Spine & Trauma 3D Navigation recallZ-1082-2019 · Brainlab AG · initiated Feb 27, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Brainlab AG or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.
FDA root cause category: Software design
What the company is doing
The firm, BrainLab, notified their existing customer base by email on with a letter entitled "URGENT FIELD SAFETY NOTICE / PRODUCT NOTIFICATION" 02/25/2019. The notice explained the issue, product and actions to be taken. The customers were instructed to do the following: "User Corrective Action: 1. Avoid workflow changes with already registered datasets, if possible. If not absolutely necessary (e.g., for intraoperative screw planning), avoid switching between the workflow application selections 3D Navigation Intraoperative Imaging and 3D Navigation during one patient treatment. 2. After restarting the application with a previously registered dataset (crash restore or workflow change), always activate and deactivate the sub-menu Orientation once to ensure that the correct, expected view representations are displayed for the current session. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning in the Brainlab Spine & Trauma 3D Navigation 1.0 Software User Guide: Before beginning any surgical step, always ensure that navigation accuracy can be sufficiently maintained. If a surgical step could easily lead to reference displacement, remount the reference at an optimal position with a reduced possibility of displacement, and then register and verify again. The following methods can be used to ensure accuracy is sufficient: - Landmark verification - Fluoroscopy (compare the instrument location in the fluoro image with the display on the navigation system)" Brainlab will provide all affected customers with a software revision of Spine & Trauma Navigation Software with the described issue corrected. Brainlab will actively contact you, starting in June 2019 to schedule the update. If you have any questions, contact the Customer Hotline: +49 89 99 15 68 1044 or +1 800 597 5911 (for US customers)
Recall details
- Recall number
- Z-1082-2019
- Status
- Terminated
- Initiated by the company
- Feb 27, 2019
- Posted by the FDA
- Apr 15, 2019
- Recalling firm
- Brainlab AG
- Quantity
- 148 units
- Product code
- HAW
- FDA submissions
- K070106
- Distribution
- Worldwide Distribution: US (Nationwide) and countries of: Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Ecuador, France, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Kazakhstan, Malaysia, New Zealand, Philippines, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand and United Kingdom.
Product description
Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery.
Code information (models, lots, serial numbers)
Version 1.0.0.249, GTIN 04056481140328
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Spine & Trauma 3D Navigation 1.0Brainlab SE · Neurological Stereotaxic Instrument1 recall4 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K070106 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Navigation Software Spine & Trauma 3D 2.5Brainlab SE · Neurological Stereotaxic Instrument5 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page