONE BUT BECAME STUCK AND COULD NOT PUSH THE COIL SHEATH ANYMORE. THE PLASTIC TUBE BECAME COMPRESSED, AND THE SHAPE CHANGED. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22728547 ↗MalfunctionJul 3, 2025Break; Break
IT WAS REPORTED THE SINGLE USE MECHANICAL LITHOTRIPTOR V HAD A FAULT OCCUR WHILST INSIDE PATIENT. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22395839 ↗MalfunctionJun 26, 2025Difficult or Delayed Positioning; Difficult or Delayed Positioning
IT WAS REPORTED THAT DURING THE ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ECPW) PROCEDURE, THE STONE HAD ALREADY BEEN CAUGHT AND WAS ABOUT TO BE CRUSHED, THE KNOB FOR CRUSHING/CLOSING THE SUBJECT DEVICE BROKE OFF WITH THE TOOL/HANDLE STUCK. THE PROCEDURE WAS COMPLETED BY REPLACING IT WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22338685 ↗MalfunctionMay 29, 2025Break; Break
IT WAS REPORTED THE SUBJECT DEVICE BASKET¿S STEM SNAPPED WHILE ATTEMPTING TO CRUSH A BILIARY STONE. THIS ISSUE OCCURRED DURING A UNKNOWN THERAPEUTIC PROCEDURE. THE EMERGENCY HANDLE WAS THEN USED TO COMPLETE THE PROCEDURE. WHILE USING THE EMERGENCY HANDLE TO CRUSH THE STONE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22109898 ↗MalfunctionMay 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED DISTAL TIP FOR PASSING GUIDEWIRE WAS TORN AND BROKEN GUIDEWIRE TIP. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22104880 ↗MalfunctionMay 28, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE SILICONE RAIL INTO WHICH THE GUIDE WIRE IS INSERTED INTO ON THE LITHOTRIPTOR WAS TORN. THE REPORTED ISSUE OCCURRED DURING SET UP AND THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22098646 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.