THUNDERBEAT 5 mm, 35 cm, Inline GripRecalls & reported problems across all models
Olympus Medical Systems Corp. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one THUNDERBEAT 5 mm, 35 cm, Inline Grip model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 1
We found 2 models of THUNDERBEAT 5 mm, 35 cm, Inline Grip by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 4 adverse event reports in the FDA's MAUDE database, received between Mar 26, 2019 and Apr 30, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
THUNDERBEAT 5 mm, 35 cm, Inline Grip recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 22, 2025Open, Classified
Firm is initiating a removal due to continued reports of adverse events.
Z-1380-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported THUNDERBEAT 5 mm, 35 cm, Inline Grip problems
Device problems
- Break3
- Detachment of Device or Device Component2
- Melted1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
THUNDERBEAT 5 mm, 35 cm, Inline Grip models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- THUNDERBEAT 5 mm, 35 cm, Inline Grip TB-0535ICElectrosurgical, Cutting & Coagulation & Accessories1 recall0 reports
- THUNDERBEAT 5 mm, 35 cm, Inline Grip TB-0535ICUnit, electrosurgical, endoscopic (with or without accessories)4 reports
Recent THUNDERBEAT 5 mm, 35 cm, Inline Grip adverse event reports
The 4 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryApr 30, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS INITIALLY REPORTED THAT THE GRASPING PART OF TWO ULTRASONIC SURGICAL DEVICES BROKE OFF. THE MOUTH WAS ALSO BROKEN OFF IN THE PACKAGING. CONSEQUENTLY, THERE WAS A 5-MINUTE DELAY TO THE LAPAROSCOPIC ANTERIOR RESECTION SURGERY TO GET A NEW DEVICE TWICE. THE CUSTOMER LATER CLARIFIED THAT THE PROBE OF THE OTHER ULTRASONIC SURGICAL DEVICE FELL INSIDE THE PATIENT AND WAS REMOVED USING FORCEPS, WHICH TOOK 1 MINUTE. THE SAME EXACT ISSUE REPORTEDLY HAPPENED FOR THE SUBJECT DEVICE. THERE WAS A 10-MINUTE DELAY TO THE THERAPEUTIC PROCEDURE WHILE THE PATIENT WAS UNDER GENERAL ANESTHESIA, WHICH WAS THEN COMPLETED WITHOUT FURTHER INCIDENT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19217201 ↗InjuryApr 30, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS INITIALLY REPORTED THAT THE GRASPING PART OF TWO ULTRASONIC SURGICAL DEVICES BROKE OFF. THE MOUTH WAS ALSO BROKEN OFF IN THE PACKAGING. CONSEQUENTLY, THERE WAS A 5-MINUTE DELAY TO THE LAPAROSCOPIC ANTERIOR RESECTION SURGERY TO GET A NEW DEVICE TWICE. THE CUSTOMER LATER CLARIFIED THAT THE PROBE OF THE SUBJECT DEVICE FELL INSIDE THE PATIENT AND WAS REMOVED USING FORCEPS, WHICH TOOK 1 MINUTE. THE SAME EXACT ISSUE REPORTEDLY HAPPENED FOR THE OTHER DEVICE. THERE WAS A 10-MINUTE DELAY TO THE THERAPEUTIC PROCEDURE WHILE THE PATIENT WAS UNDER GENERAL ANESTHESIA, WHICH WAS THEN COMPLETED WITHOUT FURTHER INCIDENT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19217092 ↗MalfunctionFeb 26, 2021Break; Break
DURING A LAPAROSCOPIC RADICAL HYSTERECTOMY, THE SUBJECT DEVICE WAS USED. AFTER 20 MINUTES SINCE THE SURGERY BEGAN, THE GENERATOR SHOWED AN ERROR IN THE PROBE. WHEN THE USER CHECKED THE SUBJECT DEVICE, THE PROBE WAS BROKEN OFF. THE FRAGMENT WAS RETRIEVED FROM THE PATIENT WITHOUT DIFFICULTIES. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11383399 ↗MalfunctionMar 26, 2019Melted; Melted
OLYMPUS WAS INFORMED THAT DURING A DIAGNOSTIC LAPAROSCOPY PROCEDURE, THE TEFLON PAD FROM THE GRASPING SECTION OF THE DEVICE MELTED. NO DEVICE FRAGMENT FELL INTO THE PATIENT. THERE WAS NO METAL EXPOSED ON THE DEVICE JAW. THE GENERATOR SETTINGS WERE UNKNOWN AND NO ERROR MESSAGES WERE OBSERVED. THE INTENDED PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONALLY, THE USER FACILITY REPORTED THAT THE DEVICE AND THE CONNECTION POINTS WERE INSPECTED PRIOR TO THE PROCEDURE WITH NO ANOMALIES FOUND.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8454215 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page