TurboHawkRecalls & reported problems across all models
Medtronic, Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 56
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls
- 0
We found 3 models of TurboHawk by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 56 adverse event reports in the FDA's MAUDE database, received between Nov 26, 2018 and Sep 17, 2026 — classified as 7 injuries, 49 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported TurboHawk problems
Device problems
- Physical Resistance/Sticking24
- Detachment of Device or Device Component23
- Material Deformation12
- Difficult to Remove9
- Mechanical Jam5
- Difficult to Advance4
- Device Damaged by Another Device3
- Mechanical Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions37
- No Consequences Or Impact To Patient17
- Vascular Dissection2
- Perforation1
- Unspecified Tissue Injury1
TurboHawk models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- TurboHawk™ TH-LX-MCatheter, peripheral, atherectomy38 reports
- TurboHawk™ TH-LX-MCatheter, peripheral, atherectomy13 reports
- TurboHawk™ THS-LX-CCatheter, peripheral, atherectomy5 reports
Recent TurboHawk adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50805233 ↗MalfunctionJul 8, 2026Deformation Due to Compressive Stress; Detachment of Device or Device Component; Difficult to Advance; Physical Resistance/Sticking
IT WAS REPORTED THAT A TURBOHAWK DIRECTIONAL ATHERECTOMY DEVICE WAS USED DURING A DIRECTIONAL ATHERECTOMY PROCEDURE TO TREAT A CHRONIC TOTAL OCCLUSION (100% STENOSIS) IN THE MID SEGMENT OF THE SUPERFICIAL FEMORAL ARTERY, WITH THE TARGET LESION DESCRIBED AS CALCIFIED PLAQUE/SOFT TISSUE. THE VESSEL WAS PRE-DILATED WITH A 3.5 BALLOON, AND THE DEVICE WAS THEN USED TO START CUTTING FROM THE ORIGIN SITE OF THE LESION. DURING THE PROCEDURE, IT WAS OBSERVED THAT THE CUTTER COULD NOT EFFECTIVELY CUT TISSUE, AND THE CATHETER KINKED WHEN ADVANCED FORWARD, PREVENTING SMOOTH ADVANCEMENT. REPEATED ATTEMPTS TO ADVANCE AND USE THE DEVICE WERE UNSUCCESSFUL. THE DEVICE WAS EXCHANGED FOR ANOTHER DEVICE OF THE SAME TYPE TO COMPLETE THE PROCEDURE. NO POST-DILATION WAS PERFORMED. INSTRUCTIONS FOR USE WERE REPORTED AS FOLLOWED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. WHEN THE DEVICE WAS RETUNED FOR EVALUATION, IT WAS NOTED THAT THE DEVICE WAS DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25750763 ↗MalfunctionMay 5, 2026Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A PERIPHERAL/ENDOVENOUS DIRECTIONAL ATHERECTOMY USING THE TURBOHAWK ATHERECTOMY CATHETER TO TREAT A PROXIMAL SUPERFICIAL FEMORAL ARTERY LESION WITH CALCIFIED, FIBROUS PLAQUE AND 85% STENOSIS, TISSUE WAS UNABLE TO BE FLUSHED. THE EXCISED PLAQUE COULD NOT BE PUSHED OUT, AND WHEN AN ATTEMPT WAS MADE TO RESTART THE DEVICE, THE THUMB SWITCH COULD NOT BE PUSHED. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. NO DEFORMATION WAS NOTICED IN THE CUTTER. THE CATHETER WAS REPLACED WITH A NEW ONE, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. WHEN THE DEVICE WAS RECEIVDD FOR EVALUATION, IT WAS NOTED THAT THE HOUSING WALL & TECOTHANE HAD A HOLE IN THE CHAMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25079018 ↗MalfunctionFeb 23, 2026Material Deformation
IT WAS REPORTED THAT DURING A TARGETED RESECTION PROCEDURE USING THE ATHERECTOMY TH-LX-M TURBOHAWK ATK, AFTER A PORTION OF PLAQUE HAD BEEN REMOVED FROM THE MID SUPERFICIAL FEMORAL ARTERY, THE CATHETER WAS WITHDRAWN FOR CLEANING. THE PLAQUE IN THE COLLECTION CHAMBER COULD NOT BE FLUSHED OUT, AND A PROTRUSION WAS OBSERVED IN THE COLLECTION CHAMBER, MAKING IT IMPOSSIBLE TO CONTINUE USING THE DEVICE. AFTER INSPECTION IT WAS CONFIRMED THAT THE TECOTHANE JACKET WAS NOT COMPROMISED. A NEW HAWK CATHETER WAS USED TO COMPLETE THE SURGERY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. WHEN THE DEVICE WAS RETURNED IT WAS NOTED THAT THE DEVICE WAS DAMAGED AT THE HINGE PIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24413809 ↗MalfunctionJan 23, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A TARGETED RESECTION PROCEDURE USING THE TURBOHAWK ATK, A PATIENT PRESENTED WITH SEVERE CALCIFICATION AND 80% STENOSIS OF THE MID SUPERFICIAL FEMORAL ARTERY. THE CUTTER HEAD DETACHED AND COULD NOT BE PULLED OUT. NO PROCEDURAL STROKE CINERADIOGRAPHY IMAGES WERE AVAILABLE. THE VESSEL WAS NOT PRE-DILATED BUT WAS POST-DILATED, AND INSTRUCTIONS FOR USE WERE FOLLOWED DURING PREPARATION, PROCEDURE, AND POST-PROCEDURE. THE PROCEDURE WAS COMPLETED BY REPLACING THE DEVICE WITH ANOTHER OF THE SAME MODEL. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24161883 ↗MalfunctionJan 23, 2026Detachment of Device or Device Component; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A TARGETED EXCISION SURGERY FOR A LESION IN THE MID SUPERFICIAL FEMORAL ARTERY, WHICH WAS CHARACTERIZED BY SEVERE CALCIFICATION, CALCIFIED PLAQUE, AND 70% STENOSIS, THE BLADE TURBO HAWK ATK DEVICE COULD NOT BE PUSHED INTO THE COLLECTION CHAMBER AFTER THE FIRST PASS THROUGH THE LESION. THE THUMB SWITCH COULD NOT BE TURNED OFF. DURING THE CUTTING PROCESS, IT WAS FOUND THAT THE BLADE COULD NOT BE RETURNED TO THE COLLECTION BIN. AFTER THAT, THE POWER WAS TURNED OFF AND THE ENTIRE DEVICE WAS SUBSEQUENTLY WITHDRAWN FROM THE BODY, SAFELY REMOVED. THE TOOL IS DEFORMED AND NO PART OF THE TOOL WAS FOUND MISSING. THE COLLECTION BIN AND CONDUIT ARE NOT COMPLETELY DISCONNECTED. BALLOON DILATION WAS THEN PERFORMED ON THE LESION, AND THE SURGERY WAS SUCCESSFULLY COMPLETED. THE VESSEL WAS BOTH PRE-DILATED AND POST-DILATED. THE PATIENT HAD A HISTORY OF SMOKING. NO SYMPTOMS, COMPLICATIONS, ADVERSE EVENTS, INFECTIONS, DEATHS, OR ANATOMICAL ABNORMALITIES WERE REPORTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. WHEN THE DEVICE WAS EVALUATED IT WAS NOTED THAT THE TIP WAS DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24160703 ↗MalfunctionNov 5, 2025Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A DIRECTED RESECTION PROCEDURE USING THE TURBOHAWK DEVICE FOR TREATMENT OF A RIGHT MID SUPERFICIAL FEMORAL ARTERY LESION WITH 70% STENOSIS, MODERATE CALCIFICATION, AND FIBROUS AND CALCIFIED PLAQUE, SEVERAL DEVICE ISSUES OCCURRED. PREOPERATIVE PREPARATION WAS PERFORMED ACCORDING TO INSTRUCTIONS, INCLUDING PRE-DILATION WITH A 3.0 MILLIMETER BALLOON. DURING THE FIRST INSERTION, THE CUTTING HEAD WAS ADVANCED THROUGH THE LESION AND THEN WITHDRAWN FOR CUTTING FROM THE PROXIMAL TO DISTAL END. UPON REACHING THE DISTAL END AND ATTEMPTING TO RETRIEVE THE CUTTING HEAD, THE THUMB SWITCH COULD NOT BE PUSHED NORMALLY. THE SURGEON TURNED OFF THE HANDLE SWITCH AND ATTEMPTED MULTIPLE TIMES TO PUSH AND RETRIEVE, BUT THE SWITCH COULD NOT BE CLOSED PROPERLY, RESULTING IN A PROCEDURAL DELAY. THE ENTIRE DEVICE WAS REMOVED FROM THE PATIENT, AND IT WAS FOUND THAT THE PROXIMAL COLLECTION CHAMBER OF THE CUTTING HEAD HAD BEEN DAMAGED BY THE CUTTING HEAD ITSELF, THE CUTTER WAS DAMAGED, AND THE FLUSH PORT ON THE CATHETER HANDLE WAS ALSO DAMAGED. THE DAMAGED CUTTER REQUIRED RETRIEVAL FROM THE PATIENT BY REMOVING THE ENTIRE DEVICE. THE SURGEON REPLACED THE DEVICE WITH THE SAME MODEL A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23477371 ↗MalfunctionSep 18, 2025Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation
IT WAS REPORTED THAT DURING A DIRECTED RESECTION PROCEDURE USING THE TURBOHAWK ATK, A PATIENT WITH A CHRONIC TOTAL OCCLUSION (100% STENOSIS) OF THE MID SUPERFICIAL FEMORAL ARTERY, SEVERE VESSEL TORTUOSITY, AND MODERATE LESION CALCIFICATION UNDERWENT PUNCTURE ANGIOGRAPHY, WHICH REVEALED FEMORAL ARTERY OCCLUSION AND POOR VISUALIZATION OF THE POPLITEAL ARTERY. AFTER GUIDEWIRE PASSAGE THROUGH THE OCCLUDED SEGMENT, THE ATHERECTOMY CATHETER WAS INSTALLED AND FLUSHED, AND A SMALL BALLOON WAS USED FOR PRE-DILATION. DURING ADVANCEMENT OF THE ATHERECTOMY CATHETER, RESISTANCE WAS FELT AND AN ABNORMAL SOUND WAS NOTED. ANGIOGRAPHY SHOWED AN ABNORMALITY AT THE TIP OF THE CATHETER. UPON WITHDRAWAL AND INSPECTION, DEFORMATION AND CRACKING OF THE TIP AND COLLECTION CHAMBER WERE OBSERVED, AND THE DEVICE COULD NOT BE USED. THE DAMAGED COMPONENT WAS RETRIEVED FROM THE PATIENT, AND THE PROCEDURE WAS COMPLETED USING AN ALTERNATIVE SURGICAL TECHNIQUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23084195 ↗MalfunctionSep 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A DIRECTED RESECTION PROCEDURE USING THE ATHERECTOMY TH-LX-M TURBOHAWK ATK, A PATIENT PRESENTED WITH A 70% STENOSIS LESION IN THE MID LEFT SUPERFICIAL FEMORAL ARTERY CHARACTERIZED AS CALCIFIED, FIBROUS, AND PLAQUE. MODERATE CALCIFICATION WAS NOTED, AND THE ARTERY DIAMETER WAS 6 MILLIMETERS. NO ISSUES DURING PREPARATION. A 5MM SPIDER FX EMBOLIC PROTECTION DEVICE WAS USED. THE VESSEL WAS PRE-DILATED AND POST-DILATED. DURING THE PROCEDURE, A DAMAGED CUTTER COMPONENT WAS DETECTED; HOWEVER, NO RESISTANCE WAS FELT DURING ADVANCEMENT, AND THE DAMAGED COMPONENT DID NOT REQUIRE RETRIEVAL FROM THE PATIENT. AFTER THE DEVICE WAS REMOVED FROM THE BODY, THE CUSTOMER REPEATEDLY INSPECTED THE DAMAGED DEVICE, AND THE PISTON INSIDE THE COLLECTION CHAMBER FELL OUTSIDE (ON THE GROUND). A REPLACEMENT TURBOHAWK LX-M WAS USED TO COMPLETE THE PROCEDURE, AND THE SURGERY PROCEEDED SMOOTHLY WITHOUT IMPACT TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22935948 ↗MalfunctionAug 8, 2025Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A DIRECTED RESECTION PROCEDURE USING THE TURBOHAWK DEVICE FOR AN 80% STENOSIS WITH PLAQUE IN THE FEMORAL ARTERY, THE HAWK BLADE COULD NOT BE FULLY RETRACTED. AFTER PLACEMENT OF AN EMBOLIC PROTECTION DEVICE AT THE DISTAL END, ROTARY CUTTING WAS PERFORMED, AT WHICH POINT THE INABILITY TO FULLY RETRACT THE BLADE WAS IDENTIFIED. TO AVOID POTENTIAL VASCULAR DAMAGE, THE DEVICE WAS PROMPTLY WITHDRAWN FROM THE BODY. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT AND THERE WAS NO VESSEL DAMAGE. SUBSEQUENTLY, BALLOON DILATATION DRUG COATED BALLOON TREATMENT WAS PERFORMED, AND THE EMBOLIC PROTECTION DEVICE WAS REMOVED. NO PATIENT SYMPTOMS, COMPLICATIONS, ADVERSE EVENTS, OR INJURIES WERE REPORTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. WHEN THE DEVICE WAS RETURNED IT WAS NOTED THAT THE TIP WAS DETACHED AT THE HINGE PIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22747228 ↗MalfunctionApr 22, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT A PHYSICIAN WAS ATTEMPTING TO USE A TURBOHAWK ATK ATHERECTOMY DEVICE FOR TREATMENT IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (POPLITEAL ARTERY) OF A PATIENT WITH CALCIFICATION, TORTUOSITY AND 80% STENOSIS. NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THE DEVICE WAS PREPPED AS PER IFU. IT WAS REPORTED THAT SURGICAL ANGIOGRAPHY REVEALED CALCIFICATION AND SHORT SEGMENT STENOSIS IN THE RIGHT SUPERFICIAL FEMORAL ARTERY, SO THE SURGEON PREPARED TO USE TURBOHAWK ATK FOR ROTATION EXCISION. NO ABNORMALITIES WERE FOUND WHEN FLUSHING THE DEVICE. IT ENTERED THE BODY, PASSED THROUGH THE LESION, AND RETURNED TO THE PROXIMAL END, TURNED ON THE SWITCH FOR ROTATION EXCISION, BUT FOUND THAT IT COULD NOT BE ADVANCED, SO THE DEVICE WAS WITHDRAWN AND OBSERVED THAT THE BLADE WAS DAMAGED, SO THE SURGEON REPLACED IT WITH TURBOHAWK PLUS FOR ROTATION EXCISION AND PERFORMED THE OPERATION NORMALLY. NO PATIENT INJURY REPORTED FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21880226 ↗MalfunctionApr 1, 2025Peeled/Delaminated; Detachment of Device or Device Component; Peeled/Delaminated; Detachment of Device or Device Component
IT WAS REPORTED THAT A PHYSICIAN WAS ATTEMPTING TO USE A TURBOHAWK ATK ATHERECTOMY DEVICE FOR TREATMENT IN A PATIENT IN THE MID/DISTAL POPLITEAL ARTERY WITH TORTUOSITY, PLAQUE AND CALCIFICATION WITH 80% STENOSIS. THE DEVICE PREPPED AS PER IFU. NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. IT WAS REPORTED THAT THE DEVICE WAS UNABLE TO FLUSH TISSUE. A HARD PLASTIC/HARD MASS WAS FOUND IN THE COLLECTION BIN, AND IT MADE IT IMPOSSIBLE TO CLEAN THE DEVICE. THE PROCEDURE WAS COMPLETED BY BALLOON DILATATION, AND A DRUG-COATED STENT WAS IMPLANTED. NO INJURY TO PATIENT REPORTED. WHEN THE DEVICE WAS RETURNED, IT WAS NOTED THAT THE DEVICE WAS DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21737887 ↗InjuryMar 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A DIRECTED RESECTION PROCEDURE INVOLVING THE ATHERECTOMY TH-LX-M TURBOHAWK ATK DEVICE, A DAMAGED COMPONENT WAS NOTED. THE PROCEDURE WAS PERFORMED ON THE PROXIMAL RIGHT SUPERFICIAL FEMORAL ARTERY, WHICH HAD A CALCIFIED TARGET LESION WITH A MODERATE DEGREE OF TORTUOSITY AND CALCIFICATION. DURING THE PROCEDURE, THE CUTTER COMPONENT WAS REPORTED AS DAMAGED. THE DAMAGED COMPONENT NEEDED TO BE RETRIEVED FROM THE PATIENT, WHICH WAS ACCOMPLISHED USING A SUCTION DEVICE. THE PROCEDURE WAS COMPLETED WITH A NEW MEDTRONIC ROTARY CUTTING DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21581632 ↗MalfunctionJan 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
PHYSICIAN ATTEMPTED TO USE A TURBOHAWK ATK ATHERECTOMY DEVICE FOR PATIENT TREATMENT IN THE MID SUPERFICIAL FEMORAL ARTERY WITH PLAQUE, CALCIFICATION AND 70% STENOSIS. NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. A NON-MEDTRONIC SHEATH WAS USED. IFU WAS FOLLOWED. VESSEL WAS PREDILATED IT IS REPORTED THAT THE BLADE OF THE DEVICE COULD NOT ROTATE DURING THE OPERATION, AND AFTER BEING REMOVED OUTSIDE THE BODY, THE BLADE BROKE AND COULD NOT BE USED. THE DEVICE WAS REPLACED WITH A NEW TURBOHAWK DEVICE, AND THE OPERATION WAS COMPLETED. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21111202 ↗MalfunctionMay 20, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
A PHYSICIAN WAS USING A TURBOHAWK ATHERECTOMY DEVICE FOR TREATMENT OF A 10MM CALCIFIED LESION WITH 50% STENOSIS IN THE MID SUPERFICIAL FEMORAL ARTERY. THE ARTERY WAS 5MM IN DIAMETER WITH MILD TORTUOSITY AND MODERATE CALCIFICATION. AN 8FR SHEATH AND A 0.0014 NON MEDTRONIC GUIDEWIRE WAS USED. THE VESSEL WAS PRE-DILATED. THE IFU WAS FOLLOWED. IT WAS REPORTED DURING THE ARTERIAL PLAQUE REMOVAL OF THE LOWER EXTREMITIES, THE GUIDE WIRE WAS USED TO GUIDE IT THROUGH THE LESION AFTER FOLLOWING THE NORMAL PROCEDURES. DURING THE PROCESS OF REMOVING THE PLAQUE, IT WAS JAMMED AND IDLING. AFTER IT WAS WITHDRAWN FROM THE BODY, IT WAS REASSEMBLED AND REMOVED, BUT THE PLAQUE STILL COULD NOT BE EFFECTIVELY REMOVED. THEN IT WAS REPLACED WITH A NEW PRODUCT AND THE SURGERY WAS SUCCESSFULLY COMPLETED. NO PATIENT INJURY. WHEN THE DEVICE WAS RETURNED IT WAS NOTED THAT THE TIP OF THE DEVICE WAS DETACHED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19349860 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.