Spine & Trauma 3D NavigationRecalls & reported problems across all models
Brainlab SE · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 131
- Deaths reported
- 0
- Injuries reported
- 131
- Recalls
- 1
We found 2 models of Spine & Trauma 3D Navigation by Brainlab SE in the FDA's device registry (GUDID). Together they are named in 131 adverse event reports in the FDA's MAUDE database, received between Aug 21, 2019 and Sep 4, 2026 — classified as 131 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (129 reports).
1 FDA recall record name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Spine & Trauma 3D Navigation recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Feb 27, 2019Terminated
In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.
Z-1082-2019 · Root cause: Software design · Recall details · FDA record ↗
Most reported Spine & Trauma 3D Navigation problems
Device problems
- Use of Device Problem129
- Improper or Incorrect Procedure or Method32
- Insufficient Device Problem Information13
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions116
- No Consequences Or Impact To Patient6
- Insufficient Information4
- Paresis4
- Cerebrospinal Fluid Leakage1
- Headache1
- Hypoesthesia1
- Movement Disorder1
Spine & Trauma 3D Navigation models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Spine & Trauma 3D Navigation 1,5Orthopedic Stereotaxic Instrument127 reports
- Spine & Trauma 3D Navigation 1.0Neurological Stereotaxic Instrument1 recall4 reports
Recent Spine & Trauma 3D Navigation adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 4, 2026Use of Device Problem; Insufficient Device Problem Information
An open surgery at the lumbar spine for a transforaminal lumbar interbody fusion(TLIF) at L4/L5 with decompression/laminectomy due to grade I spondylolisthesis, with the intended placements of 4 bilateral k-wires, was performed with Brainlab Spine & Trauma 3D Navigation 1.5.Following successful placement of the K-wires at left L4 and L5, a navigation inaccuracy was observed during drilling at right L4 after the drill entered soft tissue. The affected trajectory was not used for final K-wire placement, and no instrumentation was placed based on the inaccurate navigation guidance. Navigation was discontinued, and the remainder of the procedure was completed conventionally.According to the surgeon (treating clinician):Procedure continued conventionally without navigation and the outcome of the surgery was successful as intended, with all placements correct at the end of the surgery.There was no direct (or increased) risk of harm to a critical structure due to the reported problem/apparent navigation inaccuracy at this surgery,There was no actual harm/negative clinical effect to the patient resulting due to the initial trajectory deviation at right L4, also not due to the prolongation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50614536 ↗InjuryJul 17, 2026Use of Device Problem; Insufficient Device Problem Information
An open surgery on the lumbar spine for a fusion of vertebrae L2-L3 and L3-L4, due to degeneration of L2-L3 and L3-L4, with intended placement of 6 screws bilateral for fixation (at L2, L3 and L4), was performed with the aid of the display by the Brainlab Navigation software Spine & Trauma 3D Navigation 1.5.From an intra-operative C-arm scan, the surgeon discovered that of 3 of the 6 screws placed with Brainlab navigation involved deviated from their intended positions (at left L2, L3 and L4, all 10mm too cranial). The k-wires and screws were corrected to their intended positions with navigation at the very same surgery.According to the hospital (treating clinician): Non-removal of the screws placed in a different location than intended with Brainlab Navigation involved would potentially have led to a resulting critical/serious harm to the patient (for e.g. permanent damage/impairment), because of affecting the nerves at L2 and L3. The final outcome of the surgery was successful as intended, with all screw placements correct at the end of the surgery.Although there was a direct risk of critical harm to the patient (for e.g. permanent damage/impairment), because of the deviating pla
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25851324 ↗InjuryOct 15, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
A MINIMALLY INVASIVE SURGERY ON THE LUMBAR AND SACRAL SPINE, FOR A PERCUTANEOUS TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) OF VERTEBRAE L5 TO S1, WITH INTENDED PLACEMENT OF 4 SPINE SCREWS BILATERAL FOR FIXATION, WAS PERFORMED WITH THE AID OF THE BRAINLAB SPINE & TRAUMA 3D NAVIGATION 1.5. AFTER PLACING 3 OF THE 4 SCREWS - BILATERAL IN VERTEBRA L5 AND IN VERTEBRA RIGHT S1 - FROM INTRA-OPERATIVE FLUORO (X-RAY) IMAGING, THE SURGEON DETERMINED THAT THE SPINE SCREWS PLACED DEVIATED FROM THEIR INTENDED POSITIONS SHIFTED BY CA. 10-15MM CRANIALLY. THE SURGEON DECIDED TO REMOVE AND TO RE-PLACE THE DEVIATING SCREWS, AS WELL AS PLACING THE REMAINING INTENDED SCREW, WITH CONVENTIONAL METHODS UNDER FLUOROSCOPIC GUIDANCE TO THEIR CORRECT POSITIONS AT THE VERY SAME SURGERY. ACCORDING TO THE SURGEON (TREATING CLINICIAN): THE OUTCOME OF THE SURGERY WAS SUCCESSFUL AS INTENDED, WITH ALL PLACEMENTS CORRECT AT THE END OF THE SURGERY. THERE WAS NO DIRECT (OR INCREASED) RISK OF HARM TO A CRITICAL STRUCTURE DUE TO THE DEVIATING INITIAL PLACEMENTS. THERE WAS NO HARM NOR NEGATIVE EFFECT TO THE PATIENT, ALSO NOT DUE TO THE SURGERY/ANESTHESIA PROLONGATION OF CA. 1-1.5HRS. THERE WERE FURTHER NO REMEDIAL ACTI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23298573 ↗InjuryMay 27, 2025Use of Device Problem; Use of Device Problem
AN OPEN SURGERY ON THE LUMBAR SPINE FOR A TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) OF VERTEBRA L3 TO A FORMER EXISTING FUSION OF VERTEBRAE L4 AND L5, DUE TO DEGENERATIVE CHANGES IN THE SEGMENT L3-L5, WITH INTENDED PLACEMENT OF 2 VERTEBRA SCREWS BILATERAL FOR FIXATION, WAS PERFORMED WITH THE AID OF THE BRAINLAB SPINE & TRAUMA 3D NAVIGATION 1.5. FROM AN INTRA-OPERATIVE C-ARM SCAN, THE SURGEON DETERMINED THAT OF 1 OF THE 2 SPINE SCREWS PLACED, IN VERTEBRA LEFT L3, DEVIATED FROM ITS INTENDED TARGET POINT BY CA. 10MM. THE DEVIATING SCREW WAS REMOVED AND RE-PLACED TO ITS CORRECT POSITION WITH NAVIGATION AT THE VERY SAME SURGERY. ACCORDING TO THE SURGEON (TREATING CLINICIAN): THE FINAL OUTCOME OF THE SURGERY WAS SUCCESSFUL AS INTENDED. THERE WAS NO DIRECT (OR INCREASED) RISK TO HARM A CRITICAL ANATOMICAL STRUCTURE DUE TO THE DEVIATING PLACEMENT. THERE WAS NO HARM NOR NEGATIVE EFFECT TO THE PATIENT DUE TO THE DEVIATING INITIAL PLACEMENT, ALSO NOT DUE TO SURGERY/ANESTHESIA PROLONGATION OF CA. 20MIN. THERE WERE FURTHER NO REMEDIAL ACTIONS FOR THE PATIENT DONE, NECESSARY OR PLANNED. HOSPITALIZATION WAS ALSO NOT PROLONGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22087498 ↗InjuryMay 14, 2025Use of Device Problem; Use of Device Problem
AN OPEN SURGERY ON THE LUMBAR SPINE FOR A FUSION OF VERTEBRAE L3 TO L5, DUE TO FORAMINAL STENOSIS, WITH INTENDED PLACEMENT OF 6 VERTEBRA SCREWS BILATERAL FOR FIXATION, WAS PERFORMED WITH THE AID OF THE BRAINLAB SPINE & TRAUMA 3D NAVIGATION 1.5. FROM INTRA-OPERATIVE FLUOROSCOPY IMAGES, THE SURGEON DETERMINED THAT OF 1 OF THE 6 SPINE SCREWS PLACED WITH THE AID OF NAVIGATION, IN VERTEBRA LEFT L5, DEVIATED SHIFTED FROM ITS INTENDED POSITION. THE DEVIATING SCREW WAS REMOVED AND RE-PLACED TO ITS CORRECT POSITION, REQUIRING SEVERAL PILOT HOLE CREATION ATTEMPTS, WITH NAVIGATION AT THE VERY SAME SURGERY. ACCORDING TO THE SURGEON (TREATING CLINICIAN): THE FINAL OUTCOME OF THE SURGERY WAS SUCCESSFUL AS INTENDED. THERE WAS NO DIRECT (OR INCREASED) RISK TO HARM A CRITICAL ANATOMICAL STRUCTURE DUE TO THE DEVIATING PLACEMENTS. THERE WAS NO HARM NOR NEGATIVE EFFECT TO THE PATIENT DUE TO THE DEVIATING INITIAL PLACEMENTS, ALSO NOT DUE TO SURGERY/ANESTHESIA PROLONG OF CA. 45MIN. THERE WERE FURTHER NO REMEDIAL ACTIONS FOR THE PATIENT DONE, NECESSARY OR PLANNED. HOSPITALIZATION WAS ALSO NOT PROLONGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22018552 ↗InjuryDec 18, 2024Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
AN OPEN SURGERY ON THE LUMBAR SPINE FOR A POSTERIOR FUSION OF VERTEBRAE L4 TO L5, DUE TO SPONDYLOLISTHESIS, WITH INTENDED PLACEMENT OF 4 VERTEBRA SCREWS BILATERAL FOR FIXATION (AT L4-L5), WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB NAVIGATION SOFTWARE SPINE & TRAUMA 3D NAVIGATION 1.5.1. FROM INTRA-OPERATIVE FLUOROSCOPY IMAGES, THE SURGEON DISCOVERED THAT OF 1 OF THE 4 SCREWS PLACED WITH THE AID OF NAVIGATION DEVIATED FROM ITS INTENDED POSITION (AT LEFT L4, BY CA. 5MM LATERALLY FROM THE INTENDED ENTRY AND TARGET POSITION). THE DEVIATING SCREW WAS CORRECTED DURING THE SAME SURGERY WITH THE AID OF NAVIGATION. ACCORDING TO THE HOSPITAL: THE DEVIATION OF 1 OF THE 4 PEDICLE SCREWS PLACED WITH THE AID OF NAVIGATION WAS DETECTED BY THE SURGEON BEFORE FINALIZING THE SURGERY, AND THE SCREW WAS CORRECTED TO ITS INTENDED POSITION WITH NAVIGATION AT THE VERY SAME SURGERY. THE FINAL OUTCOME OF THE SURGERY WAS SUCCESSFUL AS INTENDED, WITH ALL SCREW PLACEMENTS CORRECT AT THE END OF THE SURGERY. THERE WAS NO DIRECT (OR INCREASED) RISK TO HARM A CRITICAL STRUCTURE REPORTED DUE TO THE DEVIATING PLACEMENTS. THERE WAS NO HARM NOR NEGATIVE EFFECT TO THE PATIENT DUE TO THE DEVIATING INITIAL
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20964445 ↗InjuryDec 13, 2024Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
A MINIMALLY INVASIVE SURGERY ON THE LUMBAR SPINE FOR A TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) OF VERTEBRAE L4 AND L5, WITH INTENDED PLACEMENT OF 4 VERTEBRA SCREWS BILATERAL FOR FIXATION, WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB SPINE & TRAUMA 3D NAVIGATION 1.5. FROM AN INTRA-OPERATIVE VERIFICATION O-ARM SCAN, THE SURGEON DETERMINED THAT 2 SPINE SCREWS PLACED IN RIGHT L4 AND L5 WITH NAVIGATION, DEVIATED FROM THEIR INTENDED POSITIONS SHIFTED CAUDAL BY CA. 17-20MM, AND INTO THE DISC SPACES. ACCORDING TO THE SURGEON (TREATING CLINICIAN): THE DEVIATION OF 2 OF 4 SPINE SCREWS PLACED WITH THE AID OF NAVIGATION WAS DETECTED BY THE SURGEON WITH A VERIFICATION O-ARM SCAN BEFORE FINALIZING THE SURGERY, AND THE SCREW PLACEMENTS WERE CORRECTED TO THEIR INTENDED POSITIONS AFTER SWITCHING TO AN OPEN SURGERY APPROACH, WITH NAVIGATION AT THE VERY SAME SURGERY. THE FINAL OUTCOME OF THE SURGERY WAS SUCCESSFUL AS INTENDED, WITH ALL PLACEMENTS CORRECT AT THE END OF THE SURGERY. THERE WAS NO DIRECT (OR INCREASED) RISK TO HARM A CRITICAL STRUCTURE DUE TO THE DEVIATING PLACEMENTS. THERE WAS NO HARM NOR NEGATIVE EFFECT TO THE PATIENT DUE TO THE DEVIATING INITIAL PLACEMENTS, ALSO NOT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20924389 ↗InjuryNov 6, 2024Use of Device Problem; Use of Device Problem
A MINIMALLY INVASIVE SURGERY ON THE LUMBAR SPINE FOR A TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) OF VERTEBRAE L3 AND L4, WITH DISC DECOMPRESSION AND CAGE PLACEMENT, WITH INTENDED PLACEMENT OF 4 VERTEBRA SCREWS BILATERAL FOR FIXATION, WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB SPINE & TRAUMA 3D NAVIGATION 1.5. FROM AN INTRA-OPERATIVE VERIFICATION C-ARM SCAN, THE SURGEON DETERMINED THAT THE 4 SPINE SCREWS PLACED IN L3 AND L4 WITH NAVIGATION, DEVIATED FROM THEIR INTENDED POSITIONS SHIFTED LATERALLY BY CA. 4-6MM, WITH THE SCREWS ON THE RIGHT SIDE MISSING THE VERTEBRAL BODY AT THEIR TARGET POINTS. ACCORDING TO THE SURGEON (TREATING CLINICIAN): - THE DEVIATION OF SPINE SCREWS PLACED WITH THE AID OF NAVIGATION WAS DETECTED BY THE SURGEON WITH A VERIFICATION C-ARM SCAN BEFORE FINALIZING THE SURGERY, AND THE SCREW PLACEMENTS WERE CORRECTED TO THEIR INTENDED POSITIONS UNDER FLUOROSCOPIC GUIDANCE AT THE VERY SAME SURGERY. - THE FINAL OUTCOME OF THE SURGERY WAS SUCCESSFUL AS INTENDED, WITH ALL PLACEMENTS CORRECT AT THE END OF THE SURGERY. - THERE WAS NO DIRECT (OR INCREASED) RISK TO HARM A CRITICAL STRUCTURE DUE TO THE DEVIATING PLACEMENTS. - THERE WAS NO HARM NOR NEGATIVE E
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20619667 ↗InjurySep 18, 2024Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
A MINIMALLY INVASIVE SURGERY, FOR A LEFT ACETABULAR FRACTURE AND PELVIC RING FIXATION AFTER A MOTOR VEHICLE ACCIDENT, WAS PERFORMED ON THE SACRAL SPINE AND ILIUM, WITH INTENDED PLACEMENT OF 1 ILIOSACRAL SCREW AT S1 AND 1 TRANS-SACRAL SCREW AT S2, WITH THE AID OF THE DISPLAY BY THE BRAINLAB SOFTWARE SPINE & TRAUMA 3D NAVIGATION 1.5. FROM AN INTRA-OPERATIVE C-ARM SCAN, THE SURGEONS DISCOVERED THAT THE SCREW AT S1 (WHICH FOLLOWED THE PATH OF A NAVIGATED GUIDE WIRE) DEVIATED FROM ITS INTENDED PATH WITH A CA. 25 MM CRANIAL SHIFT. DURING THE PROCEDURE, THE SURGEONS: - WITH THE PATIENT IN SUPINE POSITION, ATTACHED THE NAVIGATION REFERENCE ARRAY TO A SCHANZ PIN INSERTED INTO THE RIGHT ANTERIOR SUPERIOR ILIAC SPINE. - DRAPED THE PATIENT AND SECURED A UNIVERSAL AUTOMATIC IMAGE REGISTRATION (UAIR) MATRIX ON TOP OF THE DRAPE. - ACQUIRED AN INTRA-OPERATIVE C-ARM SCAN OF THE PATIENT'S REGION OF INTEREST WITH AUTOMATIC IMAGE REGISTRATION OF THE CURRENT PATIENT ANATOMY TO THE NAVIGATION. - PLANNED TRAJECTORIES FOR THE SCREWS BUT DECIDED TO PERFORM A FURTHER REDUCTION OF THE PELVIS WITHOUT NAVIGATION. - AFTERWARDS, PERFORMED ANOTHER C-ARM SCAN WITH AUTOMATIC IMAGE REGISTRATION, AND FUSED BOTH SCANS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20244616 ↗InjuryJul 11, 2024Use of Device Problem; Use of Device Problem
A MINIMALLY INVASIVE SURGERY ON THE LUMBOSACRAL SPINE FOR AN EXCISION OF INTRASPINAL TUMOR AT L5-S1 AND POSTERIOR FUSION OF VERTEBRAE L5-S1, WITH INTENDED PLACEMENT OF 4 VERTEBRA SCREWS BILATERAL FOR FIXATION, WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB SOFTWARE SPINE & TRAUMA 3D NAVIGATION 1.5.1. DURING THE PROCEDURE THE SURGEON: - WITH THE PATIENT IN PRONE POSITION ATTACHED THE NAVIGATION REFERENCE ARRAY ON THE LEFT ILIAC CREST. - ACQUIRED AN INTRA-OPERATIVE C-ARM SCAN OF THE PATIENT'S REGION OF INTEREST WITH AUTOMATIC IMAGE REGISTRATION OF THE CURRENT PATIENT ANATOMY TO THE NAVIGATION. - VERIFIED THE REGISTRATION, AND ACCEPTED THE ACCURACY TO PROCEED. - USED THE BRAINLAB NAVIGATED PEDICLE ACCESS NEEDLE (PAN) WITH INSTRUMENT POSITION DISPLAY ON THE PATIENT SCAN TO DETERMINE THE TRAJECTORY AND GET ACCESS TO THE VERTEBRAL BODY THROUGH THE SKIN. - STARTING WITH LEFT S1, PLACED THE K-WIRE THROUGH THE PEDICLE ACCESS NEEDLE. - WHEN CREATING THE SECOND PILOT HOLE IN LEFT L5 USING THE SAME METHOD, DETERMINED THAT THE TRAJECTORY AND LOCATION OF THE NAVIGATED PAN DISPLAYED BY THE NAVIGATION MAY NOT MATCH ITS PHYSICAL PATH AND POSITION IN THE BONE. - ACQUIRED A FLUOROSCOPY IMA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19721287 ↗InjuryApr 30, 2024Use of Device Problem; Use of Device Problem
AN OPEN SURGERY ON THE CERVICAL AND THORACIC SPINE FOR A FUSION AND DECOMPRESSION OF VERTEBRAE C7 TO T5, DUE TO A VERTEBRA T2 PATHOLOGICAL COMPRESSION FRACTURE, WITH INTENDED PLACEMENT OF 10 VERTEBRA SCREWS BILATERAL FOR FIXATION, WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB SPINE & TRAUMA 3D NAVIGATION 1.5. DURING THE PROCEDURE THE SURGEON: - WITH THE PATIENT IN PRONE POSITION ATTACHED THE NAVIGATION REFERENCE ARRAY ON THE SPINOUS PROCESS OF VERTEBRA T5. - ACQUIRED AN INTRA-OPERATIVE C-ARM SCAN OF THE PATIENT'S REGION OF INTEREST WITH AUTOMATIC IMAGE REGISTRATION OF THE CURRENT PATIENT ANATOMY TO THE NAVIGATION. - VERIFIED THE REGISTRATION AND ACCEPTED THE NAVIGATION ACCURACY TO PROCEED. - USED THE NAVIGATED POINTER WITH INSTRUMENT POSITION DISPLAY ON THE PATIENT SCAN TO DETERMINE THE TRAJECTORY (PATH) FOR THE SCREW PLACEMENT. - CALIBRATED A NON-BRAINLAB DRILL GUIDE, A NON-BRAINLAB TAP, AND A NON-BRAINLAB SCREWDRIVER TO THE NAVIGATION FOR INSTRUMENT POSITION DISPLAY. - STARTING WITH THE VERTEBRA LEFT C7, CREATED A PILOT HOLE DRILLING THROUGH THE NAVIGATED DRILL GUIDE, THE NAVIGATED TAP TO THREAD IT, AND THE NAVIGATED SCREWDRIVER TO PLACE THE SPINE SCREW INTO THE PILOT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19211058 ↗InjuryMar 28, 2024Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
AN OPEN SURGERY (ON (B)(6) 2024) ON THE THORACOLUMBAR SPINE FOR FUSION OF VERTEBRAE T10-L2, DUE TO A FRACTURE AT T12, WITH INTENDED PLACEMENT OF 8 SPINE SCREWS BILATERALLY FOR FIXATION (AT T10, T11, L1, L2), WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB NAVIGATION SOFTWARE SPINE & TRAUMA 3D NAV 1.5.1. DURING THE PROCEDURE THE SURGEON: - WITH THE PATIENT IN PRONE POSITION ATTACHED THE NAVIGATION REFERENCE ARRAY TO THE SPINOUS PROCESS OF L3. - ACQUIRED AN INTRA-OPERATIVE 3D C-ARM SCAN OF THE PATIENT'S REGION OF INTEREST WITH AUTOMATIC IMAGE REGISTRATION OF THE CURRENT PATIENT ANATOMY TO THE NAVIGATION. - VERIFIED THE REGISTRATION AND ACCEPTED THE ACCURACY TO PROCEED - PLACED SCREWS AT T10-L2 USING THE FOLLOWING METHOD: USED THE NAVIGATED BRAINLAB POINTER TO DETERMINE THE SCREW ENTRY POINT, USED A NON-NAVIGATED DRILL TO MARK THE ENTRY POINT AND CREATE A DIVOT TO PREVENT SKIVING, USED THE NAVIGATED BRAINLAB DRILL GUIDE TO CREATE THE PILOT HOLE, USED A NAVIGATED NON-BRAINLAB TAP TO THREAD THE PILOT HOLE, AND USED A NAVIGATED NON-BRAINLAB SCREWDRIVER TO PLACE THE SCREW. - AFTER COMPLETING SCREW PLACEMENTS AT T10 AND T11, VERIFIED NAVIGATION ACCURACY AND DETERMINED IT ACCURATE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18993117 ↗InjuryFeb 29, 2024Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
AN OPEN SURGERY ON THE LUMBAR SPINE FOR LAMINECTOMIES AND FUSION OF VERTEBRAE L4-5, DUE TO SPONDYLOLISTHESIS, WITH INTENDED PLACEMENT OF AN INTERBODY CAGE AND 4 SPINE SCREWS BILATERALLY FOR FIXATION (AT L4-5), WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB NAVIGATION SOFTWARE SPINE & TRAUMA 3D NAV 1.5. DURING THE PROCEDURE THE SURGEON: - WITH THE PATIENT IN PRONE POSITION ACQUIRED A C-ARM SCAN TO LOCALIZE AND MAKE A MEDIAL INCISION AT THE REGION OF INTEREST. - PERFORMED LAMINECTOMIES AT L4 AND L5 THROUGH THE MEDIAL INCISION BEFORE PLACING AN INTERBODY CAGE BETWEEN THOSE VERTEBRAE. - ATTACHED THE NAVIGATION REFERENCE ARRAY TO THE CAUDAL END OF THE L3 SPINOUS PROCESS AND THE TISSUE BETWEEN L3 AND L4 SPINOUS PROCESSES. - ACQUIRED AN INTRA-OPERATIVE 3D C-ARM SCAN OF THE PATIENT'S REGION OF INTEREST WITH AUTOMATIC IMAGE REGISTRATION OF THE CURRENT PATIENT ANATOMY TO THE NAVIGATION. - VERIFIED THE REGISTRATION AND ACCEPTED THE ACCURACY TO PROCEED. - USED A BRAINLAB POINTER TO DETERMINE THE LOCATION FOR SCREWS TO BE PLACED ON THE PATIENT'S LEFT SIDE AND MADE A LEFT LATERAL INCISION. - CALIBRATED A NON-BRAINLAB TAP TO THE NAVIGATION FOR INSTRUMENT POSITION DISPLAY ON THE PATIENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18806228 ↗InjuryFeb 29, 2024Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
AN OPEN SURGERY (ON (B)(6) 2024) ON THE LUMBAR SPINE FOR FUSION OF L4-S1, WITH INTENDED PLACEMENT OF 6 SPINE SCREWS BILATERALLY (AT L4-S1), FOR PAIN MANAGEMENT OF A PATIENT WITH A MEDICAL HISTORY OF MS DISEASE, WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB NAVIGATION SOFTWARE SPINE& TRAUMA 3D NAV. 1.5.1: DURING THE PROCEDURE THE SURGEONS: - WITH THE PATIENT IN PRONE POSITION ATTACHED THE NAVIGATION REFERENCE ARRAY TO SACRUM - ACQUIRED AN INTRA-OPERATIVE 3D (X-RAY) SCAN OF THE PATIENT'S REGION OF INTEREST WITH AUTOMATIC IMAGE REGISTRATION OF THE CURRENT PATIENT ANATOMY TO THE NAVIGATION, VERIFIED THE REGISTRATION AND ACCEPTED THE ACCURACY TO PROCEED - PLACED SCREWS AT L4, L5, AND S1 (LEFT SIDE FIRST, THEN RIGHT SIDE) AS FOLLOWS: -- USED THE BRAINLAB POINTER WITH INSTRUMENT POSITION DISPLAY ON THE PATIENT SCAN TO DETERMINE THE SCREW TRAJECTORY IN THE VERTEBRA -- USED A NAVIGATED NON-BRAINLAB AWL TO OPEN THE CORTICAL BONE, AND USED A NAVIGATED NON-BRAINLAB FLAT PROBE TO CANNULATE (ADDITIONALLY USED THE POINTER AND TWO OTHER NAVIGATED NON-BRAINLAB PROBES TO VERIFY LOCATION OF THE ENTRY/PILOT HOLE) -- USED A NON-NAVIGATED NON-BRAINLAB SCREWDRIVER TO PLACE THE SPINE SCREW INT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18806054 ↗InjuryFeb 15, 2024Use of Device Problem; Use of Device Problem
AN OPEN SURGERY ON THE LUMBAR SPINE, WITH INTENDED PLACEMENT OF 6 SPINE SCREWS BILATERALLY (AT L3-5), FOR AN EXTENSION OF FUSION FROM L3-5 (INCLUDING A REPLACEMENT OF PRE-EXISTING SPINE SCREWS AT L4-5), WAS PERFORMED WITH THE AID OF THE DISPLAY BY THE BRAINLAB SOFTWARE SPINE & TRAUMA 3D NAVIGATION 1.5. DURING THE PROCEDURE THE SURGEON: - WITH THE PATIENT IN PRONE POSITION ATTACHED THE NAVIGATION REFERENCE ARRAY ON THE RIGHT ILIAC CREST WITH TWO SCHANZ PINS. - ACQUIRED AN INTRA-OPERATIVE 3D C-ARM SCAN OF THE PATIENT'S REGION OF INTEREST WITH AUTOMATIC IMAGE REGISTRATION OF THE CURRENT PATIENT ANATOMY TO THE NAVIGATION. - VERIFIED THE REGISTRATION AND ACCEPTED THE ACCURACY TO PROCEED. - STARTING AT VERTEBRA LEFT L3, USED A NAVIGATED DRILL GUIDE WITH INSTRUMENT POSITION DISPLAY ON THE PATIENT SCAN TO DETERMINE THE SCREW TRAJECTORY IN THE VERTEBRA. - PREPARED THE PEDICLE BY DRILLING A PILOT HOLE THROUGH THE NAVIGATED DRILL GUIDE. - CALIBRATED A NON-BRAINLAB TAP TO THE NAVIGATION FOR INSTRUMENT POSITION DISPLAY ON THE PATIENT SCAN AND THREADED THE PILOT HOLE. - CALIBRATED A NON-BRAINLAB SCREWDRIVER TO THE NAVIGATION AND SECURED THE SPINE SCREW IN THE THREADED PILOT HOLE. - MAINTAINED TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18710220 ↗
Other Brainlab SE products
- Spine & Trauma Navigation2 models · 139 reports
- Navigation Software Spine & Trauma 3D2 models · 108 reports
- Cranial2 models · 64 reports
- Exactrac2 models · 6 reports · 1 recall
- ExacTrac Dynamic2 models · 2 reports · 1 recall
- Origin Data Management / Patient Data Manager2 models · 0 reports · 1 recall
- Spine Srs2 models · 0 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page