TOSHIBA SUPERCONDUCTING MRI SYSTEM Vantage TitanRecalls & reported problems across all models

Canon Inc. · 2 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name at least one TOSHIBA SUPERCONDUCTING MRI SYSTEM Vantage Titan model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls
2

We found 2 models of TOSHIBA SUPERCONDUCTING MRI SYSTEM Vantage Titan by Canon Inc. in the FDA's device registry (GUDID). Together they are named in 1 adverse event report in the FDA's MAUDE database, received between Jan 5, 2026 and Jan 5, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

TOSHIBA SUPERCONDUCTING MRI SYSTEM Vantage Titan recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 10, 2024Class IIOpen, Classified

    The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.

    Z-2041-2024 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jan 29, 2024Class IIOpen, Classified

    For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

    Z-1250-2024 · Root cause: Component change control · Recall details · FDA record ↗

Most reported TOSHIBA SUPERCONDUCTING MRI SYSTEM Vantage Titan problems

Device problems

  • Patient Device Interaction Problem1

Patient problems

  • Burn(s)1

TOSHIBA SUPERCONDUCTING MRI SYSTEM Vantage Titan models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.