Fhc, Inc. · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls
3

We found 3 models of Star Drive M/E by Fhc, Inc. in the FDA's device registry (GUDID). None of them is named in an adverse event report in our data.

3 FDA recall records name at least one of these models by its device identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Star Drive M/E recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 10, 2018Terminated

    The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

    Z-2359-2018 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  2. May 10, 2018Terminated

    The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

    Z-2358-2018 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  3. May 10, 2018Terminated

    The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

    Z-2352-2018 · Root cause: Device Design · Recall details · FDA record ↗