Superion™ 101-9808Recalls & reported problems

Boston Scientific Neuromodulation Corporation · Indirect Decompression System · Class 3 · Implant

All models of Superion (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884662000529), received between Oct 15, 2025 and Oct 31, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (P140004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 4 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 6, 2025Open, Classified

    As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels. Issues include breakage of the instruments and implant/device.

    Z-2485-2025 · Root cause: Device Design · Recall details · FDA record ↗

  2. Oct 12, 2022Open, Classified

    Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.

    Z-0230-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Open or Close2
  • Adverse Event Without Identified Device or Use Problem1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 31, 2025Difficult to Open or Close; Difficult to Open or Close

    DEVICE ANALYSIS PERFORMED ON THE RETURNED SUPERION INDIRECT DECOMPRESSION SPACER REVEALED A SEVERE ABRASION ON THE MATING SURFACE OF THE ACTUATOR AND SEVERE SCRAPES WERE OBSERVED ON ONE OF THE CAM LOBES. HOWEVER, THE IMPLANT FUNCTIONED ACCEPTABLY DURING THE FUNCTIONAL TEST. THE DAMAGE TO THE IMPLANT INDICATES FAILURE WAS LIKELY DUE TO DEPLOYMENT AGAINST RESISTANCE SUCH AS BONE. ADDITIONALLY, A LABELING REVIEW WAS COMPLETED AND REVEALED NO ANOMALIES. THE INSTRUCTIONS FOR USE (IFU) STATES DEPLOYMENT SHOULD NOT BE FORCED OR IMPLANT BREAKAGE OR DAMAGE TO BONY STRUCTURES MAY RESULT. FURTHERMORE, IT STATES THAT IF RESISTANCE TO DEPLOYMENT IS ENCOUNTERED, AND REPOSITIONING OF THE IMPLANT IS REQUIRED, ROTATE THE DRIVER COUNTERCLOCKWISE UNTIL A POSITIVE STOP IS REACHED AND WITHDRAW DORSALLY TO COLLAPSE THE CAM LOBES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441610 ↗
  • MalfunctionOct 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE ANALYSIS PERFORMED ON THE RETURNED SUPERION INDIRECT DECOMPRESSION SPACER REVEALED THAT THE SPINDLE CAP WAS COMPLETELY SHEARED OFF FROM THE IMPLANT BODY. THE DAMAGE TO IMPLANT INDICATED A FAILURE, LIKELY DUE TO A COMBINATION OF DEPLOYMENT AGAINST RESISTANCE (E.G., BONE) AND THE PHYSICIAN FAILING TO CORRECTLY ATTACH THE INSERTER TO THE SPACER. PRODUCT ANALYSIS CONFIRMED THAT THE MOST PROBABLE CAUSE OF THE COMPLAINT IS UNINTENDED USE ERROR CAUSED OR CONTRIBUTED TO EVENT. ADDITIONALLY, A LABELING REVIEW WAS COMPLETED AND REVEALED NO ANOMALIES. THE INSTRUCTIONS FOR USE STATE THAT DEPLOYMENT SHOULD NOT BE FORCED OR IMPLANT BREAKAGE OR DAMAGE TO BONY STRUCTURES MAY RESULT. FURTHERMORE, IF RESISTANCE TO DEPLOYMENT IS ENCOUNTERED, AND REPOSITIONING OF THE IMPLANT IS REQUIRED, THE DRIVER SHOULD BE ROTATED COUNTERCLOCKWISE UNTIL A POSITIVE STOP IS REACHED AND WITHDRAW DORSALLY TO COLLAPSE THE CAM LOBES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23432745 ↗
  • MalfunctionOct 30, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT DURING A SUPERION IMPLANT PROCEDURE, THE SPACER WOULD NOT OPEN, AND THE PROCEDURE HAD TO BE ABORTED. THE PATIENTS ANATOMY WAS CHALLENGING AS THERE WAS LIMITED SPACE IN THE SPINOUS PROCESS. THE PHYSICIAN DID NOT USE EXCESSIVE FORCE DURING THE PROCEDURE. THE BROKEN SPACER WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23432240 ↗
  • MalfunctionOct 15, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING A SUPERION IMPLANT PROCEDURE, THE SPACER BROKE INTO PIECES DURING DEPLOYMENT. THE BROKEN SPACER WAS REMOVED AND A NEW SPACER WAS USED TO COMPLETE THE PROCEDURE. THE PHYSICIAN IS BELIEVED TO HAVE USED EXCESSIVE FORCE DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23280924 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140004. One submission often covers many models, so check the product and lot details in each record.

  • Z-1972-2023 · May 12, 2023 · Open, Classified
    Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Techn

Other Prosthesis, Spinous Process Spacer/Plate devices

All Prosthesis, Spinous Process Spacer/Plate devices →