Star Drive M/E recallZ-2358-2018 · FHC, Inc. · initiated May 10, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from FHC, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
FDA root cause category: Device Design
What the company is doing
Urgent Medical Device Recall notification letters were sent to customers on 5/10/18. The letters included the following instructions: FHC, Inc. Field Representatives will review your inventory and replace effected Depth Stop Adapters you have on hand with an improved design. In the interim, we advise that you do not over-tighten the Depth Stop Adapter screw. If low impedances or a short circuit is found, physicians should replace the lead prior to use and report the issue to FHC, Inc. or your local field rep.
Recall details
- Recall number
- Z-2358-2018
- Status
- Terminated
- Initiated by the company
- May 10, 2018
- Recalling firm
- FHC, Inc.
- Quantity
- 2952 total
- Product code
- HAW
- FDA submissions
- K092562
- Distribution
- The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
Product description
microTargeting" STar" Drive System (Motorized) Distributed by Medtronic, Product Number FC8002
Code information (models, lots, serial numbers)
UDI: 00873263006375, 00873263001493
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Star™ Drive M/E FC8002-00149Fhc, Inc. · Neurological stereotaxic Instrument2 recalls0 reports
- Star™ Drive M/E FC8002-00637Fhc, Inc. · Neurological stereotaxic Instrument2 recalls0 reports
Product lines
Other devices under the same FDA submission
7 devices cleared under K092562 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- STar™ Drive MA FC8001-00148Fhc, Inc. · Neurological stereotaxic Instrument1 recall1 report
- STar™ Drive MA 70-ZD-MA-00216Fhc, Inc. · Neurological stereotaxic Instrument1 recall1 death1 report
- STar™ Drive M/E 70-ZD-ME-00215Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- STar™ Drive Lead Adapter 70-CN-DB-00264Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- STar™ Drive MA System ST-DS-MA-00441Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- STar™ Drive M/E System ST-DS-ME-00443Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- STar™ Drive MA FC8001-00635Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80148).
- Z-2345-2018 · TerminatedmicroTargeting" Drive System, Product Number MT-DS
- Z-2346-2018 · TerminatedmicroTargeting" Drive System, Product Number MT-DS-01
- Z-2347-2018 · TerminatedmicroTargeting" Drive System, Product Number 66-ZD-MD
- Z-2348-2018 · TerminatedmicroTargeting" Drive System, Product Number 66-ZD-MD-01
- Z-2349-2018 · TerminatedmicroTargeting" STar" Drive System (Manual), Product Number ST-DS-MA
- Z-2350-2018 · TerminatedmicroTargeting" STar" Drive System (Motorized), Product Number ST-DS-ME
- Z-2351-2018 · TerminatedmicroTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA
- Z-2352-2018 · TerminatedmicroTargeting" STar" Drive System (Motorized), Product Number 70-ZD-ME
- Z-2353-2018 · TerminatedmicroTargeting" Lead Adapter for microTargeting" Drive, Product Number 66-CN-DB
- Z-2354-2018 · TerminatedmicroTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB
- Z-2355-2018 · TerminatedmicroTargeting" Drive System Distributed by Medtronic, Product Number FC1006
- Z-2356-2018 · TerminatedmicroTargeting" Drive System Distributed by Medtronic, Product Number 9033G0601
- Z-2357-2018 · TerminatedmicroTargeting" STar" Drive System (Manual) Distributed by Medtronic, Product Number FC8001
- Z-2359-2018 · Terminatedmicrotargeting" Depth Stop Adapter, Product Model Number, Product Number 66-AC-DS(1.8),
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗