Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one SureSigns VM8 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls
1

We found 2 models of SureSigns VM8 by Philips Medical Systems Hsg in the FDA's device registry (GUDID). Together they are named in 14 adverse event reports in the FDA's MAUDE database, received between Jun 5, 2019 and Feb 19, 2024 — classified as 14 malfunctions.

The most frequently reported specific device problem is “No Audible Alarm” (7 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

SureSigns VM8 recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jul 13, 2018Open, Classified

    Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.

    Z-2895-2018 · Root cause: Labeling design · names 2 of these models · Recall details · FDA record ↗

Most reported SureSigns VM8 problems

Device problems

  • No Audible Alarm7
  • No Audible Prompt/Feedback3
  • Audible Prompt/Feedback Problem2
  • Application Program Freezes or Fails to Launch1
  • Defective Component1
  • Display Difficult to Read1
  • Erratic or Intermittent Display1
  • Fitting Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • No Patient Involvement2

SureSigns VM8 models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent SureSigns VM8 adverse event reports

The 14 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Philips Medical Systems Hsg products

All Philips Medical Systems Hsg devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page