TOSHIBA SUPERCONDUCTING MRI SYSTEM Vantage Titan recallZ-1250-2024 · Canon Medical System, USA, INC. · initiated Jan 29, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

FDA root cause category: Component change control

What the company is doing

On 01/29/24, correction notices were emailed to customers who were informed the following: The parts required to address the issue are expected to be available no earlier than February 2024. Until the countermeasure is performed, please use the values shown in the table shown in the correction notice. Share the contents of this letter with all users and reviewing radiologist as well as clinical engineering or biomedical group at your facility. If you have any questions regarding this letter please contact Regulatory Affairs at 800-421-1968 or your Canon service representative at 800-521-1968.

Recall details

Recall number
Z-1250-2024
Status
Open, Classified
Initiated by the company
Jan 29, 2024
Posted by the FDA
Mar 4, 2024
Recalling firm
Canon Medical System, USA, INC.
Quantity
40
Product code
LNH
FDA submissions
K122613
Distribution
Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.

Product description

Vantage Titan, Model: MRT-1504

Code information (models, lots, serial numbers)

UDI-DI: 04987670102400. Serial Numbers: U4B15Y2102, U5B15Z2001, U4B1492080, U4B1582100, U4B1582099, U4B1522091, U4B1512089, U4B1432067, U4B1522093, U4B1452073, U4B1442069, U4B1442068, U4B1642106, U4B1432066, U4B15X2101, U4B14Y2083, U4B1452071, U4B1512087, U4B1652107, U4B1552097, U4B1482078, U4B1512090, U4B1552098, U4B1442070, U4B14Y2082, U4B1482076, U4B14Z2085, U4B14Z2086, U4B1512088, U4B1612103, U4B1472075, U4B1422065, U4B1462074, U4B14Y2084, U4B1632105, U4B1452072, U4B1492081, U4B1482079, U4B1612104, U4B1482077

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93874).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page