Synergy SurfaceRecalls & reported problems across all models
Hill-Rom, Inc. · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 0
We found 4 models of Synergy Surface by Hill-Rom, Inc. in the FDA's device registry (GUDID). Together they are named in 32 adverse event reports in the FDA's MAUDE database, received between Aug 2, 2022 and Sep 10, 2026 — classified as 2 injuries, 30 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (28 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Synergy Surface problems
Device problems
- Material Split, Cut or Torn28
- Material Frayed2
- Sparking2
- Adverse Event Without Identified Device or Use Problem1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- Pressure Sore/Ulcer2
Synergy Surface models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Synergy® Surface P6881MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE28 reports
- Synergy® Surface P6881CAPBariatric bed2 reports
- Synergy® Surface P004651Bariatric bed1 report
- Synergy® Surface P6880Bariatric bed1 report
Recent Synergy Surface adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50720853 ↗MalfunctionAug 28, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50509465 ↗InjuryMay 7, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE AIR CELLS ON THE SYNERGY ELITE MATTRESS ARE SHIFTING TO ONE SIDE AND DO NOT SEEM TO FULLY INFLATE AND THE CUSTOMER HAS A NEW PRESSURE WOUND. CUSTOMER INDICATED PRIOR HISTORY OF FLAP SURGERY 3 1/2 YEARS AGO, DID NOT PROVIDE DETAILS OF LOCATION, AND ABOUT 6-8 MONTHS AGO PATIENT STARTED WITH A NEW PRESSURE SORE ON THE SACRUM AND IT IS STAGE 4. THE MEDICAL INTERVENTION PROVIDED FOR THE PRESSURE WOUND IS REGULAR DRESSING CHANGES AND THEY ARE GOING TO START USING A WOUND VAC AND THE HYPERBARIC CHAMBER. IT WAS ALSO REPORTED THE PATIENT WAS TRANSFERRED TO A REHAB FACILITY WHERE THE EXPECTED DURATION OF STAY WAS INITIALLY ESTIMATED AT 6¿8 MONTHS BUT WAS LATER REVISED TO APPROXIMATELY 20 DAYS. THE PATIENT STATED THAT THE MATTRESS WAS MANIPULATED BY A NON-BAXTER EMPLOYEE, DID NOT PROVIDE SPECIFIC DATE ON WHEN THIS OCCURRED. THE PERSON SEWED THE MATTRESS AIR CELLS AND AFTER 2-3 MONTHS THE CELLS COME LOOSE AND MATTRESS WOULD SEPARATE. INSTRUCTIONS IN THE IFU ARE OUTLINE USERS: ¿UNZIP THE TOP COVERLET FROM THE BOTTOM ENCASEMENT AND EXAMINE THE AIR CELLS AND AIR TUBES FOR LEAKS. IF A LEAK IS FOUND, CONTACT BAXTER TECHNICAL SUPPORT FOR ASSISTANCE.¿. NO FURTHER INFORMATION
Patient problems: Pressure Sore/Ulcer
MAUDE report 25115542 ↗MalfunctionMar 12, 2026Material Split, Cut or Torn
ON 18-FEB-2026, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT BLOWER ASSEMBLY, SAE (PRODUCT CODE P6881, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24584441 ↗MalfunctionMar 2, 2026Material Split, Cut or Torn
ON 03-FEB-2026, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT BLOWER ASSEMBLY, SAE (PRODUCT CODE P6881, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24483707 ↗MalfunctionJan 29, 2026Material Split, Cut or Torn
ON 12-JAN-2026, A TECHNICAL SERVICE TO REPORT THAT BLOWER ASSEMBLY, SAE (PRODUCT CODE P6881, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24216830 ↗MalfunctionAug 27, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE BLOWER ASSEMBLY, SAE POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22907158 ↗MalfunctionAug 21, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE BLOWER ASSEMBLY, SAE POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22865892 ↗MalfunctionAug 14, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE BLOWER ASSEMBLY, SAE POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22809800 ↗MalfunctionAug 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE BLOWER ASSEMBLY, SAE POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT KNOW IF THIS EVENT WAS ALREADY COMMUNICATED TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22773909 ↗MalfunctionAug 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE BLOWER ASSEMBLY, SAE POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22772897 ↗MalfunctionJul 15, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE BLOWER ASSEMBLY, SAE POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22506511 ↗InjuryJul 2, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED A CUSTOMER SUSTAINED A PRESSURE SORE WITH A SYNERGY MATTRESS. THE USER WAS UTILIZING THE SYNERGY MATTRESS AT HOME. AFTER AN UNSPECIFIED AMOUNT OF TIME, THE USER FELT THE MATTRESS WAS ¿SUNKEN IN THE MIDDLE¿ AND ¿NO ERROR MESSAGES OR BEEPING¿ TRIGGED FROM THE BED. THE CUSTOMER WENT TO THE EMERGENCY ROOM WHERE SEVEN SUTURES WERE PLACED FOR A PRESSURE INJURY. REPORTEDLY, THE PRESSURE INJURY WAS ON THE SACRAL/GLUTEAL AREA AND ¿WAS PROBABLY LIKE A STAGE FIVE, IT WAS DOWN TO THE ARTERY.¿ THE PATIENT WAS ADMITTED INTO THE HOSPITAL WHERE THEY WERE TREATED WITH TRIAD, HYDROPHILIC WOUND DRESSING, AND CAUTERIZATION. THE PATIENT WAS DISCHARGED AFTER ¿TEN OR ELEVEN DAYS¿. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Pressure Sore/Ulcer
MAUDE report 22383810 ↗MalfunctionMar 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE BLOWER ASSEMBLY, SAE POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21599319 ↗MalfunctionFeb 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT BLOWER ASSEMBLY, SAE (PRODUCT CODE P6881, SERIAL NUMBER (B)(6), HAD POWER CORD FRAYED WITH EXPOSED COPPER WIRES. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21413281 ↗
Other Hill-Rom, Inc. products
- Centrella Bed2 models · 2,021 reports · 2 recalls
- VersaCare Med Surg Bed2 models · 1,200 reports
- pro+18 models · 36 reports · 13 recalls
- Voalte Nurse Call2 models · 419 reports · 2 recalls
- ALLEN24 models · 73 reports · 5 recalls
- WatchCare4 models · 0 reports · 4 recalls
- NaviCare Nurse Call Systems3 models · 99 reports · 2 recalls
- CompellaSurface3 models · 87 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page