Skip to content

VarisoftRecalls & reported problems across all models

Unomedical A/S · 8 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

4 open recalls name at least one Varisoft model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,250
Deaths reported
0
Injuries reported
284
Recalls
4

We found 8 models of Varisoft by Unomedical A/S in the FDA's device registry (GUDID). Together they are named in 1,250 adverse event reports in the FDA's MAUDE database, received between Jan 25, 2019 and Sep 30, 2026 — classified as 284 injuries, 966 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (315 reports).

4 FDA recall records name at least one of these models by its device identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 33 reportsJun 2024: 67 reportsJul 2024: 87 reportsAug 2024: 70 reportsSep 2024: 40 reportsOct 2024: 39 reportsNov 2024: 31 reportsDec 2024: 7 reportsJan 2025: 39 reportsFeb 2025: 15 reportsMar 2025: 34 reportsApr 2025: 9 reportsMay 2025: 30 reportsJun 2025: 11 reportsJul 2025: 12 reportsAug 2025: 14 reportsSep 2025: 5 reportsOct 2025: 4 reportsNov 2025: 4 reportsDec 2025: 7 reportsJan 2026: 9 reportsFeb 2026: 4 reportsMar 2026: 49 reportsApr 2026: 123 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 21 reportsAug 2026: 229 reportsSep 2026: 161 reportsOct 2026: 0 reports202420252026229
Last 36 months. The most recent month may be incomplete.

Varisoft recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jul 23, 2024Open, Classified

    Specific VariSoftTM Infusion Sets were packaged using incorrect secondary packaging (boxes) branded "Trusteel Infusion Set".

    Z-3186-2024 · Root cause: Process change control · Recall details · FDA record ↗

  2. Oct 11, 2023Open, Classified

    The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.

    Z-0234-2024 · Root cause: Process change control · Recall details · FDA record ↗

  3. Oct 11, 2023Open, Classified

    The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.

    Z-0233-2024 · Root cause: Process change control · Recall details · FDA record ↗

  4. Oct 11, 2023Open, Classified

    The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.

    Z-0232-2024 · Root cause: Process change control · Recall details · FDA record ↗

Most reported Varisoft problems

Device problems

  • Material Twisted/Bent315
  • Loss of or Failure to Bond188
  • Detachment of Device or Device Component180
  • Premature Separation140
  • Leak/Splash112
  • Infusion or Flow Problem87
  • Adverse Event Without Identified Device or Use Problem74
  • Device Handling Problem72

Patient problems

  • No Clinical Signs, Symptoms or Conditions685
  • Hyperglycemia502
  • Elevated Ketones/ Diabetic Ketoacidosis29
  • Skin Infection10
  • Abscess9
  • Skin Inflammation/ Irritation7
  • No Known Impact Or Consequence To Patient6
  • Hemorrhage/Blood Loss/Bleeding5

Varisoft models

8 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Varisoft adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Unomedical A/S products

All Unomedical A/S devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page