Single Use Repositionable ClipRecalls & reported problems across all models
Olympus Medical Systems Corp. · 7 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
4 open recalls name at least one Single Use Repositionable Clip model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 248
- Deaths reported
- 0
- Injuries reported
- 79
- Recalls
- 4
We found 7 models of Single Use Repositionable Clip by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 248 adverse event reports in the FDA's MAUDE database, received between Jul 11, 2018 and Feb 13, 2026 — classified as 79 injuries, 169 malfunctions.
The most frequently reported specific device problem is “Break” (37 reports).
4 FDA recall records name at least one of these models by its device identifier, 4 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Single Use Repositionable Clip recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 4, 2023Open, Classified
Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.
Z-1990-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- May 4, 2023Open, Classified
Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.
Z-1989-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- May 4, 2023Open, Classified
Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.
Z-1988-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- May 4, 2023Open, Classified
Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.
Z-1987-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported Single Use Repositionable Clip problems
Device problems
- Break37
- Detachment of Device or Device Component32
- Device Fell28
- Difficult to Open or Close19
- Failure to Fire19
- Separation Failure17
- Difficult to Remove11
- Failure to Eject11
Patient problems
- No Clinical Signs, Symptoms or Conditions86
- No Known Impact Or Consequence To Patient53
- No Consequences Or Impact To Patient44
- Hemorrhage/Blood Loss/Bleeding28
- Gastrointestinal Hemorrhage10
- Foreign Body In Patient9
- Device Embedded In Tissue or Plaque5
- No Code Available5
Single Use Repositionable Clip models
7 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Single Use Repositionable Clip HX-202URHemostatic metal clip for the GI tract1 recall118 reports
- Single Use Repositionable Clip HX-202URHemostatic metal clip for the GI tract87 reports
- Single Use Repositionable Clip HX-202URHemostatic metal clip for the GI tract1 recall23 reports
- Single Use Repositionable Clip HX-202LRHemostatic metal clip for the GI tract1 recall2 reports
- Single Use Repositionable Clip HX-202LRHemostatic metal clip for the GI tract1 recall0 reports
- Single Use Repositionable Clip HX-202URHemostatic metal clip for the GI tract15 reports
- Single Use Repositionable Clip HX-202LRHemostatic metal clip for the GI tract3 reports
Recent Single Use Repositionable Clip adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2026Difficult to Open or Close
IT WAS REPORTED THE SINGLE USE REPOSITIONAL CLIP GOT STUCK AND THERE WAS NO WAY TO RELEASE IT. THE ISSUE OCCURRED DURING A THERAPEUTIC GASTROSCOPY AND WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24333120 ↗MalfunctionAug 27, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: Insufficient Information
MAUDE report 22902766 ↗MalfunctionMay 22, 2025Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED DURING A DIAGNOSTIC COLONOSCOPY PROCEDURE, THE SUBJECT DEVICE FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE AND THERE WERE NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22074917 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21107358 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM. THE FOLLOWING PATIENT IDENTIFIERS ARE LINKED : (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21107345 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21107272 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM. THE FOLLOWING PATIENT IDENTIFIERS ARE LINKED: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21107262 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21107245 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM. THE FOLLOWING PATIENT IDENTIFIERS ARE LINKED: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21107213 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106906 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM. THE FOLLOWING PATIENT IDENTIFIERS ARE LINKED : (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106697 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106544 ↗InjuryJan 8, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN THERAPEUTIC PROCEDURE, 5 UNITS OF THE SUBJECT DEVICE HAD NEVER DEPLOYED. THIS RESULTED TO THE DELAY OF THE PROCEDURE FOR MORE THAN 30 MINUTES. THE PROCEDURE WAS THEN COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM. THE FOLLOWING PATIENT IDENTIFIERS ARE LINKED: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106535 ↗InjurySep 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE ONCE THE CLIP DELIVERY SYSTEM WENT THROUGH THE SCOPE AND INTO THE COLON, THE CLIP JUST FELL OFF THE DELIVERY SYSTEM WITHOUT DEPLOYING THE CLIP. THE CUSTOMER LATER REPORTED THAT THE CLIP FELL INTO THE PATIENT AND WAS RECOVERED WITHOUT ANY INJURY OR HARM. THE PROCEDURE BEING PERFORMED WAS A COLONOSCOPY, WHICH WAS COMPLETED SUCCESSFULLY WITH A COMPETITOR PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20252788 ↗MalfunctionSep 11, 2024Difficult to Remove; Difficult to Remove
IT WAS REPORTED, THE SINGLE USE REPOSITIONABLE CLIP TOOK FORCE TO PLACE THE ENDOCLIP. BUT THEN, THE ENDOCLIP COULD NOT BE DETACHED FROM THE BASE. THE ISSUE OCCURRED, DURING A GASTROSCOPY PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20187160 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page