SonoSurg Curved Scissors,HF,Pistol Grip,5mm x 34cmRecalls & reported problems across all models
Olympus Medical Systems Corp. · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 57
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls
- 0
We found 2 models of SonoSurg Curved Scissors,HF,Pistol Grip,5mm x 34cm by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 57 adverse event reports in the FDA's MAUDE database, received between Oct 5, 2016 and Sep 27, 2024 — classified as 9 injuries, 48 malfunctions.
The most frequently reported specific device problem is “Break” (26 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported SonoSurg Curved Scissors,HF,Pistol Grip,5mm x 34cm problems
Device problems
- Break26
- Peeled/Delaminated9
- Adverse Event Without Identified Device or Use Problem6
- Fracture6
- Detachment of Device or Device Component5
- Material Separation4
- Disconnection3
- Melted2
Patient problems
- No Clinical Signs, Symptoms or Conditions42
- No Consequences Or Impact To Patient8
- Hemorrhage/Blood Loss/Bleeding3
- Fistula2
- Pneumonia2
- Ascites1
- Cyst(s)1
- Fever1
SonoSurg Curved Scissors,HF,Pistol Grip,5mm x 34cm models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- SonoSurg Curved Scissors,HF,Pistol Grip,5mm x 34cm T3905Instrument, ultrasonic surgical54 reports
- SonoSurg Curved Scissors,HF,Pistol Grip,5mm x 34cm T3905Instrument, ultrasonic surgical3 reports
Recent SonoSurg Curved Scissors,HF,Pistol Grip,5mm x 34cm adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 27, 2024Break; Break
IT WAS REPORTED, THE CURVED SCISSORS EXHIBITED A BROKEN OFF PROBE (NO FALL OUT IN THE BODY). THE ISSUE OCCURRED DURING A PROCEDURE FOR AN UNSPECIFIED PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF DELAY, PATIENT HARM, INJURY, OR DEATH. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20328148 ↗InjuryAug 27, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT THE SUBJECT DEVICE FELL INTO THE ABDOMINAL CAVITY AND THE SIZE OF THE PIECE IS 1CM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20078576 ↗MalfunctionJun 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SONOSURG SCISSORS EXHIBITED A BROKEN PROBE 15.5MM FROM THE TIP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19620738 ↗MalfunctionJun 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE CURVED SCISSORS EXHIBITED PAD DETACHMENT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19582208 ↗MalfunctionJun 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PROBE DETACHED FROM THE PISTOL GRIP DURING A THERAPEUTIC PROCEDURE. THE DEVICE WAS REPLACED WITH A SIMILAR ONE, ALLOWING THE PROCEDURE TO BE SUCCESSFULLY COMPLETED WITHOUT ANY REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19571081 ↗MalfunctionMay 24, 2024Break; Break
IT WAS REPORTED DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE SCISSORS WAS REMOVED AND WIPED OFF WITH GAUZE BEFORE THE PROCEDURE, THE PROBE BROKE OFF OUTSIDE THE BODY. THERE WERE NO REPORTS OF IT FALLING OFF INSIDE THE BODY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19397443 ↗MalfunctionApr 15, 2024Pitted; Pitted
IT WAS REPORTED, THE SONOSURG CURVED SCISSORS,HF,PISTOL GRIP,5MM X 34CM EXHIBITED THAT THE TIP OF THE PROBE WAS CHIPPED. THE ISSUE OCCURRED DURING A THERAPEUTIC LAPAROSCOPIC INGUINAL HERNIA REPAIR PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19112199 ↗InjuryFeb 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) RECEIVED A LITERATURE TITLED "MO36-1 A CASE OF SINGLE HOLE LAPAROSCOPIC WINDOW OPENING FOR GIANT HEPATIC CYST WITH INTRAOPERATIVE PREVENTION OF BILE LEAK BY COMBINED USE OF ICG FLUORESCENCE." ¿LITERATURE SUMMARY¿ THE PATIENT IS AN 84-YEAR-OLD MAN. HE WAS REFERRED TO OUR DEPARTMENT AFTER AN ABDOMINAL CT SCAN REVEALED AN 18.7 CM DIAMETER LIVER CYST IN THE S2 OF THE LIVER. ON PHYSICAL EXAMINATION AT THE TIME OF ADMISSION, A HUGE ABDOMINAL BULGE WITH MOBILITY WAS SEEN IN THE LEFT UPPER ABDOMEN. THE DIAGNOSIS OF A SYMPTOMATIC GIANT SIMPLE HEPATIC CYST WAS MADE, AND A SINGLE HOLE LAPAROSCOPIC WINDOW OPENING WAS PERFORMED. THE LAPAROSCOPE WAS AN OLYMPUS VISERA ELITE II WITH IR MODE AFTER INTRAVENOUS ICG. FIRST, THE LIVER CYST WAS PUNCTURED WITH A SAND BALLOON, AND AFTER ASPIRATING THE FLUID (1900 ML), THE CYST WALL WITH POOR FLUORESCENCE WAS DISSECTED WITH SONOSURG WHILE USING IR MODE. AFTER CAUTERIZING THE CYST WALL, THE CYST WALL WAS OBSERVED AGAIN WITH IR MODE, AND A SMALL AMOUNT OF ICG WAS FOUND LEAKING FROM THE DISSECTED EDGE OF THE CYST WALL. WE CONSIDERED IT TO BE A BILE SCAR AND CLIPPED THE SAME AREA, AFTER WHICH THE SURGERY WAS COMPLETED.
Patient problems: Fistula; Pneumonia
MAUDE report 18646226 ↗MalfunctionJan 22, 2024Melted; Excessive Heating; Melted; Excessive Heating
THE CUSTOMER REPORTED TO OLYMPUS THAT WHEN USING THE SONOSURG CURVED SCISSORS FOR THE FIRST TIME, THE TIP OF THE PAD BURNED AND MELTED AFTER SEVERAL OUTPUTS. NO INTERNAL SHEDDING OCCURRED. THE THERAPEUTIC PROCEDURE WAS COMPLETED WHILE USING A REPLACEMENT DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18553579 ↗MalfunctionJan 10, 2024Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT WHEN USING THE SONOSURG CURVED SCISSORS, HF, PISTO, THE PROBE BROKE AND FELL OUTSIDE OF THE PATIENT'S BODY CAVITY. THE BROKEN PARTS WERE COLLECTED. THERE WERE NO DELAY WITH THE PROCEDURE. THE ISSUE OCCURRED DURING PROCEDURE, AND THE THERAPEUTIC LAPAROSCOPIC CHOLECYSTECTOMY WAS COMPLETED BUT WAS UNSPECIFIED IF IT WAS COMPLETED WITH THE SAME SET OF EQUIPMENT OR SIMILAR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18483972 ↗MalfunctionDec 20, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THAT THE TIP OF THE PROBE BROKE WHILE USING THE SONOSURG CURVED SCISSORS. SHORTLY AFTER THE POWER WAS TURNED ON, THE CUSTOMER STATED THAT THERE WERE NO FORCE BEING APPLIED TO THE TIP. NO BURNT MARKS OR INTERNAL SHEDDING WERE FOUND. THE ISSUE OCCURRED DURING THE THERAPEUTIC PROCEDURE, AND IT WAS COMPLETED BY SWITCHING TO A SIMILAR DEVICE, WITH A DELAY IN WHICH THE TIMEFRAME WAS NOT SPECIFIED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18367115 ↗MalfunctionDec 4, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE SONOSURG CURVED SCISSORS EXPERIENCED AN OUTPUT PROBLEM, AND THE TISSUE PAD WAS WORN OUT. THE ISSUE WAS FOUND DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18254189 ↗MalfunctionNov 28, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THAT THE ULTRASONIC SURGICAL DEVICE PROBE TIP BROKE. THE ISSUE WAS FOUND DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18219905 ↗MalfunctionNov 17, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT THE SONOSURG CURVED SCISSORS,HF,PISTOL GRIP,5MM X 34CM HAD THE WHITE PAD ON THE SIDE PARTIALLY PEEL OFF. THE ISSUE WAS FOUND DURING THE THERAPEUTIC LAPAROSCOPIC APPENDECTOMY PROCEDURE WHICH WAS COMPLETED WITH A SIMILAR DEVICE. THE DEVICE WAS RETURNED FOR EVALUATION AND DURING THE EVALUATION THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: ONE OF THE PINS THAT SERVES AS THE FULCRUM OF THE GRIP PART IS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18160300 ↗MalfunctionOct 31, 2023Peeled/Delaminated; Peeled/Delaminated
THE ULTRASONIC SURGICAL DEVICE WAS SENT TO AN OLYMPUS SERVICE CENTER FOR EVALUATION WITH A REPORT THAT THERE WAS BLACK RUBBER MATERIAL THAT CAME OUT FROM INSIDE THE HANDLE WHILE CLEANING. THE CUSTOMER ALSO NOTED THAT THEY WERE NOT ABLE TO COLLECT THE RUBBER PARTS, SO THEY WOULD LIKE TO CONFIRM THAT THERE ARE NO MISSING PARTS INSIDE. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT. DURING INSPECTION, IT WAS FOUND THAT SOME OF THE INTERNAL SILICONE RUBBER HAD PEELED OFF, AND THE GRIPPING SURFACE HAD PEELED OFF. THE MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18039005 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page