2 open recalls name at least one Superion model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,488
Deaths reported
0
Injuries reported
408
Recalls
2

We found 5 models of Superion by Boston Scientific Neuromodulation Corporation in the FDA's device registry (GUDID). Together they are named in 1,488 adverse event reports in the FDA's MAUDE database, received between Nov 14, 2018 and Jun 29, 2026 — classified as 408 injuries, 1,080 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (994 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 14 reportsDec 2023: 10 reportsJan 2024: 14 reportsFeb 2024: 6 reportsMar 2024: 5 reportsApr 2024: 18 reportsMay 2024: 10 reportsJun 2024: 10 reportsJul 2024: 7 reportsAug 2024: 4 reportsSep 2024: 2 reportsOct 2024: 12 reportsNov 2024: 9 reportsDec 2024: 2 reportsJan 2025: 7 reportsFeb 2025: 9 reportsMar 2025: 12 reportsApr 2025: 5 reportsMay 2025: 2 reportsJun 2025: 10 reportsJul 2025: 8 reportsAug 2025: 7 reportsSep 2025: 2 reportsOct 2025: 447 reportsNov 2025: 3 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports202420252026447
Last 36 months. The most recent month may be incomplete.

Superion recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 6, 2025Open, Classified

    As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels. Issues include breakage of the instruments and implant/device.

    Z-2485-2025 · Root cause: Device Design · names 5 of these models · Recall details · FDA record ↗

  2. Oct 12, 2022Open, Classified

    Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.

    Z-0230-2023 · Root cause: Process change control · names 3 of these models · Recall details · FDA record ↗

Most reported Superion problems

Device problems

  • Material Integrity Problem994
  • Adverse Event Without Identified Device or Use Problem256
  • Difficult to Open or Close150
  • Migration127
  • Lack of Effect51
  • Device Dislodged or Dislocated40
  • Material Twisted/Bent40
  • Malposition of Device12

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,061
  • Pain166
  • Insufficient Information99
  • No Code Available88
  • Bone Fracture(s)52
  • Vertebral Fracture43
  • Therapeutic Effects, Unexpected26
  • Discomfort25

Superion models

5 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Superion adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Boston Scientific Neuromodulation Corporation products

All Boston Scientific Neuromodulation Corporation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page