AMS 800 Urinary Control System Control Pump, REF 72400098… recallZ-2669-2011 · American Medical Systems, Inc. · initiated May 9, 2011
Terminated All recalls posted in June 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
American Medical Systems is recalling all unexpired AMS 800 and AMS Acticon Neosphincter Control Pumps. Based upon a review of our product test procedures, we are unable to confirm that all Control Pumps have met our requirements. Test errors could potentially result in a failure to identify a malfunction in the Control Pump activation mechanism. This malfunction may lead to inadvertent activa
FDA root cause category: Nonconforming Material/Component
What the company is doing
American Medical System sent an "URGENT MEDICAL DEVICE RECALL letter dated May 9, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer Customers were instructed to return any existing stock . Please contact American Medical System at 1 (800) 328-3881 to obtain a Return Authorization number. Customer Service will also provide a FedEx shipping number for your use to return affected product. For further questions please contact American Medical Systems at 1- (800)328-3881.
Recall details
- Recall number
- Z-2669-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 9, 2011
- Posted by the FDA
- Jun 27, 2011
- Recalling firm
- American Medical Systems, Inc.
- Quantity
- 15,466
- Product code
- EZY
- FDA submissions
- P000053
- Distribution
- Worldwide Distribution--USA (nationwide) including Washington D.C. and Puerto Rico and the countries FRANCE, GERMANY, CANADA, ISRAEL, SOUTH AFRICA, LEBANON, AUSTRIA, ARGENTINA, SAUDI ARABIA, SPAIN, BELGIUM NETHERLANDS, PERU, VENEZUELA, SRI LANKA, PORTUGAL, TUKEY, BRAZIL AUSTRALIA, CZECH REPUBLIC, TUNISIA. LUXWMBOURG, QATAR, SOLOMON ISLANDS, BOSNIA & HERZEGOVIA, CHILE, CHINA, COLOMBIA, COSTA RICA, DENMARK, ECUADOR, ESTONIA, FINLAND, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRELAND, ITALY, JAPAN, JORDAN, REPUBLIC OF KOREA, MALAYSIA, MEXICO, NEW ZEALAND, NORWAY, PANAMA, POLND, ROMANIA, RUSSIAN FEDERATION, SINGAPORE, SLOVAKIA, SLOVENIA, SWEDEN, SWITZERLAND, TAIWAN PROVIDENCE OF CHINA, THAILAND, UNITED ARAB EMIRATES, UNITED KINGDOM and URUGUAY.
Product description
AMS 800 Urinary Control System Control Pump, REF 72400098, (720172-01 Japan only ), Sterilization Method Steam, American Medical systems Inc., 10700 Bren Road West, Minnetonka, MN 55343 InhibiZone Antibiotic Surface Treatment, AMS 800 Urinary Control Pump with InhibiZone, REF 72404127, Sterilization Method ETO, American Medical Systems Inc., 10700 Bren West, Minnetonka, MN 55343 The AMS Urinary Control System (of which the pump is a component) is used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) following prostate surgery.
Code information (models, lots, serial numbers)
Show the full code information (145k characters)
Lot Serial Number Model 72400098 331544008 364861014 378181008 379515015 379515016 381584018 383424011 384677004 387975018 390279019 391689015 392472008 392802016 394617015 397986004 398690003 399152019 399886014 401696018 401698014 403662016 403773004 404506011 405093015 406168014 408796016 409096007 410461012 410463012 411503008 412906004 412913009 413757006 413757008 413860005 413860006 414275001 414757005 414757006 414760017 415271016 415274002 415275003 415276011 415846005 416339019 416340016 416340019 416342013 416932019 416933018 417674002 417776008 418275016 418277017 418773010 418915006 418916015 419374018 419832017 420062013 420063011 420133008 420656014 420658010 421033012 421602010 421603012 421603017 421680018 422720006 422720009 422729014 422802010 423278001 423278009 423797018 424126016 424182002 424234012 424234018 424743014 424810010 425289018 425403005 426068017 426071017 426282003 426376012 426377008 426782010 427219003 427428007 427428014 427825003 427827005 427827018 427828004 427828010 428257008 428278007 428278008 428278009 428278015 428279015 428495001 428495010 428720008 429139012 429580019 429750008 429750010 429750012 429750018 429751012 429856008 429856011 430013010 430180006 430375008 430376005 430376007 430376008 430376010 430592005 430592015 430592016 430748019 430749020 430855005 430855010 430857008 431130002 431130006 431130011 431263011 431263015 431263019 431264003 431264009 431264012 431264014 431264017 431548001 431902008 432044012 432045013 432045016 432045020 432174004 432174009 432174017 432175007 432175010 432175015 432175017 432177008 432177010 432202011 432202014 432203006 432450004 432450007 432748009 432748015 432829005 432830007 432831001 432831007 433031001 433031005 433031009 433031010 433031017 433070006 433070008 433070010 433070019 433071005 433071009 433071011 433071013 433071019 433139001 433139006 433139007 433139008 433139010 433139011 433139012 433438008 433438009 433438011 433438012 433438017 433489002 433489006 433489009 433505001 433505002 433505004 433505005 433505010 433505015 433505017 433567006 433567008 433567009 433567015 433568001 433568002 433568012 433568017 433568018 433643002 433643007 433643015 433643018 433721002 433721005 433721006 433721012 433920001 433920005 433920006 433920010 433920011 433920016 433920019 433920020 434103006 434103007 434103014 434103019 434103020 434191007 434191016 434295017 434613002 434613004 434614001 434614002 434614005 434614013 434942001 434942003 434942009 434943005 434943017 435297004 435297010 435298002 435298003 435298005 435298011 435298014 435298016 435524004 435524005 435524015 435549006 435549007 435549008 435549009 435549012 435549015 435549017 435549018 435759003 435759004 435759007 435759010 435760002 435760009 435760010 435760013 435769009 435769010 435769012 435770003 435770014 435770015 435939011 435939012 435980007 435980008 435980010 435980012 435980013 435980014 435980015 435980016 435980017 435980018 436001019 436177004 436177005 436177009 436177010 436177013 436177014 436177015 436177016 436177017 436177018 436227005 436227006 436227007 436227011 436227012 436227017 436227019 436258001 436258002 436258004 436258008 436258013 436258018 436259002 436259003 436259005 436259006 436259008 436476003 436476005 436476009 436476013 436476016 436476018 436477002 436477005 436477009 436477015 436616001 436616003 436616004 436616006 436616007 436616009 436616011 436616012 436616014 436617001 436617003 436617006 436617016 436759001 436759002 436759003 436759005 436759006 436759007 436759008 436759009 436759011 436759013 436759014 436759015 436759016 436759017 436759018 436760009 436760010 436760011 436760012 436760013 436760016 436760017 436760019 436895004 436895005 436895006 436895007 436895008 436895009 436895010 436895012 436895013 436895014 436895017 436895018 436895019 436896003 436896006 436896007 436896011 436896016 436896018 436896020 436915001 436915002 436915003 436915011 436915012 436915013 436915014 436915015 436915017 436915019 436916001 436916002 436916003 436916004 436916005 436916006 436916007 436916008 436916009 436916010 436916012 436916013 436916014 436916015 436916016 436916017 437149001 437149002 437149003 437149004 437149006 437149007 437149008 437149010 437149011 437149012 437149013 437149014 437149015 437149017 437149019 437150002 437150003 437150004 437150005 437150006 437150009 437150012 437150014 437150015 437150017 437150018 437150020 437269001 437269002 437269003 437269004 437269005 437269006 437269008 437269013 437269014 437269015 437270001 437270003 437270004 437270005 437270006 437270007 437270008 437270010 437270011 437270013 437270015 437270016 437270018 437270019 437312001 437312002 437312003 437312004 437312007 437312008 437312009 437312010 437312012 437312013 437312014 437312016 437312017 437313001 437313002 437313004 437313005 437313006 437313007 437313008 437313009 437313010 437313012 437313013 437313014 437313019 437432004 437432006 437432008 437432010 437432011 437432012 437432013 437432018 437755001 437755002 437755003 437755004 437755005 437755006 437755007 437755008 437755009 437755010 437755011 437755012 437755013 437755014 437755015 437755016 437755017 437755018 437755019 437755020 437757001 437757002 437757003 437757004 437757005 437757006 437757007 437757008 437757009 437757010 437757015 437824013 437824016 437824017 437824018 437824019 437824020 437825001 437825002 437825003 437825004 437825005 437825006 437825008 437825009 437825010 437825011 437825012 437825013 437825016 437825017 437832001 437832002 437832003 437832004 437832006 437832007 437832010 437832011 437832012 437832014 437832015 437832016 438511003 438511004 438511005 438511006 438511007 438511008 438511009 438511010 438511012 438511013 438511016 438511017 438511018 438511019 438512002 438512003 438512004 438512006 438512007 438512008 438512009 438512010 438512011 438512012 438512013 438512014 438512015 438512016 438512017 438512018 438512019 438512020 438763002 438763006 438763008 438763011 438763013 438763014 438763015 438763016 438763018 438763019 438764001 438764003 438764004 438764005 438764007 438764012 438764013 438764014 438764015 438764016 438764019 438795001 438795002 438795003 438795005 438795006 438795007 438795008 438795009 438795010 438795011 438795012 438795013 438795015 438795016 438795018 438795019 438796001 438796003 438796007 438796008 438796010 438796012 438796015 438796016 438796018 438796019 438945001 438945003 438945004 438945005 438945006 438945007 438945008 438945009 438945010 438945011 438945012 438945013 438945019 438945020 438946003 438946004 438946005 438946007 438946008 438946010 438946011 438946012 438946013 438946014 438946015 438946016 438946017 438946018 438946019 438946020 439150004 439150006 439150008 439150009 439150010 439150011 439150012 439150013 439150015 439150016 439150018 439150019 439150020 439151001 439151002 439151003 439151004 439151005 439151006 439151008 439151009 439151012 439151013 439151014 439151016 439151018 439151019 439151020 439168001 439168002 439168003 439168004 439168005 439168006 439168007 439168008 439168009 439168010 439168012 439168013 439168014 439168015 439168016 439168017 439168018 439169002 439169003 439169004 439169005 439169006 439169007 439169008 439169010 439169011 439169012 439169013 439169015 439169016 439169017 439169018 439169019 439169020 439301002 439301003 439301004 439301005 439301006 439301007 439301009 439301012 439301013 439301014 439301015 439301016 439301017 439302001 439302003 439302004 439302005 439302006 439302007 439302008 439302009 439302010 439302011 439302013 439302014 439302015 439302017 439302018 439302019 439302020 439303001 439303002 439303003 439303004 439303005 439303006 439303007 439303010 439303011 439303012 439303013 439303014 439303015 439303016 439303017 439303018 439303019 439524001 439524002 439524004 439524005 439524006 439524007 439524009 439524010 439524011 439524012 439524013 439524014 439524015 439524016 439524017 439524018 439524020 439525001 439525002 439525003 439525005 439525006 439525008 439525009 439525010 439525011 439525012 439525013 439525014 439525015 439783001 439783002 439783003 439783004 439783005 439783007 439783008 439783010 439783011 439783013 439783014 439783016 439783017 439783018 439784001 439784002 439784004 439784005 439784006 439784008 439784010 439784011 439784012 439784013 439784014 439784015 439784017 439784018 439897001 439897005 439897008 439897009 439897010 439897011 439897013 439897014 439897016 439897017 439897018 439897020 439898001 439898002 439898003 439898004 439898005 439898006 439898008 439898009 439898010 439898011 439898012 439898014 439898015 439898016 439898017 439898018 439898019 440009001 440009002 440009003 440009004 440009005 440009006 440009009 440009012 440009013 440009014 440009016 440009018 440009020 440010001 440010002 440010003 440010004 440010006 440010007 440010008 440010009 440010011 440010012 440010014 440010016 440010017 440010018 440010019 440010020 440151001 440151002 440151003 440151004 440151005 440151006 440151007 440151009 440151010 440151011 440151013 440151014 440151015 440151016 440151017 440151018 440151019 440151020 440152001 440152003 440152006 440152007 440152008 440152009 440152010 440152011 440152012 440152013 440152014 440152015 440152017 440152018 440152019 440152020 440249003 440249004 440249005 440249006 440249007 440249009 440416001 440416002 440416003 440416004 440416005 440416006 440416007 440416008 440416009 440416010 440416012 440416014 440416015 440416017 440416018 440416019 440416020 440417001 440417002 440417004 440417005 440417006 440417009 440417010 440417011 440417013 440417014 440417017 440417018 440417019 440417020 440738001 440738002 440738003 440738004 440738005 440738006 440738007 440738009 440738011 440738012 440738013 440738014 440738015 440738016 440738017 440738018 440753004 440753005 440753006 440753007 440753009 440753010 440753011 440753014 440753015 440753016 440753017 440753018 440753019 440753020 440989001 440989002 440989003 440989004 440989006 440989007 440989008 440989009 440989010 440989011 440989012 440989013 440989014 440989015 440989016 440989017 440989018 440989019 440990001 440990002 440990003 440990004 440990006 440990007 440990008 440990009 440990010 440990013 440990015 440990016 440990018 440990019 440990020 441225001 441225002 441225003 441225004 441225005 441225007 441225008 441225009 441225012 441225013 441225015 441225016 441225017 441225018 441225019 441225020 441226001 441226002 441226004 441226005 441226006 441226008 441226009 441226011 441226015 441226016 441226018 441226019 441226020 441487001 441487002 441487003 441487005 441487006 441487008 441487009 441487010 441487011 441487012 441487013 441487016 441487017 441487018 441487019 441488001 441488002 441488003 441488004 441488006 441488007 441488008 441488009 441488010 441488011 441488015 441488016 441488017 441488018 441488020 441554001 441554002 441554003 441554004 441554005 441554006 441554007 441554008 441554010 441554011 441554013 441554014 441554018 441554019 441555001 441555003 441555004 441555005 441555006 441555007 441555008 441555009 441555010 441555011 441555014 441555015 441555016 441555017 441555018 441555019 441555020 441677004 441677005 441677006 441677007 441677008 441677009 441677010 441677011 441677012 441677013 441677015 441677020 441678002 441678003 441678004 441678005 441678006 441678007 441678008 441678009 441678011 441678012 441678013 441678015 441678016 441678018 441678020 441679001 441679002 441679004 441679005 441679011 441679013 441679016 441679018 441679019 441679020 441946001 441946005 441946006 441946007 441946009 441946010 441946014 441946017 441946020 441947001 441947002 441947004 441947005 441947006 441947007 441947008 441947009 441947010 441947011 441947013 441947015 441947016 441947018 441947019 442162001 442162003 442162005 442162006 442162007 442162008 442162009 442162011 442162012 442162013 442162014 442162015 442162016 442162017 442162018 442162019 442162020 442163001 442163002 442163003 442163004 442163005 442163006 442163007 442163011 442163014 442163015 442163016 442163017 442163018 442163019 442163020 442331002 442331003 442331004 442331005 442331006 442331007 442331009 442331010 442331011 442331013 442331014 442331018 442332001 442332002 442332003 442332004 442332006 442332007 442332008 442332010 442332011 442332014 442332016 442332017 442332018 442332019 442332020 442437008 442437009 442437010 442437011 442437012 442437013 442437014 442437015 442437016 442437017 442437018 442437019 442437020 442750001 442750002 442750003 442750004 442750005 442750006 442750009 442750010 442750011 442750014 442750016 442750017 442750018 442750019 442752001 442752002 442752003 442752004 442752006 442752007 442752011 442752012 442752013 442752014 442752015 442752016 442766003 442766004 442766006 442766007 442766008 442766009 442766010 442766011 442766012 442766015 442766016 442766017 442766018 442766019 442766020 442767001 442767002 442767004 442767006 442767010 442767011 442767012 442767013 442767014 442767015 442767017 442767018 442938003 442938006 442938007 442938011 442989001 442989002 442989005 442989006 442989007 442989008 442989009 442989010 442989011 442989013 442989014 442989015 442989016 442989017 442989019 442989020 442991002 442991003 442991004 442991005 442991006 442991007 442991008 442991009 442991011 442991012 442991014 442991015 442991016 442991018 443297001 443297002 443297004 443297006 443297009 443297010 443297011 443297012 443297013 443297015 443297016 443297017 443297019 443298001 443298002 443298003 443298004 443298005 443298006 443298007 443298008 443298010 443298011 443298013 443298014 443298019 443298020 443525008 443525009 443525013 443526001 443526002 443526004 443526012 443526013 443526015 443526017 443526018 443526019 443843001 443843002 443843003 443843005 443843006 443843009 443843010 443843013 443843014 443843015 443843017 443843020 443844002 443844008 443844009 443844010 443844011 443844016 443844017 443844020 443908001 443908002 443908003 443908006 443908007 443908011 443908012 443908014 443908016 443908017 443908019 443908020 444226002 444226003 444226006 444226007 444226008 444226010 444226011 444226012 444226014 444226018 444226020 444227002 444227004 444227005 444227007 444227008 444227009 444227010 444227011 444227012 444227013 444227014 444227016 444227017 444427001 444427002 444427003 444427006 444427007 444427008 444427009 444427011 444427014 444427019 444508002 444508003 444508004 444508005 444508006 444508008 444508009 444508013 444508014 444508015 444508017 444509002 444509003 444509004 444509006 444509007 444509008 444509010 444509011 444509012 444509013 444509014 444509016 444509017 444509019 444509020 444751001 444751002 444751004 444751007 444751008 444751010 444751013 444751018 444751019 444751020 444752001 444752005 444752007 444752010 444752014 444752015 445114004 445114005 445114007 445114008 445114010 445114011 445114013 445114016 445114017 445114018 445114019 445114020 445115001 445115002 445115006 445115007 445115008 445115011 445115014 445115015 445115016 445115017 445115018 445115019 445115020 445327005 445327007 445327008 445327011 445327012 445327013 445327015 445327016 445328003 445328004 445328005 445328006 445328007 445328009 445328010 445328012 445328013 445328014 445328017 445328019 445534002 445534005 445534006 445534007 445534008 445534009 445534012 445534014 445534015 445534016 445534019 445535002 445535003 445535004 445535005 445535006 445535012 445535015 445535016 445535019 445535020 445563002 445563003 445563005 445563006 445563008 445563009 445563010 445563011 445563013 445563014 445563015 445563016 445563017 445563018 445563019 445563020 445843003 445843007 445843010 445843013 445843015 445843016 445843017 445843020 445844014 446060002 446060007 446060011 446060012 446060014 446060015 446061004 446061005 446061006 446061007 446061016 446180003 446180005 446180006 446180007 446180008 446180009 446180010 446180017 446181001 446181002 446181006 446181008 446181009 446181010 446181011 446181015 446181018 446181019 446295002 446295009 446295011 446295012 446295013 446295014 446295015 446295016 446295017 446295018 446295019 446296002 446296004 446296006 446296007 446296010 446296011 446296014 446296018 446481007 446481017 446481019 446481020 446482001 446482003 446482006 446482008 446482009 446482011 446482016 446482018 446482019 446788001 446788003 446788005 446788008 446788013 446788016 446788017 446788019 446788020 446789006 446789008 446789009 446789011 446789014 446789017 446789019 446864003 446864004 446864005 446864006 446864008 446864014 446941005 446944002 446944013 447231005 447231006 447231007 447231008 447231009 447231015 447231018 447232012 447232018 447254002 447254003 447254007 447254018 447255002 447255005 447255006 447255007 447255011 447378013 447378016 447520002 447520003 447520005 447520006 447520007 447520014 447520019 447521013 447521015 447521019 447834004 447834005 447834012 447834014 447835002 447835004 447835018 448131001 448131002 448131005 448131013 448131019 448132010 448132012 448162002 448162011 448162012 448162013 448361001 448361005 448361014 448361020 448362001 448362010 448362019 448423008 448609002 448609011 448868001 448868002 448868003 448868004 448868005 448868012 448868014 448869011 449031003 449031015 449032003 449032006 449032012 449032015 449032016 449096007 449097005 449097008 449216004 449216009 449216015 449216018 449217020 449622004 449623004 449623011 449623015 449623016 449623018 449624009 449624011 449624013 449624016 449624018 449624020 449625002 449625005 449625010 449625012 449625015 449820002 449820011 449820012 449821007 449821014 449821015 449821016 449824014 449824017 449824018 449825010 449825015 449825016 449825020 450223006 450223017 450223018 450223019 450224001 450224009 450224019 450309003 450309015 450310007 450310008 450310012 450310017 450310018 450480001 450480005 450480013 450480016 450481013 450481014 450676002 450676008 450676010 450677006 450677020 450940001 450940002 450940004 450940006 450940007 450940008 450940009 450940010 450940015 450940016 450940018 450940020 450941002 450941003 450941010 450941012 450941020 451147011 451147019 451251020 451252001 451252020 451526013 451526014 451527001 451527003 451527005 451527015 451527019 451527020 451728004 451728005 451728007 451728010 451728016 451729005 451729006 451963011 451964007 451964014 451964015 451964018 452039002 452039011 452039012 452039013 452039016 452040006 452040008 452085004 452085011 452085013 452085014 452085016 452085018 452086005 452086007 452755011 452755012 452756001 452756004 452756013 452756014 452756016 452924005 452924006 452924012 452924018 452925009 452925016 453251002 453252013 453252015 453346001 453346002 453346012 453347016 453471001 453471004 453471011 453471012 453471020 453472003 453841004 453841007 453841008 453841009 453841013 453841016 453841017 453842014 453875001 453875014 453875018 453876006 453876010 453876016 453876017 453985008 453985011 453986015 453986018 453987009 453987011 453987015 453987019 454357001 454357012 454357013 454357018 454358002 454358017 454590003 454590010 454590011 454590012 454591011 454591012 454591013 454591016 454591020 454671008 454671014 454672003 454672006 454672010 454672016 454715001 454715002 454715006 454715007 454715008 454715009 454715012 454808016 454809001 454810001 454810004 454810007 454811002 454811003 454811004 454811005 454811006 454811007 454811008 454855002 454903001 454903007 454903010 454903014 454903016 454904008 454904013 454904014 454904015 454904016 455293007 455293009 455293015 455294002 455294009 455294020 455539006 455539010 455543005 455543006 455543017 455544003 455544005 455544007 455544009 455912002 455912006 455912009 455912019 455916008 456373002 456373003 456373013 456373014 456373020 456471003 456471013 456471015 456471018 456517001 456517003 456517004 456517005 456517006 456517007 456517008 456517010 456517011 456517019 456520001 456520006 456520007 456520011 456520012 456520013 456520014 456520015 456520017 456520018 456520019 456520020 456792007 456792011 456792015 456793002 456793004 456793006 456793007 456793008 456793010 456793011 456793017 456793019 457099001 457099002 457099003 457099008 457101002 457101010 457101013 457101014 457101015 457101017 457101018 457101020 457260008 457260009 457260017 457261004 457261005 457261013 457261014 457261015 457261016 457261017 457522001 457522013 457523001 457523002 457523018 457637009 457762002 457762003 457762004 457762005 457762012 457762013 457762016 457762019 457878007 457878017 457914001 457914005 458022003 458024001 458026001 458026002 458026005 458026007 458026016 458026017 458327001 458327002 458327003 458327004 458327005 458327006 458327010 458327016 458328011 458328017 458328018 458329004 458329013 458329015 458329016 458329017 458489004 458489005 458489007 458489008 458489009 458489010 458489013 458489014 458489017 458489018 458490012 458569002 458569005 458569014 458569018 458569020 458570004 458570006 458570017 458875004 458875008 458875010 458875013 458875019 458876003 458876005 458876006 458876007 458876008 458876009 458876010 458876014 459234004 459234005 459234006 459234019 459235003 459235004 459235006 459235017 459262006 459262007 459262008 459379001 459379008 459379010 459379014 459379018 459379019 459380002 459380003 459380004 459380005 459380006 459380007 459380008 459380009 459380010 459380016 459380017 459380018 459380019 459380020 459456007 459456013 459759001 459759003 459759005 459759007 459759008 459759009 459759010 459759020 460088005 460088010 460089001 460089015 460171004 460171011 460171015 460171016 460171020 460172004 460172016 460172017 460246015 460247004 460247006 460247007 460247009 460247013 460247014 460247016 460319006 460319008 460319012 460319014 460319015 460319019 460319020 460320009 460320012 460320015 460503019 460746002 460746009 460746012 460746013 460746014 460746016 460746018 460747001 460747005 460747010 460818001 460818003 460819017 460819018 461064002 461064006 461064008 461064010 461064012 461064016 461065010 461065018 461230004 461230010 461233003 461233006 461233007 461233014 461233018 461374003 461374004 461374008 461374011 461374016 461374017 461375008 461375009 461375011 461375013 461770003 461770011 461770013 461770016 461771017 461771020 461772006 461772007 461772008 461772009 461772010 461772016 461773001 461773003 461773004 461773006 461773007 461773012 461773013 461773014 461773015 461773016 461773017 461773018 461773019 461773020 462003001 462003004 462003005 462003011 462004001 462004007 462004011 462054017 462055006 462055013 462055017 462323001 462497001 462497003 462498011 462509005 462509008 462509014 462509017 462510002 462510006 462839006 462839012 462839013 462839014 462839015 462840003 462840008 462840011 462840012 462840013 462840014 462851004 462851009 462851013 462851019 462852006 462852012 463197007 463197018 463197019 463197020 463432015 463433001 463433004 463433007 463517007 463517008 463517009 463517010 463517012 463517018 463517019 463517020 463518005 463518011 463518012 463518014 463518017 463634016 463634018 463635009 463776005 463776006 463776011 463776014 463776015 463776016 463776018 463776019 463777004 463777016 463786004 463787002 463787019 464065003 464065004 464065010 464066002 464066007 464066008 464066009 464066018 464230015 464230016 464285004 464285010 464285013 464285015 464567004 464567005 464567006 464567007 464567008 464567011 464567012 464567019 464568001 464568004 464568009 464568018 464581001 464581002 464581008 464581010 464581015 464582004 464582007 464582010 464582011 464582012 464582013 464582014 464582015 464582019 465050013 465051003 465051004 465051006 465051007 465051011 465153001 465153006 465154008 465154009 465205003 465205005 465205006 465205008 465205017 465205018 465397002 465397003 465397012 465398007 465398008 465398011 465398013 465398016 465398017 465831001 465831004 465831008 465831018 465832004 465832005 465832007 465832008 465832016 465832020 465952018 466145001 466146002 466146005 466146006 466146009 466146010 466146012 466146013 466146015 466146017 466198011 466198014 466198015 466199001 466199008 466199011 466199013 466461007 466461008 466461012 466461017 466461019 466461020 466481005 466481006 466481007 466481008 466481010 466483002 466483003 466483004 466483005 466483006 466483009 466483014 466483019 466483020 466869014 467009003 467009004 467009010 467010002 467010015 467010017 467010020 467216005 467216006 467216011 467216012 467347013 467347019 467347020 467526001 467526003 467657004 467657008 467657016 467657017 467657018 467657019 467660014 467660016 467660017 467660019 467747012 467747013 467748001 467748004 467748006 467748012 468066013 468194012 468194015 468349006 468349013 468349015 468349016 468349018 468496006 468496008 468496009 468496010 468496011 468634003 468634004 468634005 468634007 468634012 468799001 468799003 468799005 468799006 468799008 468799019 468799020 468800008 468800009 468800011 468800013 468800016 468800020 469066002 469066010 469066011 469066012 469182005 469182007 469182008 469182010 469182018 469202001 469202003 469202005 469202006 469202012 469202015 469202017 469202018 469426001 469426012 469534004 469534005 469534020 469575006 469575011 469575014 469575017 469578002 469578005 469578008 469578009 469578016 469578018 469578019 469692003 469692011 469692015 469692016 469692017 469692018 469692019 469692020 469809001 469809007 469809008 469809009 469809010 469809011 469965005 469965010 469965017 470175001 470175010 470175012 470175014 470335008 470335012 470335017 470432001 470432004 470432007 470432011 470432013 470432016 470432017 470492001 470492002 470492003 470492004 470492007 470492009 470492011 470492013 470492015 470615004 470615008 470615009 470615010 470615011 470615016 470856015 470857005 470857013 470857020 470863001 470863004 471025001 471025008 471025017 471040004 471040006 471040007 471040012 471041003 471232011 471336010 471336014 471336015 471336020 471339016 471622004 471622008 471622014 471622018 471797004 471797005 471797010 471797015 471818012 471818020 471819003 471819006 471819017 472015004 472015005 472015006 472015007 472015008 472015009 472015015 472015016 472015018 472203002 472203007 472203014 472203018 472203019 472204004 472204006 472204015 472432002 472434007 472434016 472434019 472434020 472619001 472619002 472619003 472619006 472619008 472619009 472619010 472619011 472619015 472619017 472619018 472619019 472619020 472768006 472768010 472768019 472769002 472769003 472824006 472824007 472826001 472826005 472826006 472826017 472826019 472826020 472981002 472981003 472981004 472981007 472981008 472981015 472981017 472981018 472982001 472982003 472982015 472982019 473073006 473073008 473073009 473073013 473073019 473073020 473547001 473547016 473547020 473547027 473712010 473712012 473712015 473712026 473840006 473840023 473840024 473840027 473840029 473840030 473900002 473900006 473900013 473900019 473900024 474043018 474043019 474134003 474134005 474134006 474134007 474134009 474134011 474134017 474134019 474244001 474244006 474244007 474244015 474244019 474244025 474244027 474244028 474244029 474393016 474393019 474393020 474393021 474393022 474393023 474428004 474428011 474428018 474428024 474428027 474428028 474501008 474501016 474501018 474501019 474571001 474571002 474571003 474571015 474948002 474948003 474948005 474948009 474948010 474948012 474948013 474948015 474948016 474948018 474948020 474948021 474948022 474948023 474948025 474948026 474972013 474972017 474972018 474972029 475096001 475096004 475096012 475099008 475099013 475099018 475099025 475099027 475328001 475328004 475328005 475328007 475328015 475328024 475328026 475328028 475540002 475540003 475540004 475540005 475540006 475540008 475540011 475540012 475540020 475540025 475719001 475719002 475719005 475719007 475719009 475719011 475719012 475719019 475719020 475719022 475719023 475719024 475719026 475754001 475754002 475754003 475754004 475754006 475754014 475754015 475754016 475754018 475754022 475754023 475754024 475754025 475754027 475754028 475956001 475956003 475956004 475956005 475956007 475956010 475956011 475956012 475956021 475956022 475956025 475972001 475972002 475972003 475972005 475972006 475972009 475972010 475972011 475972012 475972023 475972026 475972028 475972029 476174002 476174012 476174017 476174018 476175001 476175006 476175007 476175009 476175010 476704001 476704003 476704006 476704007 476704010 476704012 476704014 476704017 476710004 476710005 476710007 476710009 476710026 476710029 476710030 476721004 476721008 476721011 476721012 476721013 476721015 476721017 476721018 477004003 477004004 477004005 477004006 477004012 477004016 477004020 477004028 477004029 477004030 477315001 477315002 477315003 477315004 477315005 477315006 477315007 477315008 477315009 477315012 477315014 477315019 477322002 477322004 477322006 477322008 477322009 477322010 477322013 477322014 477322015 477322016 477322017 477322019 477618001 477618007 477618011 477618012 477637001 477637002 477637005 477637008 477637011 477637012 477863001 477870005 477870008 477870010 477870013 477983011 477983016 477984001 477984002 477984003 477984010 477984012 478520009 478520017 478520026 478520028 478670019 478670025 478702002 478702003 478702006 478702007 478702008 478702011 478702013 478702014 478702015 478702016 478702017 478702022 478702026 478702027 478702029 478702030 478876004 478876005 478876009 478915007 478915008 478915019 478915029 478915030 478982001 478982013 479100002 479100003 479100014 479100023 479100025 479100027 479126006 479126007 479126008 479126009 479126011 479126012 479126013 479126015 479126016 479126017 479126018 479263013 479263014 479263017 479276010 479276018 479462004 479462009 479462010 479463001 479463002 479463005 479463010 479463015 479551004 479551005 479551006 479551009 479551010 479551011 479551012 479551013 479551014 479551017 479551018 479551019 479551020 479830003 479830005 479830016 479830021 479830023 479830024 479839001 479839002 479839003 479839005 479839006 479839009 479839012 479839013 479839015 479839018 479839021 479839022 479839024 479839030 480008002 480008004 480008005 480008020 480040003 480040005 480040008 480040009 480040024 480167001 480167005 480167006 480167008 480167011 480167013 480167019 480167020 480167021 480167022 480167026 480167028 480167030 480225001 480225002 480225003 480225010 480225011 480225012 480225013 480225015 480225016 480225018 480225020 480225021 480711003 480711004 480711012 480711013 480711014 480711027 480711029 480719003 480719012 480719015 480719016 480719019 480719022 480719029 480841007 480841010 480907003 480907009 480907018 480907019 480907023 481067004 481067013 481270011 481561006 481561008 481561012 481561014 481561015 481561024 481561026 481561028 481561029 481583001 481583017 481583020 481583024 481583025 481583027 481583028 481658003 481658004 481658012 481658019 481658020 481658021 481658022 481658027 481753008 481753016 481753021 481753024 481753025 481753028 481947004 481947005 481947006 481947007 481947010 481947014 481947015 481947016 481947024 481947028 482407001 482407003 482407020 482407027 482607005 482607012 482610006 482610008 482610009 482610011 482610016 482610018 482610021 482610024 482610026 482610028 482730002 482730012 482730023 482730025 482957006 482957007 482957013 483160001 483160005 483160008 483160009 483160010 483160012 483160013 483160016 483160017 483160019 483263004 483263005 483263007 483263009 483263025 483263028 483342005 483342006 483342008 483342019 483432008 483432010 483432011 483432012 483432015 483432019 483432020 483432024 483432026 483432030 483582007 483582009 484053009 484053011 484053027 484053029 484110014 484110016 484110019 484230002 484230003 484230004 484230008 484230009 484230011 484230012 484230014 484230015 484230017 484230018 484230023 484230024 484230025 484230028 484230029 484275001 484275002 484275003 484275004 484275005 484275006 484275007 484275008 484275010 484275013 484374008 484374009 484382001 484382002 484382006 484382010 484615003 484615004 484615007 484615015 484615025 484619002 484619004 484619011 484619016 484619018 484619020 484619021 484619023 484619027 484892001 484892009 484892010 484892016 484892017 484892022 484892028 484915001 484915002 484915005 484915006 484915008 484915012 484915013 484915014 484915015 484915016 484915017 484915020 484915022 484915025 484915026 484915029 485198009 485198018 485198019 485198022 485198023 485198024 485198025 485198027 485198030 485199003 485199004 485199006 485199007 485199010 485199013 485199015 485199019 485261002 485261006 485261007 485261010 485261012 485261016 485261017 485261018 485261019 485261021 485261022 485261027 485534001 485534005 485534006 485534008 485534012 485534014 485534015 485534017 485534018 485534024 485534028 485806009 485806011 485806012 485806014 485856008 485856028 485856029 486040001 486040002 486040003 486040006 486040007 486040010 486040025 486040027 486250005 486250007 486250011 486250014 486250015 486250016 486250017 486250018 486250019 486250020 486250025 486381002 486381004 486381020 486381024 486461003 486616006 486616008 486616017 486616018 486616019 486616020 486616023 486616025 486616028 486802003 486802007 486802009 486802013 486802023 486802027 486802030 486851001 486851003 486851004 486851005 486851006 486851026 487043006 487043008 487043012 487043015 487043020 487043024 487043025 487043028 487143003 487143004 487143008 487143011 487143017 487143019 487143021 487143022 487143026 487143029 487187002 487187003 487187005 487187010 487187011 487187014 487187017 487187018 487187019 487187020 487239001 487239002 487239003 487239004 487239006 487239007 487239008 487239011 487239018 487239019 487239021 487239023 487239024 487239026 487239028 487239029 487239030 487394009 487396004 487396005 487396006 487396008 487396011 487396016 487396017 487396018 487396019 487396020 487538005 487538007 487538010 487538014 487538015 487538017 487538019 487538022 487582002 487582003 487582004 487582010 487582014 487582024 487582025 487582028 487624007 487911014 487911015 487911016 487911017 487911018 487959001 487959002 487959005 487959010 488292009 488292011 488292015 488292017 488292023 488292027 488292029 488292030 488580001 488580016 488580019 488580025 488580028 488580029 488927002 488927003 488927004 488927011 488927012 488927014 488927020 489108002 489108007 489108014 489108023 489108030 489298006 489298014 489298021 489378001 489378003 489378011 489378020 489378025 489378028 489378029 490059005 490059008 490059010 490059011 490059016 490059017 490059018 490060002 490060007 490060008 490060011 490060012 490060017 490060020 490060021 490060029 490060032 490060035 490060036 490060037 490469001 490469002 490469003 490469004 490469016 490482001 490482005 490482006 490482008 490482010 490482018 490829004 490829005 490829007 490829009 490829025 490829028 490829030 490997011 490997012 490997019 490997030 491094004 491094007 491282004 491282006 491282008 491282010 491282012 491282014 491282015 491282016 491282017 491282021 491282022 491282024 491282028 491282029 491428005 491428013 491488004 491488006 491488007 491488008 491488009 491488012 491488013 491488018 491488020 491499002 491499004 491499005 491499014 491499017 491499022 491499026 491499027 491570008 491570015 491570018 491660002 491660012 491660013 491660015 491660016 491731009 491731015 491732007 491732008 491732010 491732014 491829009 491829011 491953001 491953002 491953003 491953005 491953012 491953017 491966004 491966008 491966011 491966012 491966014 492223002 492223005 492223007 492223012 492223017 492223019 492404002 492404003 492404004 492404019 492626002 492626005 492626007 492626010 492626012 492626014 492626017 492647001 492647002 492647003 492647005 492647006 492647007 492647008 492647009 492647010 492647011 492647012 492647013 492647014 492647015 492647016 492647017 492647018 492647019 492648002 492648003 492648005 492648009 492648010 492648013 492844001 492844003 492844005 492844010 492844012 492844015 492845004 492845010 492845016 493073002 493073004 493073016 493073018 493073020 493075006 493217007 493217011 493217016 493217018 493309001 493309003 493309005 493309008 493309009 493309010 493309013 493309017 493605002 493605004 493605005 493605007 493605012 493605013 493605016 493605018 493605019 493605020 493957001 493957003 493957004 493957008 494004017 494004018 494004019 494384006 494384008 494384010 494384019 494384020 494385003 494385004 494385011 494385013 494385014 494385016 494385017 494667002 494667004 494667005 494667013 494667017 494667019 494667020 494668008 494668010 494668014 494668018 494772005 494773011 494773012 494773015 494773017 494773019 495303002 495303003 495303006 495303007 495303012 495303015 495303016 495303017 495304003 495304005 495304008 495304012 495344001 495344003 495344009 495344010 495344011 495360002 495360003 495360008 495360018 495360020 495596002 495596003 495596020 495597001 495597002 495597010 495597017 495597018 495880001 495880011 495880012 495880015 496437008 496437010 496437014 496437017 496437018 496461002 496461003 496461013 496461015 496466001 496466006 496466008 496466010 496466016 496466018 496466019 496466020 496688008 496688017 496698006 496698007 496698009 496698018 496921003 496921009 496921014 496921017 496922002 496922003 496922010 497017006 497017009 497017012 497017015 497017020 497223002 497223020 497590016 497590017 497590018 497600004 497600012 497600014 497600019 497601001 497601007 497601008 497601010 497601013 497601014 497601015 497601016 497601018 497601020 497726017 497726018 497726019 497726020 497788004 497788006 497788007 497788009 497788017 497790001 497790005 497790010 497790012 497790013 497790016 497790017 497791004 497791008 497791010 497791011 497791012 497791014 497791017 498003002 498003007 498003008 498003011 498003012 498003018 498003019 498047002 498047003 498047004 498047007 498047010 498047011 498047013 498047017 498048004 498048008 498048017 498150011 498150017 498150018 498150019 501115004 501115013 501115014 501115015 501115018 501117003 501117005 501117006 501117008 501117011 501117013 501117015 501117016 501117020 501433004 501559004 501559008 501559010 501559015 501560001 501560002 501560005 501560011 501560017 501809002 501809003 501809004 501809009 501809012 501809013 501809014 501809015 501809016 501809019 501810004 501810005 501810006 501810007 501810008 501810009 501810012 501810013 501810019 501976001 501976005 501976008 501976009 501976016 501976019 502193002 502193003 502193009 502295001 502295002 502295003 502295014 502295016 502295017 502552005 502552007 502552011 502552013 502552014 502552017 502552018 502552019 502552020 502662007 502662017 502663001 502663004 502663011 502663015 502663018 502663020 503042001 503042006 503042009 503042010 503042012 503042013 503042015 503043007 503043015 503062002 503062015 503062016 503062018 503062019 503549001 503549003 503549004 503549005 503549009 503549010 503568012 503568014 503622001 503622002 503622005 503622007 503622010 503622011 503622012 503622013 503622014 503622015 503622016 503622017 503622018 503622019 503622020 503795006 503795013 503795016 503795017 503795018 503795019 504673009 504673014 505056001 505056014 505056016 505057003 505057017 505133005 505133008 505133011 505133012 505133018 505164001 505164020 505250003 505250011 505250013 505478002 505478007 505478008 505478009 505478014 505478020 505920001 505920002 505920003 505920006 505920007 505920010 505920011 505920012 505920013 505920015 505920016 505920017 505920018 506169015 506169017 506169018 506224002 506224010 506224015 506224017 506224019 506384008 506384017 507127004 507195001 507195002 507195004 507195007 507195009 507195010 507195011 507252001 507252002 507252003 507252013 507252014 507252015 507252016 507318002 507318010 507318011 507318012 507318014 507318015 507318016 507318017 507386002 507386008 507386009 507386010 507386018 507452001 507452003 507452004 507452006 507452007 507452008 507452017 507452020 507687002 507687003 507687006 507687010 507687014 507693007 507693013 507838002 507838004 507838005 507838011 507838015 507929003 507929004 507929010 507929013 507929016 507929017 507929018 508048001 508048002 508048012 508238001 508238002 508238003 508238008 508532013 508532014 508532015 508532017 508532019 508562001 508562003 508562006 508562007 508562008 508562009 508562010 508562013 508562014 508562015 508562016 508562017 508563001 508563002 508563003 508563011 508563012 508563014 508563015 508563016 508563017 508666004 508666005 508666006 508666017 508812004 508812005 508812008 508812011 508812012 508812017 508812019 508831003 508831004 508831009 508831011 508831019 508831020 509019001 509019007 509019008 509019009 509019010 509019011 509019012 509019016 509042003 509042010 509042012 509090006 509090007 509092002 509092003 509092005 509092008 509193003 509225001 509225004 509225005 509225011 509225012 509225017 509225018 509225020 509500002 509500003 509500007 509741003 509741005 509741020 509775001 509775007 509775011 509775012 509775017 509816001 509816002 509816003 509816004 509816005 509816006 509816007 509816008 509816009 509816010 509816011 509816012 509816013 509816014 509816015 509816016 509816017 509816018 509816019 509816020 509829005 509829006 509829009 509829010 509829019 509978003 509978004 509978015 509978019 510022002 510022007 510022008 510022009 510022010 510022011 510022012 510022014 510022019 510023001 510023004 510023012 510023015 510023018 510023019 510023020 510139001 510139002 510139004 510139005 510139006 510139007 510139008 510139009 510139010 510139011 510139012 510139013 510139014 510139015 510139017 510139018 510139019 510139020 510232002 510232006 510232007 510232014 510232016 510232019 510232020 510513001 510513002 510513010 510513011 510513012 510513013 510513014 510513015 510513016 510513017 510513018 510513019 510513020 510615002 510615004 510615009 510615010 510615012 510615014 510719005 510719006 510719009 510719010 510719011 510719012 510719014 510719015 510719016 510719019 510757005 510757015 510757016 510757017 510757018 510757020 510834001 510834002 510834008 510980001 510980004 510980009 510980010 510980014 510980015 510980016 510980019 510980020 511337001 511337002 511337003 511337004 511337005 511337006 511337007 511337009 511337010 511337011 511337012 511337013 511337014 511337015 511337016 511337017 511337019 511337020 511671004 511671007 511671008 511671010 511671011 511671012 511671015 511671016 511671019 511674011 512131003 512131004 512131012 512131014 512131015 512422007 512422008 512606003 512606012 512606018 512606019 512606020 512645003 512645005 512645006 512645007 512645008 512645010 512645011 512645012 512645013 512645014 512764010 512764013 512764017 512764019 512764020 512932002 512932003 512932010 512932012 512932015 512932018 512932020 513316001 513316004 513316005 513316006 513316007 513316008 513316010 513316014 513316015 513316017 513541001 513541002 513541003 513541005 513541012 513541013 513541017 513648004 513648007 513648008 513648009 513648010 513648011 513648012 513648013 513714001 513714006 513714008 513714009 513714010 513714012 513714013 513714014 513937002 513937003 513937004 513937005 513937008 513937010 513937014 513937018 513937019 514766001 514766003 514766004 514766006 514766008 514766009 514766014 514766016 514766018 514776001 514776006 514776007 514776017 514776018 514892001 514892002 514892004 514892005 514892006 514892008 514892009 514892010 514892011 514892012 514892013 514892014 514892016 514892017 514892020 514919001 514919003 514919004 514919006 514919010 514919011 514919014 514919016 514919019 514919020 514983001 514983002 514983004 514983005 514983008 514983011 514983013 514983019 514983020 515140001 515140002 515140003 515140006 515140007 515140011 515140012 515140019 515192001 515192002 515192003 515192010 515192011 515192012 515192013 515192014 515192015 515192016 515192017 515192018 515192019 515192020 515705001 515705003 515705005 515705006 515705008 515705009 515705011 515705013 515705020 515711009 515711011 515711016 515711018 515711019 515893002 515893006 515893007 515893008 515893012 515893014 515893020 516107001 516107002 516107003 516107012 516107015 516107018 517057001 517057002 517057004 517057009 517057014 517057015 517057016 517057018 517330001 517330002 517330003 517330005 517330018 517408002 517408003 517408007 517408009 517408010 517408013 517408015 517408016 517408017 517408018 517506001 517506007 517506009 517506010 517506011 517506012 517506013 517506016 518243001 518243002 518243008 518243010 518243020 518362004 518362007 518362010 518362011 518362013 518362015 518362017 518362020 518444006 518444007 518444008 518444010 518444011 518444012 518444014 518444015 518501008 518501013 518501014 519226005 519226014 519226019 519319001 519319005 519319006 519319009 519319010 519319012 519319013 519319014 519319015 519384001 519384008 519384014 519384015 519384017 519925010 519925012 520078001 520078016 520231001 520231003 520231004 520231008 520231010 520231011 520231013 520231016 520231019 520281018 520372005 520372008 520372011 520456002 520456004 520456007 520456010 520456015 520499001 520499002 520499003 520499005 520499009 520499015 520499016 520499017 520499019 520640003 520640004 520640006 520640007 520640008 520640009 520640010 520640011 520640012 520640014 520640015 520640017 520733012 520733013 520856004 520856005 520856008 520856020 520929007 520929008 520929009 520929010 520929011 520929012 520929013 521039004 521039006 521039009 521039011 521039012 521039013 521039016 521039017 521039018 521039019 521039020 521109010 521109011 521109016 521109017 521109020 521236004 521236006 521236008 521236009 521236011 521236015 521236017 521359003 521359004 521359006 521359008 521359009 521359010 521359012 521359015 521359018 521359020 521624002 521624004 521624005 521624006 521624007 521624008 521624012 521624016 521842004 521842006 521842007 521842009 521842011 521842014 521842018 521842019 521905004 521905006 521905007 521905008 521905010 521905015 521905016 521905019 521954001 521954002 521954015 522115001 522115003 522115007 522115009 522115010 522115012 522115015 522115017 522935001 522935002 522935003 522935004 522935005 522935007 522935008 522935010 522935015 522935016 522935018 522935019 522976006 522976007 522976008 522976018 522976019 523065004 523065010 523065015 523135002 523135003 523135004 523135005 523135011 523135012 523135013 523135014 523135015 523135016 523135017 523135018 523135019 523135020 523296002 523296004 523296006 523296014 523296015 523296016 523296018 523296019 523296020 523367003 523367007 523367015 523367020 523485001 523485002 523485007 523485009 523485011 523485012 523485013 523485016 523485017 523658001 523658005 523658009 523658017 523693001 523693009 523693010 523693012 523693015 523693016 523693018 523782001 523782003 523782006 523782007 523782008 523782016 523782017 523871001 523871002 523871003 523871004 523871011 523871018 523871020 523924001 523924002 523924010 523924013 523924014 523924016 523924017 523924018 523924019 523924020 524217007 524217013 524217018 524235005 524235009 524235010 524235018 524235019 524334009 524334010 524334013 524334014 524334016 524334017 524334018 524415002 524415013 524415017 524415018 524415019 524520005 524520007 524520008 524520014 524520017 524520018 524560004 524560009 524560011 524560014 524560017 524560018 524586001 524586003 524586011 524586012 524586013 524586014 524586019 524650007 524650011 524650013 524650016 524650017 524650018 524650020 524719002 524719003 524719005 524719006 524719007 524719008 524719009 524719010 524719011 524719012 524719013 524719014 524719016 524782001 524782002 524782004 524782005 524782006 524782007 524782008 524782010 524782011 524782013 524782014 524782015 524782016 524782018 524854002 524854003 524854005 524854007 524854009 524854010 524854011 524854016 524854018 524854019 528180002 528180004 528180005 528180006 528180007 528180008 528180009 528180010 528180011 528180013 528180014 528180017 528181001 528181002 528181004 528181005 528181006 528181008 528181009 528181011 528181012 528181016 528181017 528181018 528181020 529634001 529634002 529634007 529634008 529634009 529634011 529634013 529634015 529634016 529975001 529975002 529975003 529975004 529975005 529975006 529975007 529975008 529975009 529975010 529975011 529975012 529975013 529975014 529975015 529975016 529975017 529975018 529975019 529975020 530057001 530057004 530057005 530057006 530057007 530057008 530057009 530057010 530057011 530057012 530057013 530057014 530057015 530057016 530057017 530057018 530202001 530202002 530202003 530202004 530202005 530202006 530202007 530202008 530202009 530202010 530202012 530202013 530202014 530202016 530202017 530202019 530277002 530277005 530277006 530277012 530277016 530277018 530277020 530852002 530852003 530852004 530852005 530852006 530852007 530852008 530852009 530852010 530852012 530852013 530852014 530852015 530852016 530852017 530852018 530852020 530917007 530917008 530917012 530917017 530917018 530917020 531046020 531105001 531105002 531105007 531105018 531251009 531251011 531251012 531337001 531337002 531337003 531337004 531337008 531337011 531337013 531337015 531337016 531337019 531353004 531353015 531353017 531353018 531374002 531374003 531374008 531374011 531374015 531374016 531374017 531374018 531374019 531463003 531463005 531463006 531463007 531463008 531463009 531463010 531463011 531463012 531463013 531463015 531463016 531466001 531466003 531466005 531466008 531466010 531466011 531466012 531466015 531466019 531558001 531558009 531558013 531558014 531558015 531558016 531613001 531613002 531613003 531613004 531613009 531613013 531613015 531613016 531613017 531613018 531613020 531643001 531643004 531643006 531643010 531643011 531643013 531643015 531643016 531643018 531776003 531776004 531776005 531776006 531776011 531776012 531776013 531776020 531851004 531851005 531851008 531907008 531907015 531907016 531907017 531907020 531981009 531981010 531981011 531981013 531981014 531981018 532039014 532304002 532304004 532304005 532304007 532304008 532304011 532304012 532304013 532304014 532304015 532304016 532350001 532350004 532350005 532350006 532350007 532350010 532350014 532446010 532446011 532446012 532547002 532547006 532547007 532547008 532547009 532547011 532547013 532547014 532547015 532547017 532547018 532641001 532641003 532641005 532641009 532641010 532641013 532641014 532641015 532641017 532641020 532722001 532722003 532722004 532722006 532722008 532722009 532722010 532722011 532722013 532722014 532773001 532773003 532773006 532773008 532773010 532773011 532773012 532773013 532773015 532773018 532773019 532875003 532875006 532875009 532933002 532933003 532933005 532933008 532933009 532933010 532933011 532933012 532933013 532933014 533051006 533051007 533051009 533051012 533051015 533051016 533051017 533051018 533051019 533051020 533165001 533165004 533165010 533165011 533165012 533165017 533165018 533165019 533261007 533261014 533376003 533376004 533376005 533376012 533436001 533436003 533436004 533436005 533436006 533436007 533436013 533436016 533436017 533436019 533557001 533557002 533557005 533557007 533557008 533557010 533557011 533557014 533610005 533682003 533682006 533682007 533682008 533682009 533682011 533682013 533682014 533682015 533682016 533944001 533944006 533944011 533944013 533944014 534269002 534269005 534269008 534269009 534269012 534269018 534269019 534477002 534477007 534477008 534477009 534477010 534477013 534477016 534477017 534477018 534477019 534477020 534559002 534559003 534559004 534559006 534559008 534559009 534559013 534559014 534559015 534559017 534559018 534591003 534633001 534633002 534633003 534633004 534633005 534633008 534633009 534633011 534633017 534725001 534725002 534725003 534725010 534725012 534725019 534746002 534746003 534746004 534746008 534746010 534746011 534746014 534746019 535132002 535211005 535211006 535211009 535211015 535211017 535211018 535301002 535301005 535301007 535301008 535301010 535301011 535301012 535301017 535373002 535373003 536005004 536005010 536005012 536005019 536172004 536172007 536172009 536172010 536172014 536172016 536252001 536252010 536252011 536252014 536252016 536252020 536767001 536767002 536767003 536855008 536855010 536855013 536951001 536951002 536951005 536951006 536951009 536951010 536951011 536951015 536951020 537016004 537016006 537016017 537072006 537072010 537072011 537072012 537072013 537072015 537072016 537072017 537072019 537072020 537346001 537346003 537346004 537346007 537346008 537346014 537350001 537350004 537350005 537350006 537350013 537350014 537350016 537616001 537616002 537616003 537616004 537616009 537616014 537616015 537616016 537616017 537616019 537707003 537707006 537707010 537707011 537707013 537707016 537707019 537794003 537794008 537794014 537794015 537848001 537848002 537848003 537848004 537848005 537848006 537848008 537848015 537848018 537932001 537932002 537932004 537932005 537932006 537932016 537932017 537964001 537964002 537964004 537964005 537964006 537964007 537964013 537964014 537964015 537964017 538481005 538481013 538481015 538481017 538481019 538610001 538610002 538610003 538610006 538610007 538610010 538610011 538610012 538610013 538610014 538610015 538610016 538610017 538773003 538773014 538773018 538773019 539362004 539362005 539362008 539362009 539362010 539362013 539362016 539362019 539362020 539510007 539510011 539510012 539510013 539510015 539510016 539510018 539510019 539511001 539511003 539511004 539511005 539511006 539511009 539511010 539511012 539511013 539511014 539511015 539511017 539511018 540116005 540116008 540116010 540116012 540116013 540116020 540117001 540117005 540117006 540117012 540117013 540117015 540117016 540117018 540117019 540263002 540263003 540263004 540263005 540263007 540263008 540263009 540263010 540263012 540263014 540263017 540263019 540263020 540446001 540446003 540446004 540446005 540446007 540446008 540446009 540446010 540446011 540446014 540446015 540446016 540446017 540773001 540773002 540773004 540773005 540773007 540773008 540773009 540773010 540773014 541026002 541026009 541026013 541026014 541026015 541026016 541108002 541108003 541108004 541108007 541108011 541108015 541108017 541108020 541138001 541138004 541138006 541138007 541138012 541138017 541138018 541138019 541397002 541397003 541397005 541397006 541397008 541397011 541397012 541397017 541398001 541398002 541398003 541398006 541398014 541398015 541398016 541496008 541496010 541496012 541496016 541601005 541601006 541601007 541601011 541601013 541601015 541601017 541601018 541667002 541667003 541667004 541667015 541667016 541667018 541786001 541786007 541786008 541786010 541786015 541786017 541786018 541786019 541842002 541842003 541842007 541842008 541842011 541842014 541842015 541842016 541842018 541842020 541989001 541989002 541989004 541989006 541989008 541989013 541989014 541989015 542060002 542060003 542060004 542060006 542060007 542060008 542060009 542060010 542060011 542060012 542060014 542060017 542336001 542336012 542336013 542336014 542336018 542451001 542451005 542451006 542451008 542451009 542451016 542523001 542523003 542523004 542523005 542523008 542523009 542523010 542523012 542523013 542523015 542523017 542523018 542523019 542623001 542623005 542623007 542623009 542623011 542623012 542623013 542623014 542623015 542623017 542708001 542708005 542708010 542708012 542708014 542708016 542783001 542783002 542783004 542783008 542783012 542883001 542883003 542883006 542883008 542883009 542883010 542883011 542883014 542883016 542883017 542883018 542987011 542987012 542987013 542987017 542987018 542987020 543189005 543189008 543189010 543189011 543230001 543230003 543230008 543230012 543230013 543230014 543230015 543268004 543268005 543268007 543268014 543268015 543268016 543268020 543322001 543322016 543322018 543322020 543438002 543438003 543438004 543438012 543438015 543600002 543600003 543600013 543729001 543729008 543729009 543729010 543845005 544124002 544124010 544124012 544124014 544124017 544124018 544124019 544124020 544194003 544194004 544194005 544194008 544194009 544194011 544194012 544194014 544194015 544194016 544194019 544194020 544326002 544326004 544326005 544326011 544326012 544326013 544326014 544326015 544326016 544326017 544326018 544326019 544326020 544429004 544429008 544429009 544429014 544429015 544429018 544520002 544520004 544520005 544520011 544520012 544520014 544520017 544571002 544571003 544571005 544571006 544571008 544571009 544571011 544571016 544571019 544571020 544629001 544629002 544629003 544629005 544629012 544629014 544629015 544629017 544629018 544671003 544671004 544671006 544671007 544671015 544671016 544671017 544671019 544738007 544738009 544738011 544738012 544738014 544738015 544738016 544738019 544967001 544967006 544967010 544967011 544967012 544967013 544967014 544967015 544967016 544967019 544967020 545101001 545101002 545101003 545101004 545101005 545101006 545101008 545101009 545101010 545101011 545101012 545101013 545101015 545101016 545101017 545101018 545101019 545219003 545219004 545219008 545219009 545219010 545219011 545219013 545219014 545219018 545219020 545300001 545300002 545300010 545300011 545300015 545469003 545469004 545469005 545469008 545469009 545469011 545469012 545469014 545469015 545469017 545612004 545612005 545612006 545612011 545612012 545612014 545612015 545612017 545612018 546307001 546307002 546307003 546307006 546307009 546307010 546307011 546307012 546307016 546307017 546307018 546307019 546307020 546448001 546448006 546448008 546448009 546448017 546448019 546496001 546496004 546496006 546496007 546496009 546496010 546496017 546496018 547224005 547224006 547224009 547224010 547224013 547224014 547224016 547224018 547224019 547368001 547368002 547368005 547368008 547368009 547368010 547368011 547368012 547368016 547368018 547368020 549923003 549923004 549923005 549923006 549923008 549923010 549923014 549923015 549923016 549923017 549923018 549923020 549966013 549966014 549966017 549966019 549966020 549967003 549967004 549967005 549967009 549967011 549967013 549967015 549967018 549967019 550642001 550642007 550642009 550642014 550642015 550642016 550642019 550767004 550767006 550767007 550767008 550767009 550767017 550767020 550824001 550824003 550824005 550824006 550824009 550824011 550824012 550824013 550824014 550824020 550926002 550926003 550926007 550926009 550926011 550926013 550926020 551629003 551629005 551776001 551776002 551776006 551776012 551776013 551776014 551776015 551776016 551776017 551776018 551776019 551777002 551777003 551777005 551777008 551777018 551886001 551886004 551886009 551886010 551886011 551886015 551886016 551886017 551886018 552015014 552015015 552113002 552113003 552113008 552113009 552113010 552113015 552113016 552113018 552113019 552118002 552118003 552118004 552118005 552118006 552118007 552118009 552118011 552118012 552118017 552364001 552364006 552364009 552364010 552364011 552364013 552364015 552364017 552364019 552419002 552419003 552419009 552419010 552419020 552534001 552534005 552534008 552534010 552534011 552534012 552534014 552534018 552534019 552534020 552535005 552535007 552535008 552535009 552535011 552535015 552535020 552819002 552819005 552819007 552819008 552819010 552819017 552819020 552944001 552944004 552944005 552944006 552944011 552944015 552944017 552944018 553218001 553218002 553218003 553218004 553218005 553218007 553218008 553218010 553218011 553218015 553218016 553218017 553218020 553327001 553327002 553327005 553327006 553327007 553327008 553327009 553327012 553327013 553327014 553327018 553457003 553457005 553457007 553457008 553457009 553457011 553457013 553457014 553457015 553457016 553457017 553457018 553528001 553528002 553528003 553528005 553528007 553528009 553528016 553528019 553718002 553718016 553718017 553768002 553768003 553768004 553768005 553768006 553768009 553768010 553768012 553768013 553768014 553768016 553768017 553768020 553851001 553851002 553851003 553851004 553851006 553851009 553851011 553851014 553851016 553851017 553851018 553851019 554039002 554039003 554039006 554039009 554039017 554039020 554457002 554457003 554457011 554457016 554457017 554457019 554566001 554566007 554566011 554566013 554566014 554566015 554566017 554566018 554566019 554698016 554698018 554698019 555217001 555217003 555217006 555217007 555217008 555217010 555398007 555398009 555398010 555398011 555398012 555398016 555398018 555398019 555513002 555513003 555513004 555513005 555513006 555513011 555513012 555513013 555671001 555671008 555671010 555671011 555671012 555671013 555671014 555671015 555671017 555671018 555671019 555773002 555773003 555773004 555773005 555773019 556259005 556259006 556259007 556259009 556259010 556259011 556259012 556259015 556259016 556259019 556259020 556408001 556408002 556408003 556408004 556408006 556408007 556408008 556408009 556408010 556408012 556408013 556408014 556408017 556444001 556444002 556444003 556444004 556444005 556444006 556444007 556444008 556444009 556444010 556444011 556444012 556444013 556444014 556444017 556444018 556522001 556522002 556522003 556522005 556522006 556522008 556522014 556522015 556522016 556522017 556603002 556603003 556603004 556603005 556603006 556603007 556603008 556603009 556603010 556603011 556603012 556603013 556603014 556603015 556603016 556603017 556603018 556603020 557082002 557082006 557082008 557082010 557082016 557190004 557190014 557190018 557241002 557241003 557241006 557241008 557241013 557241015 557241016 557241017 557241018 557349003 557349005 557349012 557349020 557396003 557396005 557396009 557396010 557396012 557396013 557396018 557396019 557396020 557471003 557471005 557471009 557471010 557471014 557471015 557471016 557542003 557542009 557542013 557542014 557542015 557542017 557542018 558152001 558152003 558152005 558152007 558152008 558152009 558152010 558152014 558152015 558152018 558152020 558239001 558239002 558239003 558239004 558239005 558239006 558239007 558239009 558239010 558239011 558239014 558239016 558239018 558518002 558518004 558518009 558610005 558610009 558610010 558610011 558610012 558610013 558610015 558610016 558657002 558657003 558657009 559311005 559311009 559311010 559311015 559311018 559395003 559395004 559395005 559395006 559395007 559395008 559395011 559395015 559456002 559456004 559456005 559456007 559456010 559456015 559456016 559456017 559456018 559533002 559533006 559533007 559533008 559533009 559533011 559533015 559533016 559533017 559558001 559558003 559558004 559558005 559558007 559558009 559558010 559558011 559558014 559558015 559558017 559558020 560424003 560424007 560424013 560424016 560424017 560424018 560424020 560504001 560504015 560504016 560504020 560835003 560835004 560835007 560835010 560835018 560835019 561231002 561231004 561231007 561231008 561231009 561231011 561231012 561231014 561231015 561231016 561231019 561380002 561380003 561380006 561380007 561380009 561380013 561380016 561380020 561507004 561507012 561507016 563151003 563151006 563207001 563207003 563207004 563207006 563207007 563207010 563207012 563207013 563207014 563207015 563207017 563232002 563232003 563232004 563232009 563232012 563232013 563232014 563232015 563232017 563232019 563311002 563311003 563311005 563311007 563311010 563311011 563913002 563913003 563913004 563913011 563913012 563913013 563913015 563913016 564113002 564113004 564113005 564113007 564113010 564113011 564113012 564113015 564116004 564116005 564116006 564116009 564116011 564116012 564116017 564116019 564116020 564285004 564285006 564285010 564285014 564285015 564285016 564285017 564871009 564871010 564871011 564871012 564871015 564871017 564964004 564964005 564964009 564964010 564964011 564964012 564964013 564964015 564964018 565018001 565018002 565018008 565018009 565018013 565018015 565018016 565018018 565018019 565018020 565200001 565200005 565200006 565200007 565200008 565200010 565200011 565200012 565200013 565200014 565200016 565913014 565913015 565913018 566135007 566135012 566135015 566135016 566135017 566135018 566135019 566352001 566352002 566352005 566352006 566352012 566352016 566352017 566352019 566352020 566464001 566464002 566464003 566464004 566464007 566464008 566464009 566464011 566464014 566464018 567086005 567086006 567086019 567234002 567234005 567234007 567234011 567234014 567234015 567234018 567309001 567309002 567309005 567309008 567309010 567309011 567309012 567309015 567309016 567309018 567309019 567386003 567386007 567386008 567386009 568037002 568037003 568037004 568037006 568037017 568037018 568037019 568104005 568104006 568104007 568104008 568104009 568104010 568104011 568104012 568104013 568104014 568104016 568224003 568224004 568224012 568224015 568224017 568224018 568224019 568439002 568439003 568439013 568439014 568439015 568439017 568439018 568439020 568520003 568520012 568520016 568520018 568520019 568548004 568548006 568548007 568548012 568548013 568548019 569116001 569116003 569116005 569116006 569116007 569116008 569116009 569116010 569116013 569116014 569116015 569116016 569116017 569116019 569116020 569337002 569337003 569337004 569337010 569337014 569337016 569337017 569337018 569337019 569398001 569398003 569398012 569398014 569398015 569398017 569522002 569522003 569522004 569522006 569522008 569522009 569522011 569522012 569522014 569522015 569522017 569522018 569522019 569833001 569833002 569833004 569833008 569833010 569833011 569833017 569996001 569996009 569996011 569996015 569996016 569996017 569996019 569996020 570348001 570348002 570348003 570348004 570348005 570348006 570348009 570348012 570348016 570348017 570348018 570348020 570394005 570394006 570394007 570394008 570394009 570394010 570394011 570394012 570394015 570394016 570394017 570394018 570493003 570493004 570493007 570493011 570493012 570493014 570493016 570493018 570545008 570545010 570545014 570545016 570545020 570660002 570660003 570660006 570660010 570660012 570660014 570660017 570660018 570660019 571536004 571536005 571536006 571536008 571536009 571536010 571536012 571536016 571536018 571536020 571821004 571821005 571821007 571821008 571821009 571821011 571821012 571821015 571821016 571821017 571821018 571821019 571921001 571921002 571921004 571921006 571921007 571921010 571921011 571921012 571921013 571921017 571922001 572438009 572438017 572438018 572438019 572438020 572439001 572439002 572439006 572439008 572439011 572439012 572439013 572439014 572439015 572439019 572482001 572482005 572482006 572482011 572482012 572482015 572482020 572536005 572536006 572536007 572536010 572536012 572536013 572536018 572537001 572537002 572537005 572537007 572537008 572537009 572537014 572537015 572537017 572537019 572604010 572604012 572604017 572690001 572690002 572690005 572690006 572690007 572690010 572690011 572690014 572690016 572690017 572690019 573311002 573311004 573311005 573311006 573311007 573311008 573311010 573311017 573311019 573415005 573415006 573415007 573415009 573415010 573415011 573415012 573415018 573415019 573415020 573518001 573518002 573518009 573518015 573518016 573518017 573559004 573559008 573559010 573559011 573559015 573559016 573559017 573559018 573559020 573626005 573626006 573626009 573626010 573626018 574463009 574463012 574463015 574463017 574524001 574524002 574524003 574524010 574524011 574524015 574524020 574661001 574661003 574661005 574661006 574661008 574661011 574661015 574661016 574661020 576134001 576134002 576134003 576134009 576134010 576134011 576134013 576134017 576274002 576274007 576274008 576274009 576274010 576274011 576274012 576274013 576274014 576274015 576274016 576274017 576274018 576274019 576317001 576317007 576317008 576317011 576317012 576317015 576317016 576317018 577100007 577100016 577100017 577100020 577121008 577121013 577121017 577218001 577218002 577218005 577218006 577218007 577218008 577218011 577218012 577218013 577218014 577218015 577218016 577218018 577218019 577218020 577350001 577350002 577350003 577350005 577350006 577350009 577350010 577350012 577350014 577350016 577350017 577463001 577463002 577463003 577463004 577463005 577463007 577463008 577463009 577463010 577463011 577463012 577463013 577463015 577463016 577463018 577463019 577463020 578013007 578013009 578013010 578013012 578013013 578013015 578013016 578013019 578091005 578091009 578091011 578091012 578091014 578091017 578185001 578185002 578185003 578185004 578185005 578185007 578185013 578185015 578185016 578185018 578285002 578285007 578285014 578285015 578285017 578389003 578389007 578389008 578389012 578389014 578389015 578389016 578604001 578604003 578604004 578604006 578604007 578604010 578604013 578604015 578604016 578604017 578604018 578604019 578694002 578694003 578694004 578694005 578694006 578694007 578694009 578694011 578694012 578694014 578694015 578694017 579303003 579303004 579303006 579303007 579303008 579303010 579303012 579303016 579303017 579303018 579303019 579334002 579334008 579334010 579334011 579334015 579334019 579445005 579445006 579445007 579445011 579445013 579445014 579445015 579445016 579445017 579445018 579445019 579522002 579522012 579522019 579522020 579644001 579644002 579644004 579644006 579644007 579644010 579644011 579644012 579644013 579644014 579644015 579644017 579644020 579685002 579685003 579685004 579685006 579685007 579685008 579685009 579685010 579685011 579685015 579685017 579685019 579685020 579816001 579816002 579816003 579816004 579816009 579816010 579816015 579816017 579816018 579913001 579913002 579913003 579913005 579913006 579913007 579913015 579913019 580339002 580339004 580339005 580339006 580339007 580339008 580339011 580339012 580339014 580479001 580479002 580479003 580479006 580479007 580479009 580479011 580479012 580479013 580479014 580479017 580479018 580479019 580653001 580653002 580653003 580653004 580653007 580653012 580653013 580653015 580653017 580653018 580653019 581534002 581534011 581534014 581534015 581534017 581534018 581641001 581641002 581641007 581641011 581641012 581641013 581641014 581641015 581641020 581644001 581644002 581644003 581644005 581644007 581644009 581644010 581644013 581644016 581644017 581646003 581646005 581646007 581646009 581646011 581646015 581646016 581646018 581646019 581646020 581647002 581647003 581647004 581647006 581647009 581647011 581647012 581647013 581647014 581647015 581647017 581842003 581842005 581842012 581842013 581842014 581842015 581842016 581842017 581842018 581842019 581842020 581995002 581995003 581995005 581995007 581995010 581995012 581995013 581995014 581995015 581995016 581996002 581996003 581996005 581996006 581996007 581996008 581996014 581996016 581996017 581996019 581996020 581997001 581997009 581997012 581997016 581998002 581998005 581998007 581998010 581998011 581998012 581998018 581998019 581998020 581999005 581999009 581999010 581999011 581999012 581999013 581999014 581999015 581999016 581999017 581999019 582000002 582000004 582000005 582000007 582000008 582000009 582000010 582000012 582000013 582000017 582000018 582000020 582001005 582001009 582001016 582002001 582002002 582002003 582002004 582002005 582002006 582002007 582002009 582002011 582002012 582002015 583060001 583060005 583060009 583060011 583060012 583060014 583060015 583060016 583060017 583060018 583060019 583060020 583061003 583061011 583061012 583061014 583061016 583061020 583062006 583062007 583062010 583062012 583062015 583062017 583062020 583063003 583063005 583063006 583063007 583063011 583063012 583063014 583064001 583064005 583064007 583064013 583065003 583065005 583065006 583065009 583065010 583065013 583065014 583065015 583065017 583066004 583066005 583066006 583066010 583066014 584059002 584059005 584059007 584059008 584059011 584059012 584059013 584059014 584059015 584059017 584059019 584060003 584060004 584060012 584060014 584061010 584061017 584062004 584062008 584062009 584062010 584062011 584062014 584062015 584062016 584062018 584062019 584062020 584063003 584063008 584063010 584063012 584063013 584063019 584063020 584064001 584064002 584064003 584064005 584064006 584064008 584064009 584064010 584064011 584064014 584064015 584064016 584064017 584064020 584065001 584065002 584065008 584065011 584065016 584065018 585097011 585097012 585097013 585097018 585097019 585097020 585098014 585098015 585098016 585098017 585098018 585098019 585099002 585099003 585099004 585099005 585099006 585099008 585099009 585099011 585100001 585100002 585100004 585100008 585100009 585100013 585100014 585100015 585100019 585100020 585102004 585102007 585102009 585102014 585102015 585102018 585850001 585850002 585850003 585850005 585850011 585850012 585850013 585850014 585850015 585850016 585850017 585852001 585852003 585852004 585852005 585852006 585852007 585852009 585852010 585852013 585852014 585852015 585852016 585852017 585852018 585852019 585852020 585853002 585853010 585853014 585853016 585854005 585854006 585854008 585854016 585854017 585854018 585854020 586213003 586213004 586213005 586213006 586213007 586213008 586213011 586213019 586214001 586214002 586214003 586214004 586214005 586214006 586214012 586214013 586214014 586214020 586215002 586215003 586215007 586215009 586215016 586215017 586215018 586215019 586215020 586216004 586216009 586216011 586216013 586216018 586216019 586216020 587057002 587057005 587057006 587057007 587057010 587057016 587057019 587058005 587058006 587058007 587058010 587058011 587058012 587058014 587059001 587059002 587059006 587059007 587059009 587059010 587059018 587059020 587993003 587993008 587993011 587993014 587993017 587993019 587994002 587994005 587994007 587994009 587994014 587994015 587994016 587994018 587994019 587994020 587995001 587995013 587995014 587995017 587995018 587995019 587995020 587996001 587996002 587996003 587996004 587996008 587996009 587996010 587996011 587996012 587996013 587996014 587996015 587996017 587996019 587996020 587997001 587997009 587997010 587997012 587997016 587997018 589122001 589122002 589122003 589122004 589122006 589122007 589122008 589122011 589122013 589122014 589122017 589122019 589123003 589124002 589124003 589124004 589124005 589124007 589124009 589124010 589124011 589124014 589124015 589125001 589125008 589125009 589125011 589125017 589125018 589126003 589126004 589126005 589126007 589126008 589126009 589126013 589126015 589126018 589126019 590150001 590150003 590150005 590150008 590150010 590150011 590150012 590150017 590151001 590151004 590151006 590151007 590151008 590151009 590151010 590151011 590151016 590152001 590152002 590152003 590152004 590152005 590152006 590152011 590152012 590152015 590152016 590153002 590153003 590153005 590153006 590153010 590153011 590153016 590154001 590154002 590154003 590154005 590154011 590154014 590154018 590155002 590155003 590155007 590155008 590155011 590155013 590155014 590155015 590155016 590155017 590155018 590155019 590156006 590156008 590156010 590156011 590156013 590156014 590156016 590156017 591315004 591315009 591315011 591315015 591315016 591315017 591316003 591316004 591316006 591316010 591316012 591316014 591316016 591316017 591316018 591317005 591317006 591317007 591317009 591317011 591317012 591317013 591317015 591317016 591317020 591318016 591318018 591319007 591319009 591319018 592091002 592091005 592091006 592091007 592091010 592091011 592091013 592091014 592091017 592092003 592092004 592092011 592092016 592092017 592092018 592093001 592093002 592093003 592093004 592093005 592093006 592093007 592093009 592093012 592093015 592093016 592094002 592094004 592094013 592094018 592095002 592095003 592095004 592095006 592095007 592095008 592095009 592095010 592095011 592095013 592095015 592095018 592095019 592524002 592524003 592524006 592524008 592524010 592524011 592524013 592524014 592524015 592524016 593320003 593320006 593320010 593320013 593320014 593321002 593321005 593321008 593321012 593321014 593321015 593321016 593321018 593321019 593321020 593322002 593322003 593322004 593322005 593322007 593322010 593322011 593322015 593322016 593322019 593322020 593323003 593323006 593323007 593323011 593323013 593323018 593323019 593323020 593324005 593324010 593324013 593324015 593324016 594309001 594309002 594309003 594309005 594309007 594309010 594309011 594309012 594309015 594310001 594310002 594310003 594310005 594310006 594310008 594310009 594310010 594310011 594310012 594310013 594310015 594310016 594310017 594310018 594310019 594311003 594311005 594311007 594311013 594311018 594311019 594311020 595404001 595404002 595404004 595404005 595404006 595404009 595404010 595404011 595404012 595404013 595404014 595404015 595404019 595405006 595405007 595405011 595405012 595405015 595405016 595405018 595405019 595406002 595406003 595406013 595406015 595406018 595406019 595406020 595407003 595407013 596399016 596399017 596399018 596399020 596400001 596400003 596400004 596400005 596400007 596400008 596400011 596400012 596400016 596400017 596400018 596400019 596401003 596401007 596401015 596402002 596402003 596402004 596402006 596402009 596402011 596402012 596402015 596402016 596402017 596402018 596402020 596403001 596403002 596403006 596403007 596403009 596403010 596403011 596403012 596403013 596403014 596403015 596403017 596529008 596529009 596529016 596529018 596529019 596529020 597241001 597241003 597241004 597241005 597241006 597241009 597241010 597241017 597241018 597242001 597242002 597242003 597242004 597242005 597242007 597242011 597242016 597242018 597242019 598029003 598029005 598029009 598029011 598029014 598029015 598029016 598029018 598029020 598030003 598030007 598030011 598030018 598030019 598765002 598765004 598765005 598765006 598765015 598765018 598765019 598766006 598766011 598766012 598766014 598766016 598767003 598767005 598767007 598767013 598767017 598767018 598767020 598768001 598768002 598768006 598768007 598768008 598768010 598768017 598769001 598769004 598769007 598769008 598769009 598769012 598769013 598769014 598769017 598769018 599412005 599412006 599412008 599412009 599412010 599412012 599412013 599412015 599412018 599413001 599413002 599413004 599413005 599413006 599413014 599413016 599413017 599414001 599414002 599414005 599414006 599414007 599414008 599414012 599414019 599415003 599415004 599415005 599415012 599415013 599415017 599415018 600392004 600392005 600393001 600393002 600393003 600393006 600393007 600393008 600393009 600393010 600393014 600393015 600393016 600393017 600394002 600394018 600395007 600395010 600395016 600395017 600395018 600395019 600396003 600396008 600396016 600396018 600396020 601240006 601240009 601240011 601240015 601240016 601240017 601240018 601243001 601243012 601243014 601243015 601243017 601244008 601244010 601244013 601244015 601244019 601245003 601245007 601861004 601861005 601861006 601861011 601861012 601861013 601861014 601862004 601862006 601862007 601862008 601862015 601862016 601862018 601862019 601863001 601863002 601863004 601863008 601863011 601863015 601863016 601863020 601864001 601864002 601864004 601864006 601864010 601864014 601864016 601864017 601864019 601865009 601865010 601865011 601911001 601911002 601911005 601911006 601911008 601911017 601912002 601912003 601912005 601912007 601912015 601912018 601913001 601913002 601913003 601913005 601913011 601913012 601913013 601913017 601913020 603965004 603965011 603965015 603965016 603965018 603965019 603965020 603966003 603966004 603966011 603966013 603966019 603966020 605174001 605174002 605174003 605174004 605174005 605174007 605174009 605174011 605174014 605174015 605174017 605174019 605174020 605175002 605175004 605175005 605175006 605175007 605175008 605175009 605175010 605175011 605175015 605175016 605175018 605175019 605175020 606154002 606154010 606157001 606157007 606157010 606157013 606157015 606157017 606157019 606158003 606158004 606158008 606158010 606158016 606158017 606158019 606159002 606159005 606159007 606159008 606159011 606159012 606159013 606159016 606159019 607163001 607163003 607163007 607163008 607163011 607163014 607163015 607163016 607163017 607163018 607163020 607165001 607165003 607165005 607165006 607165008 607165009 607165014 607165015 607165016 607165017 607165018 607166002 607166003 607166004 607166005 607166006 607166007 607166008 607166009 607166015 607166016 607166019 607166020 608410001 608410002 608410006 608410008 608410009 608410011 608410012 608411002 608411003 608411004 608411012 608411013 608411014 608411017 608411018 608412002 608412005 608412006 608412007 608412010 608412011 608412012 608412013 608412014 608412015 608412017 608412018 608412019 608413001 608413002 608413004 608413005 608413007 608413011 608413013 608413020 608427006 609420002 609420004 609420017 609421001 609421003 609421005 609421007 609421008 609421011 609421013 609421014 609421017 609421018 609422001 609422003 609422004 609422005 609422009 609422012 609422013 609422015 609422016 609422019 609423001 609423002 609423003 609423005 609423006 609423007 609423010 609423011 609423012 609423013 609423015 609423016 609423017 609423018 610075002 610075003 610075005 610075010 610075011 610076001 610076003 610076008 610076009 610076014 610076015 610076016 610076017 610076018 610076019 610077001 610077003 610077006 610077008 610077009 610077011 610077016 610077018 610077019 610078002 610078008 610078010 610078011 610078013 610078014 610078020 611392002 611392004 611392005 611392013 611392016 611393005 611393006 611393007 611393009 611393014 611393015 611393016 611393020 611394002 611394003 611394004 611394005 611394011 611394012 611394013 611394014 611395002 611395003 611395004 611395005 611395006 611395007 611395008 611395009 611395010 611395011 611395012 611395013 611395015 611395017 611395019 611395020 611396001 611396002 611396005 611396011 611396012 611396013 611396014 611396015 611396018 612493005 612493012 612493015 612493016 612493017 612493018 612494006 612494007 612494009 612494010 612494011 612494013 612495002 612495003 612495011 612495013 613523001 613523002 613523004 613523005 613523011 613523012 613523013 613523015 613523016 613524002 613524003 613524007 613524017 613525003 613525004 613525005 613525006 613525009 613525014 613525015 613525016 613525019 614612001 614612002 614612003 614612004 614612012 614612013 614612014 615573001 615573002 615573004 615573005 615573006 615573007 615573010 615573016 615573017 615574001 615574002 615574003 615574004 615574009 615574013 615574015 615574017 615574018 615574020 616866004 616866005 616866009 616866010 616866012 616866015 616866016 616866017 616866018 616866019 616867002 616867005 616867008 616867009 616867010 616867011 616867012 616867013 616867014 616867015 616867016 616867017 616867019 616868001 616868003 616868005 616868006 616868008 616868009 616868011 616868012 616868014 616868015 616868017 616868019 616868020 618093002 618093003 618093006 618093008 618093010 618093013 618093017 618093018 618093019 618093020 618094001 618094003 618094004 618094005 618094007 618094011 618094012 618094015 618094020 618095001 618095002 618095004 618095005 618095008 618095009 618095010 618095016 618095017 618095018 618095019 618096001 618096002 618096003 618096005 618096008 618096009 618096010 618096014 618096015 618096016 618096018 619092005 619092006 619092008 619092011 619092014 619092015 619092016 619092017 619092018 619092019 619092020 619093002 619093003 619093004 619093005 619093006 619093013 619093016 619094002 619094003 619094004 619094005 619094006 619094007 619094008 619094010 619094011 619094012 619094014 619094015 619094017 619095005 619095007 619095008 619095009 619095012 619095015 619095019 619095020 619096001 619096003 619096004 619096005 619096006 619096007 619096008 619096009 619096010 619096011 619096016 620464004 620464013 620464016 620464018 620464019 620465001 620465004 620465006 620465007 620465008 620465009 620465010 620465011 620465013 620465017 620465018 620466001 620466002 620466004 620466005 620466006 620466007 620466008 620466010 620466012 620466013 620466014 620466015 620466019 620466020 620467008 620467014 620467015 620467016 620467017 620467018 620467020 620468001 620468007 621704001 621704002 621704003 621704004 621704005 621704006 621704013 621704014 621704015 621704016 621704017 621704018 621704019 621705001 621705004 621705005 621705006 621705007 621705010 621705012 621705014 621705015 621705016 621705017 621705018 621705020 621706003 621706007 621706009 621706010 621706019 621706020 623011001 623011002 623011005 623011006 623011007 623011011 623011012 623011013 623011014 623011015 623011016 623011017 623011018 623011019 623012001 623012002 623012003 623012004 623012005 623012006 623012007 623012008 623012009 623012010 623012011 623012012 623012013 623012015 623012016 623012017 623012018 623013003 623013007 623013008 623013009 623013011 623013012 623013013 623013014 623013015 623013020 623014002 623014004 623014006 623014007 623014009 623014011 623014015 623014016 623015001 623015003 623015010 623015011 623015012 623015013 623015014 623015015 623015017 624152013 624153001 624153002 624153003 624153004 624153005 624153006 624153007 624153011 624153012 624153013 624153014 624153015 624153016 624153018 624154006 624154008 624154009 624154010 624154018 624154019 624154020 624155002 624155003 624155005 624155006 624155009 624155010 624155011 624155012 624155015 624155017 624155018 624156001 624156002 624156003 624156004 624156005 624156006 624156008 624156011 624156015 624156016 624156019 625246002 625246003 625246004 625246005 625246012 625246013 625247011 625247016 625247017 625248001 625248004 625248005 625248006 625248008 625248017 625248018 625248019 625249012 625249013 625249014 625249015 625249016 625249017 625249018 625249019 626532011 626532013 626532014 626532015 626532017 627473001 627473002 627473003 627473005 627473010 627473011 627473012 627473017 627473018 627474005 627474006 627474007 627474008 627474009 627474011 627474012 627474013 627474014 627474016 627474017 627474018 627474019 627475002 627475010 627475011 627475013 627475015 627475017 627475019 627476001 627476002 627476003 627476004 627476005 627476006 627476007 627476008 627476009 627476011 627476014 627476015 627476016 627476017 627476018 627476019 627477002 627477003 627477004 627477005 627477006 627477008 627477009 627477013 627477014 627477015 627477016 627477017 628874003 628874004 628874005 628874006 628874007 628874008 628874014 628874016 628874017 628874018 628874019 628874020 628875006 628875010 628875011 628875013 628875014 628875015 628875018 628875019 628875020 628876001 628876002 628876004 628876005 628876006 628876007 628876008 628876009 628876011 628876012 628876014 628876015 628876017 628876018 628876019 628876020 628877001 628877002 628877003 628877004 628877012 628877013 628877014 628877015 628877016 628877017 628877019 628878001 628878002 628878003 628878004 628878005 628878006 628878007 628878008 628878009 628878010 628878011 628878012 628878013 628878014 628878015 628878016 628879001 628879004 628879005 628879006 628879007 628879012 628879014 628879016 628879017 628879020 629988003 629988004 629988005 629988006 629988008 629988009 629988013 629989001 629989003 629989005 629989006 629989007 629989008 629989016 629989017 629989019 629989020 629991003 629991005 629991012 629991013 629991016 629991019 629991020 629992004 629992013 629992014 629992015 629992017 629992018 629992020 629993004 629994001 629994002 629994005 629994007 629994010 629994012 629994014 629994015 629994016 629994017 629994018 629995002 629995003 629995008 629995009 629995015 629995020 631107004 631107005 631107006 631107007 631107008 631107009 631107010 631107014 631107017 631107018 631107019 631107020 631108001 631108003 631108004 631108007 631108008 631108010 631108011 631108019 631108020 631109001 631109004 631109010 631109014 631109017 631109020 631110002 631110003 631110004 631110005 631110006 631110007 631110011 631110012 631110013 631110014 631110015 631110016 631110017 631110018 631824002 631824011 631824012 631824013 631824015 631824017 631824018 631825001 631825004 631825006 631825009 631825011 631825013 631825016 631825018 632526006 632526008 632526009 632526010 632526013 632526015 632526017 632527002 632527012 632527013 632527014 632527015 632527016 632527017 632527018 632527019 632527020 632528001 632528002 632528003 632528004 632528005 632528007 632528008 632528009 632528010 632528011 632528012 632528014 632528015 632528019 632528020 632529008 632529009 632530003 632530004 632530005 632530006 632530007 632530012 632530014 633722001 633722003 633722004 633722005 633722011 633722015 633722016 633722017 633722019 633723003 633723012 633723014 633723015 633723016 633723017 633723018 633723019 634930006 634930007 634930008 634930009 634930012 634930013 634930017 634941004 634941006 634941011 634941012 634941017 634941018 636264002 636264003 636264005 636264006 636264007 636264018 636266010 636266018 637165003 637165005 637165006 637165008 637165009 637165011 637165012 637165013 637165014 637165015 637165016 637165017 637165019 637166003 637166004 637166006 637166007 637166010 637166012 637167003 637167004 637167005 637167006 637167007 637167008 637167009 637167011 637167014 637167015 637167016 637167017 637168004 637168005 637168006 637168007 637168010 637168011 637168012 637168014 637168018 637168019 637169001 637169003 637169004 637169012 637169014 637169017 637169018 637170001 637170003 637170004 637170007 637170009 637170010 637170011 637170014 637170017 637170018 638247002 638247008 638247010 638247013 638247015 638247016 638247019 638247020 638248001 638248003 638248008 638248009 638248010 638248013 638248014 638248018 638248020 639515004 639515005 639515006 639515007 639515008 639515009 639515010 639515012 639516002 639516004 639516005 639516006 639516013 639516014 639517002 639517003 639517004 639517005 639517006 639517008 639517015 639517016 639517017 639517018 639517019 639518002 639518004 639518011 639518014 639519001 639519002 639519003 639519008 639519009 639519011 639519013 639519014 639519016 639519017 639519018 640711002 640711004 640711012 640711013 640711016 640711017 640711019 640712004 640712006 640712011 640712012 640712018 640712020 640713006 640713007 640713008 640713009 640713010 640713011 640713012 640713014 640713015 640713016 640714002 640714015 640714017 640715001 640715002 640715003 640715004 640715005 640715006 640715007 640715008 640715010 640715014 640715018 640716014 640716016 641688006 641688007 641688008 641688010 641688012 641688015 641688019 641689005 641689006 641689007 641689013 641689014 641689016 641689017 641689018 641689019 641689020 641690007 641690009 641690010 641690014 641690015 641690017 641691001 641691002 641691004 641691007 641691008 641691009 641691010 641691011 641691013 641691016 641691019 641692003 641692005 641692008 641692009 641692011 641692014 641692015 641692016 641692017 641692018 643335002 643335003 643335006 643335007 643335011 643335013 643335015 643335017 643335020 643337002 643337003 643337004 643337009 643337010 643337011 643337012 643337013 643337018 643337019 643337020 643338009 643338010 643338013 643338014 643338018 643339001 643339002 643339003 643339006 643339008 643339009 643339019 644543004 644543005 644543008 644543009 644543013 644543014 644543019 644543020 644544002 644544003 644544004 644544005 644544006 644544007 644544008 644544009 644544011 644544013 644544019 644545002 644545003 644545004 644545005 644545006 644545007 644545008 644545014 644545016 644545017 644545018 644546010 644546011 644546017 645568004 645568006 645568007 645568009 645568010 645568014 645568015 645568016 645568019 645569003 645569004 645569005 645569006 645569007 645569009 645569010 645569016 645569019 645570001 645570002 645570003 645570007 645570009 645570011 645570012 645570013 645570014 645570015 645570020 645571001 645571004 645571016 645571017 645571018 645572001 645572002 645572003 645572004 645572005 645572007 645572008 645572009 645572010 645572011 645572013 645572014 645573001 645573002 645573003 645573005 645573006 645573008 645573009 645573010 645573011 645573012 645573013 645573014 645573015 645573016 645573017 647135003 647135005 647135007 647135008 647135009 647135010 647135011 647135012 647135016 647135017 647136001 647136004 647136005 647136006 647136007 647136008 647136013 647136014 647137001 647137002 647137003 647137005 647137007 647137008 647137009 647137010 647137012 647137014 647138001 647138006 647138009 647138012 647138013 647138014 647138015 648365001 648365006 648365011 648365012 648365013 648365014 648365015 648365016 648366001 648366002 648366004 648366006 648366007 648366009 648366013 648366015 648366016 648366018 648366019 648367001 648367003 648367010 648367012 648367014 648367015 648367016 648367017 648367020 648368001 648368003 648368004 648368005 648368012 648368016 648368019 648368020 649655011 649655015 649656001 649656007 649656008 649656010 649656012 649656016 649656017 649656020 650604001 650604002 650604003 650604004 650604006 650604009 650604012 650604013 650604016 650604017 650605002 650605003 650605004 650605005 650605006 650605007 650605009 650605010 650605011 650605012 650605015 650605016 650606002 650606003 650606004 650606005 650606010 650606011 650606014 650606015 650606018 650606019 650606020 650607002 650607003 650607004 650607007 650607014 650607018 650607019 650607020 651922002 651923001 651923006 651923009 651923010 651923013 651923014 651923016 651923018 651923019 651923020 651924002 651924004 651924005 651924006 651924015 651924016 651924018 651924020 651925002 651925003 651925004 651925005 651925006 651925007 651925008 651925009 651925011 651925012 651925013 651925014 651925017 651926001 651926002 651926004 651926005 651926008 651926009 651926010 651926013 651926014 651926015 651926016 651926018 653029002 653029003 653029007 653029008 653029013 653029015 653029019 653029020 653030001 653030002 653030004 653030008 653030010 653030011 653030015 653030017 653031001 653031002 653031011 653031016 653031017 653031018 653031020 653032004 653032005 653032007 653032008 653032009 653032010 653032013 653032014 653032015 653032016 653032018 653033001 653033004 653033006 653033008 653033011 653033012 653033016 653033017 653033019 654726006 654726008 654726009 654726010 654726011 654726013 654726014 654726015 654726017 654727001 654727005 654727006 654727009 654727012 654727014 654728001 654728002 654728004 654728006 654728007 654728009 654728010 654728011 654728012 654728013 654728015 654728016 654729001 654729003 654729006 654729008 654729013 654730006 654730015 656100001 656100002 656100003 656100004 656100005 656100006 656101001 656101004 656101005 656101006 656101007 656101008 656102001 656102002 656102003 656102004 656102005 656102008 656102009 656102010 656102011 656102013 656103003 656103005 656103006 656103007 656103008 656103010 656103011 656103012 656103018 656103019 656975004 656975006 656975016 656976001 656976009 656976010 656976011 656976012 656976013 656976014 656976015 656977006 656977007 656977008 656977010 658954008 658954012 658955002 658955003 658955005 658955006 658955007 658955009 658955010 658955011 658955013 658955015 658955017 658956002 658956003 658956004 658956005 658956007 658956008 658956013 658956014 658956017 658956018 658956019 658957001 658957003 658957004 658957005 658957006 658957009 658957010 658957011 658957012 658957013 658957014 658957015 658957016 658957018 658957019 658957020 659018002 659018003 659018004 659018006 659018008 659018011 659018013 659018016 659297003 659297006 659297009 659297010 659297011 659297012 659297013 659297014 659297015 659297016 659298005 659298006 659298007 659298010 659299007 659299008 659299009 659299010 659299011 659299012 659299013 659299014 659299015 659299016 659299017 659299018 659300001 659300003 659300004 659300006 659300007 659300009 659300011 659300012 659300019 659301001 659301002 659301003 659301005 659301006 659301007 659301009 659301010 659301011 659301012 659301014 659301015 659301016 660618001 660618003 660618004 660618005 660618006 660618007 660618008 660618009 660618010 660618011 660618015 660618017 660618018 660618019 660619001 660619002 660619003 660619004 660619005 660619007 660619008 660619009 660619010 660619011 660619012 660619013 660619014 660620002 660620003 660620004 660620005 660620006 660620007 660620008 660620009 660620010 660620011 660620012 660620013 660620014 660620015 660620016 660620017 660620018 660621001 660621002 660621006 660621009 660621011 660621012 660621013 660621014 660621015 660621016 660621017 660621018 660621019 660621020 660622002 660622003 660622004 660622006 660622007 660622008 660622009 660622010 660622011 660622012 660622013 660622014 660622015 660622016 660622017 660622019 660622020 661014001 661014008 661014014 661014015 661014016 661014017 661014018 661617002 661617003 661617004 661617007 661617009 661617014 661617015 661617016 661618001 661618002 661618003 661618004 661618005 661618006 661618007 661618010 661618012 661618013 661619003 661619006 661619009 661619010 661619012 661619015 661619016 661619018 661947001 661947004 661947005 661947013 661947015 661947016 661947017 661947019 661948001 661948002 661948003 661948005 661948007 661948008 661948009 661948010 661948011 661948013 661948015 661948017 661948018 661948019 661948020 661949001 661949004 661949007 661949008 661949009 661949010 661949011 661949013 661949018 661949019 661949020 661950002 661950003 661950009 661950011 661950012 661950017 661950018 661950019 661951001 661951004 661951005 661951006 661951007 661951008 661951011 661951012 661951016 661951017 662563001 662563002 662563003 662563004 662563005 662563006 662563007 662563010 662563011 662563013 662563014 662563016 662563017 662563019 662563020 664413003 664413005 664413006 664413007 664413012 664413013 664413014 664413018 664413020 666531002 666531004 666531005 666531006 666531007 666531008 666531011 666531012 666531013 666531014 666531017 666531018 666531019 666532001 666532007 666532011 666532013 666532014 666532015 666532017 666532018 666533001 666533011 666534002 666534003 666534004 666534008 666534009 666534010 666534011 666534012 666534013 666534014 666534016 666534018 667898002 667898004 667898009 667898010 667898011 667898014 667898017 667899001 667899002 667899003 667899004 667899005 667899006 667899008 667899011 667899012 667899013 667899014 667900001 667900002 667900003 667900004 667900005 667900006 667900007 667900008 667900009 667900010 667900011 667900012 667900015 667900016 667900018 667900019 667900020 667901002 667901004 667901006 667901007 667901008 667901013 667901014 667901018 669104001 669104009 669104011 669104012 669104013 669104014 669104015 669105001 669105002 669105003 669105004 669105005 669105006 669105007 669105008 669105009 669105010 669105011 669105012 669105013 669105014 669105015 669105016 669105017 669106002 669106003 669106004 669106005 669106006 669106007 669106009 669106014 669106016 669106017 669106018 669106019 669107001 669107002 669107003 669107006 669107007 669107008 669107009 669107012 669107017 670140001 670140002 670140006 670140007 670140008 670140009 670140010 670140011 670140012 670140013 670140016 670140017 670141002 670141003 670141008 670141009 670141010 670141013 670141014 670141015 670141017 670141018 670141019 670142002 670142005 670142007 670142008 670142012 670142013 670142014 670142019 670142020 670260002 670260003 670260006 670260007 671428002 671428004 671428005 671428006 671428010 671428011 671428012 671428013 671428014 671428016 671428018 671429001 671429004 671429005 671429008 671429009 671429010 671429017 671429018 671430001 671430002 671430006 671430008 671430014 671430015 671430016 671430017 671430019 671431001 671431002 671431004 671431005 671431006 671431007 671431008 671431009 671431010 671431011 671431012 671431013 671431014 671431015 671431016 671431017 671431018 671431019 671431020 671432002 671432008 671432011 671432012 671432013 671432014 671432015 671432016 671432017 671432018 671432020 672569002 672569003 672569008 672569010 672569015 672569020 672570004 672570005 672570008 672570009 672570010 672570011 672570012 672570013 672570014 672570015 672570018 672571001 672571002 672571003 672571004 672571005 672571006 672571009 672571010 672571012 672571014 672571018 672571019 672572001 672572002 672572003 672572006 672572007 672572011 672572012 672572013 672572014 672572015 672572017 672572018 672572019 673690002 673690004 673690007 673690009 673690010 673690015 673690017 673690019 673691004 673691005 673691007 673691011 673691013 673691014 673691017 673691020 673692001 673692002 673692005 673692006 673692007 673692008 673692009 673692011 673692013 673692014 673692017 673692018 673692019 673693004 673693005 673693006 673693008 673693010 673693011 673693014 673693016 673693018 673694001 673694007 673694008 673694010 673694014 673694015 673694016 673694017 673694018 674810002 674810004 674810005 674810006 674810007 674810008 674810011 674810012 674810013 674810015 674810016 674810017 674810018 674810019 674810020 674811002 674811004 674811005 674811007 674811008 674811009 674811015 674811018 674811019 674812002 674812003 674812009 674812010 674812011 674812012 674812013 674812015 674812016 674812017 674812018 674812019 674813002 674813003 674813005 674813006 674813008 674813011 674813012 674813015 674813016 674813018 675683001 675683002 675683006 675683008 675683010 675683012 675683015 675683018 675683019 675683020 675684001 675684002 675684004 675684005 675684006 675684007 675684010 675684011 675684012 675684013 675684015 675684016 675684018 675685001 675685006 675685008 675685009 675685012 675685013 675685015 675685016 675685018 676943001 676943003 676943005 676943006 676943007 676943008 676943010 676943012 676943013 676943014 676943015 676943017 676944001 676944002 676944003 676944005 676944006 676944007 676944008 676944009 676944010 676944012 676944013 676944014 676945001 676945002 676945003 676945004 676945005 676945006 676945007 676945008 676945009 676945010 676945011 676945012 676945013 676945014 676945015 676945016 676945017 676945018 676945019 676945020 677855001 677855002 677855003 677855007 677855011 677855016 677855019 677855020 677856001 677856002 677856004 677856008 677856011 677856014 677856018 677856019 677857001 677857002 677857003 677857004 677857005 677857006 677857007 677857009 677857011 677857012 677857015 677857017 677857018 677857019 677857020 677858001 677858011 677858012 677858013 677858015 677858018 677859001 677859002 677859003 677859004 677859005 677859007 677859008 677859011 677859016 677859017 677859018 678944001 678944002 678944004 678944005 678944007 678944008 678944009 678944015 678944017 678944018 678944020 678945001 678945008 678945009 678945010 678945011 678945013 678945014 678945015 678945018 678945020 678946003 678946005 678946008 678946010 678946012 678946013 678946016 678946017 678946018 678946019 680109001 680109004 680109005 680109006 680109007 680109008 680109009 680109011 680109012 680109014 680109016 680109017 680109019 680110001 680110002 680110003 680110004 680110005 680110007 680110008 680110009 680110010 680110011 680110012 680110013 680110014 680110015 680110016 680110018 680110019 680110020 680111003 680111004 680111008 680111011 680111012 680111013 680111014 680111016 680111017 680111019 680111020 680112001 680112002 680112003 680112004 680112005 680112007 680112008 680112010 680112011 680112013 680112016 680112019 680112020 680113001 680113002 680113005 680113006 680113007 680113008 680113009 680113010 680113011 680113012 680113013 680113014 680113015 680113016 680113017 680113018 680113019 680113020 680114001 680114002 680114003 680114005 680114007 680114013 680114014 680114015 680114016 680114017 680114018 680114019 681338001 681338002 681338003 681338004 681338005 681338006 681338007 681338010 681338012 681338013 681338014 681338016 681338017 681339001 681339005 681339008 681339012 681339014 681339015 681339018 681339019 681340001 681340002 681340003 681340004 681340005 681340006 681340007 681340008 681340009 681340010 681340011 681340012 681340013 681340014 681340015 681340016 681340017 681340018 681341001 681341002 681341003 681341004 681341005 681341009 681341011 681341012 681341013 681341014 681341016 681341017 681342001 681342002 681342003 681342004 681342005 681342006 681342008 681342010 681342011 681342012 681342013 681342014 681342015 681342017 681342018 681342020 681343004 681343005 681343007 681343008 681343010 681343011 681343013 681343016 682273001 682273003 682273007 682273009 682273010 682273012 682273013 682273014 682273015 682273016 682273019 682273020 682274002 682274003 682274005 682274006 682274007 682274008 682274009 682274010 682274011 682274012 682274013 682274014 682274015 682274016 682274017 682274018 682274019 682274020 682275003 682275008 682275009 682275010 682275011 682275012 682275014 682275015 682275018 682275019 682275020 682276002 682276004 682276005 682276007 682276009 682276010 682276011 682276015 682276016 682276017 682276018 682276019 683451001 683451002 683451003 683451004 683451005 683451008 683451009 683451010 683451011 683451015 683451016 683451018 683451019 683452001 683452004 683452005 683452006 683452007 683452008 683452009 683452010 683452011 683452012 683452013 683452016 683452017 683452019 683452020 683453001 683453003 683453004 683453006 683453008 683453009 683453010 683453011 683453012 683453013 683453014 683453015 683453016 683453017 683453018 683453019 683454005 683454006 683454007 683454008 683454011 683454014 683454018 683454019 684636001 684636002 684636005 684636006 684636007 684636008 684636011 684636014 684636015 684636016 684636017 684636018 684636019 684637001 684637003 684637004 684637005 684637007 684637009 684637011 684637016 684637017 684637018 684637019 684638002 684638004 684638006 684638007 684638012 684638013 684638014 684638015 684638016 684638017 684638018 685844001 685844002 685844003 685844004 685844005 685844006 685844007 685844010 685844011 685844012 685844013 685844014 685844015 685844016 685844017 685844018 685844019 685844020 685845001 685845002 685845004 685845005 685845006 685845007 685845008 685845009 685845010 685845011 685845017 685846001 685846002 685846005 685846006 685846007 685846008 685846009 685846010 685846011 685846012 685846014 685846015 685846016 685846017 685846018 685846019 685846020 685847001 685847002 685847003 685847005 685847006 685847010 685847012 685847013 685847014 685847015 685848001 685848002 685848003 685848005 685848006 685848007 685848009 685848011 685848012 685848013 685848014 685848015 685848016 685848018 685848019 685849003 685849004 685849005 685849009 685849010 685849011 685849013 685849016 685849018 685849020 685850002 685850003 685850004 685850005 685850006 685850007 685850008 685850012 685850013 685850015 685850016 685850017 685850018 685850019 685850020 687052001 687052002 687052004 687052005 687052006 687052007 687052008 687052009 687052010 687052011 687052012 687052013 687052017 687052018 690077002 690077007 690077008 690077009 690077010 690077011 690077014 690077018 690077019 690077020 690078001 690078004 690078006 690078009 690078010 690078011 690078012 690078014 690078015 690079002 690079003 690079004 690079005 690079006 690079008 690079009 690079010 690079011 690079012 690079014 690079015 690079017 690079018 690079019 690850001 690850002 690850003 690850004 690850005 690850006 690850007 690850008 690850009 690850010 690851001 690851002 690851003 690851004 690851005 690851006 690851007 690851008 690851011 690851015 690851017 690851018 691436001 691436002 691436003 691436004 691436005 691436007 691436009 691436010 691436011 691436012 691436013 691436015 691436017 691436018 691436019 691436020 691437001 691437002 691437006 691437007 691437008 691437009 691437010 691437011 691437012 691437013 691437014 691437015 691437017 691437018 691437019 691437020 691438002 691438004 691438005 691438007 691438009 691438010 691438011 691438014 691438015 691438019 692239001 692239002 692239003 692239004 692239005 692239006 692239007 692239008 692239009 692239010 692239011 692239012 692239013 692239014 692239015 692239016 692239017 692240003 692240004 692240005 692240006 692240008 692240009 692240010 692240011 692240012 692240013 692240014 692240015 692240016 692240019 692241001 692241003 692241006 692241007 692241008 692241009 692241010 692241011 692241012 692241013 692241014 692241016 692241017 692242001 692242007 692242009 692242010 692242013 692242014 692242016 692242019 694561001 694561004 694561005 694561006 694561008 694561009 694561010 694561011 694561013 694561015 694561017 694562002 694562004 694562005 694562006 694562007 694562008 694562010 694562011 694562012 694562013 694562015 694562016 694562017 694562018 694562019 694562020 695794001 695794003 695794004 695794005 695794006 695794007 695794008 695794010 695794011 695794012 695794013 695794014 695794015 695794016 695794017 695794018 695795003 695795004 695795005 695795006 695795009 695795010 695795011 695795012 695795013 695795015 695795017 695795018 695795020 696786001 696786002 696786003 696786004 696786005 696786007 696786008 696786009 696786010 696786011 696786013 696786015 696786016 696786017 696786018 696787001 696787002 696787003 696787004 696787005 696787006 696787007 696787008 696787009 696787010 696787013 696787014 696787015 696787018 696787019 696787020 696788002 696788003 696788004 696788005 696788008 696788012 696788015 696788016 696788019 698169001 698169002 698169003 698169004 698169005 698169006 698169008 698169009 698169010 698169011 698169012 698169013 698169014 698169015 698169016 698169017 698169018 698169019 700706001 700706003 700706004 700706005 700706006 700706007 700706008 700706009 700706010 700706011 700706012 700706013 700706014 700706015 700706016 700706017 700706019 700706020 700707001 700707002 700707003 700707004 700707005 700707006 700707007 700707008 700707010 700707011 700707012 700707013 700707014 700707016 700707017 700707018 700707019 700707020 701582001 701582002 701582004 701582007 701582008 701582009 701582010 701582011 701582012 701582013 701582014 701582015 701582018 701582020 701740001 701740002 701740003 701740005 701740006 701740007 701740011 701740012 701740016 701740017 701741001 701741005 701741006 701741007 701741008 701741009 701741011 701741014 701741016 701741017 701741018 701741019 701742001 701742002 701742003 701742004 701742006 701742008 701742009 701742011 701742013 701742014 701742015 701742016 701742017 701742018 701742019 701742020 701743001 701743002 701743003 701743004 701743005 701743006 701743007 701743008 701743010 701743015 701743016 701743018 701743019 701744002 701744003 701744004 701744010 701744012 703117001 703117002 703117003 703117004 703117005 703117006 703117008 703117009 703117011 703117012 703117013 703117017 703117018 703117019 703118001 703118002 703118003 703118004 703118005 703118006 703118007 703118008 703118009 703118013 703118014 703118015 703118016 703118017 703118018 703118019 704337001 704337002 704337003 704337004 704337005 704337006 704337007 704337008 704337009 704337010 704337011 704337012 704337013 704337014 704337015 704337016 704337017 704337018 704337019 704337020 705324004 705324005 705324006 705324008 705324010 705324011 705324013 705324016 705325001 705325005 705325006 705325011 705325012 705325013 705325015 705325016 705325019 705325020 705326001 705326003 705326004 705326006 705326010 705326011 705326012 705326013 705326014 705326016 705326017 706627005 706627007 706627013 706627016 706627017 706627018 706627019 706628016 707613016 707616001 707616003 707616005 707616007 707616010 707616011 707616013 708953019 Lot Serial Number Model number 72404127 517943019 529147013 538287013 539938005 548507003 549620010 558049003 559261007 560759014 560999004 562043017 562043019 562149001 564666007 564666015 565713012 566725004 566725009 566843002 566886014 567627016 568635014 569630013 570062016 570954005 570954011 571253005 571362018 571414001 572251009 572894001 572894016 573095001 573185006 573870014 574040003 574040006 574040009 574092004 574092008 574092009 574092010 574093003 574093007 574093017 574093020 574218009 574255005 574332003 574332004 574332010 574332012 574332015 574332016 574586002 574586005 574586006 574702005 574702008 574702011 574702017 574831007 574868007 574868009 574868016 574868019 574927012 574927016 575014006 575014008 575014011 575014017 575066019 575138006 575138018 575138020 575355003 575493012 575555005 575555015 575673007 575726006 575726015 575726020 575805013 575940003 575940004 575940008 575940011 576017011 576017015 576476002 576476007 576476008 576476010 576476015 576541001 576541002 576541003 576541010 576541012 576541014 576541018 576597007 576597010 576676007 576676008 576676011 576676016 576768006 576768007 576768018 576842008 576842009 576842015 576843005 576843006 576843018 576843019 576957011 577496002 577496018 577496020 577681010 577681012 577681015 577681016 577681018 577682002 577682005 577682008 577682009 577682010 577682017 577848003 577899001 577899009 577899010 578469011 578469016 578469019 579064008 579064016 579128010 579128011 579128013 579128015 579128018 579203002 579203003 579203004 579203005 579203006 579203009 580032002 580032004 580032006 580032009 580032013 580032016 580093001 580093002 580093008 580093010 580093012 580093019 580145002 580145005 580145010 580145011 580197002 580197015 580272012 580982004 580982007 580982009 580982010 580982011 580982013 580982014 581135007 581135009 581135011 581135015 581135016 581135018 581315004 581315007 581315015 581476005 581476013 581476014 581630003 581630004 581986001 581986002 581986003 581986010 581986011 581986016 581986017 581987001 581988005 581988008 581988010 581988017 581989011 581989013 581990004 581990011 581990017 581991012 581991013 581991017 581991020 581992003 581992017 581993002 581993006 581993007 581993008 581993011 581994010 581994011 583051006 583051009 583051010 583051015 583051016 583051019 583052002 583052011 583052012 583052014 583053010 583053016 583054002 583054004 583055011 583055013 583055017 583056011 583056012 583056019 583503008 583504005 583505006 583505019 583506015 583507003 583507007 583507017 584053001 584053005 584053018 584054005 584054009 584054010 584054011 584055004 584055007 584055010 584057002 584057004 584057007 584057016 584057017 584058002 584058019 585093004 585093010 585093018 585095001 585095015 585096004 585096006 585096017 586207003 586207004 586207007 586207008 586207014 586207018 586207019 586208005 586208014 586208017 586209004 586209009 586209013 586210006 586210009 586210016 586210017 586211005 586211015 586211016 586212001 586212019 586480001 586480004 586480006 586480014 586480015 586480020 587052005 587052015 587052019 587053003 587054017 587054019 587055002 587055007 587056006 587056008 587056010 587056011 587984001 587984006 587984011 587984015 587984016 587984018 587984020 587985007 587985009 587985011 587985016 587985017 587986003 587986005 587986007 587986012 587986016 587989002 587989014 587990001 587991002 587991011 587991012 587991016 587992008 587992013 589114001 589114007 589114015 589115004 589115006 589115007 589115008 589116009 589117001 589117002 589117013 589117015 589117018 589118010 589118012 589118013 589118016 589119003 589119007 589119014 589119019 589120012 589121003 589121015 589121018 590142005 590142007 590142012 590142014 590142017 590143002 590143009 590144001 590144006 590144012 590145008 590146004 590146009 590146010 590146012 590146017 590147001 590147002 590147006 590147010 590147017 590148013 590149001 590149002 590149003 590149011 590149018 590149019 591309001 591309010 591309013 591309014 591310009 591310010 591310012 591310016 591310017 591311001 591311002 591311007 591311008 591311009 591311010 591312004 591312006 591312014 591313005 591313010 591314003 591314005 591314015 592085001 592085004 592087013 592088001 592088002 592088008 592089005 592089012 592089016 592089017 592090007 592090016 592090019 593316001 593316005 593317002 593317013 593317016 593318006 593318008 593318013 593318018 593319016 593319018 593495005 593495007 593495008 593495009 594302005 594302011 594303002 594303005 594303009 594303013 594303015 594303016 594304004 594304008 594304015 594304016 594304017 594304019 594305002 594305006 594305009 594306004 594306007 594306009 594306010 594306013 594307001 594307005 594307012 594307017 594308002 594308007 594308011 594308016 594308017 595398009 595398020 595399001 595399016 595400012 595402003 595402006 595402007 595402017 595402019 595402020 595403006 595403008 595403011 596395011 596395012 596395014 596395018 596396010 596396018 596397008 596397012 596397013 596397014 596397017 596397020 596398015 596398016 596398017 597237009 597237011 597237019 597238001 597238007 597238008 597238010 597238013 597238019 598027001 598027007 598027009 598027010 598027013 598027019 598028005 598028019 598763002 598763004 598763006 598763013 598763014 598764009 598764014 598764015 598764017 600389004 600389008 600389009 600389013 600390005 600390006 600391004 600391011 601235006 601235014 601235015 601235017 601236001 601236003 601236017 601237001 601237005 601237017 601237020 601238003 601238005 601238016 601238017 601238018 601239018 601856007 601856012 601857003 601857012 601857013 601858009 601858012 601859002 601859010 601859013 601859015 601860005 601860016 601907004 601907010 601907017 601908006 601908007 601909010 601909014 601910010 601910020 603957002 603957013 603957014 603958012 603958015 603958018 603959004 603959012 603959014 603960003 603960004 603960007 603960011 603960015 603960016 603960019 603961006 603961007 603961009 603962010 603962013 603963011 603963013 603963014 603963016 603963017 603964006 603964016 603964018 605176003 605176007 605176009 605177014 605177019 605178002 605178016 605180008 605180013 605180014 605181004 605181006 605181011 605181013 605181014 605181018 605181019 606146010 606146018 606148006 606148007 606149010 606150015 606151002 606151003 606151004 606151005 606151006 606151007 606151008 606151009 606151010 606151011 607156001 607156015 607157006 607157012 607157017 607157020 607158003 607158012 607159001 607159002 607159006 607159007 607159009 607159012 607160006 607160015 607161011 608405002 608405003 608405004 608405009 608405011 608405013 608405014 608405015 608405018 608405019 608406005 608406016 608406017 608407003 608407013 608407018 608408002 608408008 608408011 608408017 608408019 609414005 609414006 609414009 609415005 609415006 609415015 609415018 609415019 609416003 609416010 609416013 609416018 609418007 609418009 609418012 609418019 609419005 609419007 610068006 610068007 610068010 610068018 610069005 610069006 610069009 610069010 610069013 610069014 610069017 610069020 610070010 610070015 610071007 610071008 610071009 610071019 610071020 610072002 610072005 610073009 610073011 610073016 610074004 610074010 611386003 611386007 611386008 611386009 611386010 611386020 611387008 611387011 611387012 611387015 611387016 611387017 611388003 611388004 611388005 611388016 611389001 611389011 611389012 611389019 611390001 611390004 611390009 612488002 612488006 612488009 612488014 612488019 612489004 612489017 612490017 612490020 612491001 612491002 612491003 612491004 612491005 612491006 612491007 612491008 612491009 612491010 612491014 612491019 612491020 612492002 612492018 613519019 613520006 613521006 613521009 613521016 613521019 614607006 614607016 614607017 614608005 614608009 614608019 614609011 614609013 614609017 614611007 614611010 614611012 614611015 614611016 615568001 615568007 615568011 615569011 615569020 615570003 615570007 615570008 615570011 615570013 615570015 615570019 615571010 615571018 615572002 615572005 615572009 615572013 615572019 616860014 616861019 616862010 616862015 616862018 616862019 616863002 616863004 616863012 616863013 616863014 616863019 616864002 616864006 616864007 616864012 616864014 616864015 618086014 618086015 618086016 618086017 618088010 618088019 618089004 618089006 618089010 618090003 618091009 618091010 618091012 618092002 618092003 618092004 618092005 618092007 618092008 618092009 618092010 618092011 618092012 618092015 618092016 619085001 619085003 619085014 619085019 619086014 619086015 619086018 619087007 619087012 619087019 619088002 619088005 619089008 619089015 619090002 619090005 619090008 619090011 619090015 619090019 619091004 619091014 619091018 619091019 620456007 620456014 620456020 620457004 620457005 620457007 620457008 620457016 620457019 620458001 620458012 620459009 620459010 620459014 620459020 620460012 620460013 620460015 620460017 620460020 620461005 620461006 620461008 620461010 620461012 620461013 620461019 620462003 620462004 620462005 620462006 620462019 620463004 620463007 620463010 620463012 620463018 620463020 621697003 621697008 621697009 621697010 621697015 621697018 621698008 621698012 621698014 621698015 621698020 621699015 621700001 621701002 621701009 621701010 621701011 621701013 621701014 621701015 621701016 621701017 621702004 621703005 621703008 621703009 623004003 623004014 623004015 623004017 623004018 623004020 623005013 623005017 623006006 623006010 623006015 623006019 623007006 623007007 623008001 623008007 623008008 623008014 623009004 623009009 623010003 623010007 623010013 624139004 624139007 624139008 624139018 624140001 624140018 624140019 624140020 624143001 624143002 624143003 624143007 624143020 624144002 624144009 624144013 624144014 624144015 624144017 624144018 624144019 624145005 624145019 624146002 624146003 624146004 624146005 624146008 624147004 624147006 624147007 624147011 624147014 625237013 625237014 625238004 625238008 625238018 625238020 625239015 625239016 625239017 625239018 625242001 625242006 625242007 625242012 625242016 625243010 625243019 625244009 625245004 625245009 625245011 625245012 625245014 625245016 625245017 626526006 626526009 626526010 626526012 626526014 626526016 626527003 626527018 626528002 626528013 626528017 626528018 626529011 626529019 626530003 626530016 626530020 626531011 626531016 627381008 627381012 627381013 627466008 627466016 627467005 627467006 627467011 627468016 627468019 627469004 627469009 627469010 627469015 627469016 627469017 627470009 627470013 627470017 627471013 627471018 627472006 627472011 627472016 627472017 627472020 628867009 628867013 628867018 628868001 628868019 628869003 628869018 628870002 628870009 628870014 628871010 628871016 628871019 628871020 629979004 629979009 629979019 629980006 629980011 629980012 629980015 629980016 629981001 629981002 629981009 629981010 629981011 629981013 629981018 629981019 629981109 629982002 629982004 629982006 629983008 629983010 629983013 629983020 629984003 629984004 629984005 629985005 629985012 631102001 631102002 631102003 631102004 631102014 631102017 631103001 631103007 631103008 631103011 631104002 631104008 631104010 631104011 631104013 631104014 631104020 631105006 631105009 631105011 631105012 631105014 631105018 631106005 631106014 631106015 631106019 631821002 631821004 631821018 631822005 631822007 631823001 631823006 631823007 632372008 632372014 632373006 632522009 632522017 632522020 632523013 632523015 632523016 632523018 632523019 632523020 632524001 632524012 632524017 632525009 633717002 633717003 633718013 633718014 633718015 633718016 633718020 633719004 633719013 633719016 633719017 633720002 633720008 633720010 633720019 633720020 633721002 633721003 633721008 633721012 633721013 634927006 634927011 634927017 634928001 634928010 634928016 634929001 634929003 634929007 634929010 636262004 636262006 636262012 636262016 636262020 636263001 636263017 637158009 637158014 637159018 637161008 637161014 637161017 637161019 637163002 637163016 637163018 637163020 637164005 637164011 637164019 637564001 637564004 637564017 638239006 638239010 638240007 638240013 638240015 638240019 638241005 638241007 638241009 638241013 638241018 638242003 638242007 638242016 638242018 638243007 638244001 638244006 638244008 638244013 638244014 638244018 638245003 638245004 638246004 638246010 638246019 639506002 639506003 639506004 639506005 639506011 639506012 639506019 639507003 639507009 639507013 639507014 639507016 639508006 639508007 639508017 639509004 639509006 639509010 639510006 639510012 639510015 639510018 639511006 639511008 639511009 639511011 639512003 639512011 639512014 639512018 639513003 639513005 639513015 639514001 639514006 639514011 639514014 639514015 639514016 639514020 640703002 640703003 640703006 640703014 640704002 640704009 640704014 640705003 640705007 640705011 640705012 640705019 640706007 640706009 640706016 640707004 640707010 640707012 640707016 640708001 640709013 640709015 640709020 640710002 640710009 641679001 641679005 641679008 641679014 641680002 641680003 641680009 641681002 641681005 641681007 641682004 641682005 641682006 641682008 641682009 641682011 641682013 641683004 641683009 641683012 641684017 641684018 641685003 641686001 641686002 641686006 641686017 641687004 641687008 641687009 641687012 641687013 643329003 643329007 643329008 643329009 643329013 643329015 643330002 643330005 643330008 643330016 643330018 643331002 643331005 643331009 643331012 643332001 643332004 643332005 643332006 643333007 643333014 643333015 643333019 643334002 643334013 643334015 643334017 644535005 644535008 644535015 644536006 644536016 644536020 644537007 644537015 644537018 644538017 644538020 644539010 644539013 644540005 644540010 644540012 644540013 644540014 644540020 644541003 644541010 644541011 644542010 645560003 645560006 645561006 645561007 645561012 645561014 645561016 645562001 645562017 645562018 645563008 645563016 645563017 645564003 645564012 645564015 645565007 645565011 645566002 645566003 645566013 645567008 647128006 647128015 647129003 647129012 647129015 647129018 647130011 647131001 647131015 647131016 647131018 647134006 647134011 647134015 647134016 648357011 648357017 648357018 648358005 648358006 648358008 648358011 648358018 648359012 648359017 648360002 648360008 648360010 648360016 648360020 648361005 648361015 648361017 648362011 648362012 648362013 648362015 648362019 648363004 648363005 648363009 648364004 648364006 648364008 648364011 648364012 648364015 648364018 648364019 649653006 649653007 649653010 649653014 649653017 649653018 649654012 649654013 649654014 649654015 649654016 649654018 650595001 650595007 650595014 650595017 650596003 650596012 650597001 650597004 650597006 650598001 650598006 650598013 650598018 650599001 650599015 650600001 650600004 650600008 650600013 650600020 650601007 650601016 650602007 650602014 650602017 651914001 651914004 651914005 651914010 651914013 651914014 651914016 651914019 651915004 651915007 651915012 651915013 651915019 651916005 651916009 651917007 651917013 651917015 651917018 651918003 651919007 651919017 651919019 651920005 653021001 653021006 653021010 653022001 653022013 653022018 653023001 653023002 653024003 653024017 653025009 653025011 653025012 653026009 653026012 653026013 653026014 653026016 653027012 653027014 653028004 653028010 653028012 653028017 654718011 654718012 654719001 654719007 654719008 654719014 654719015 654719017 654720001 654720004 654720008 654720009 654720010 654721002 654721003 654721011 654722010 654722018 654723011 654724005 654724008 654724010 654724013 654725006 654725007 654725011 656105004 656105006 656105013 656105015 656106001 656106004 656106011 656106012 656106018 656107002 656107004 656107008 656107009 656108002 656108008 656109007 656109015 656110002 656110008 656110009 656110016 656982007 656982008 656982009 656982011 656982012 656982013 656983001 656983013 656984001 656984003 656984004 656984005 656984010 656984011 656984018 656985003 656985004 656985005 656985018 656986001 656986002 656986005 656986011 656986017 656987002 656987003 656987004 656987007 656987015 656988012 656988014 658163005 658163011 658163012 658163013 658163017 658287012 658287018 658288006 658288013 658289003 658289009 658289011 658289013 658289016 658291001 658291002 658291005 658291013 658291016 658293001 658293005 659290002 659290004 659290005 659290006 659290008 659290010 659291009 659291012 659292004 659292008 659292010 659293001 659293008 659293009 659293012 659293013 659294001 659294005 659294017 659295010 659295014 659296007 659296012 660612007 660613002 660613011 660613013 660613014 660613016 660615001 660615004 660615005 660615007 660615011 660615012 660615013 660615015 660615018 660616001 660616003 660616008 660616009 660616012 660617001 660993003 660993009 660993011 660993012 660993018 661006001 661006005 661006007 661006008 661006011 661006012 661006013 661006015 661610011 661610016 661611016 661611019 661612001 661612010 661612016 661612020 661613001 661613003 661613009 661614008 661615006 661615007 661616008 661616011 661616013 661939001 661939004 661939005 661939006 661939007 661939014 661939018 661940008 661940013 661940015 661941001 661941004 661941005 661941007 661941010 661941011 661941012 661941013 661941014 661941015 661941018 661942002 661942011 661942012 661942017 661943001 661943002 661943003 661943004 661943008 661943011 661943013 661943016 661944001 661944003 661944006 661944008 661944011 661945002 661946006 661946012 661946013 661946018 662536005 662536011 662536012 663379009 663379012 663808010 664406004 664406007 664406013 664406018 664407005 664407019 664408005 664408007 664408008 664408010 664408011 664408015 664408018 664409007 664410001 664410002 664410006 664410010 664410017 664411006 664411008 664411012 664411015 664411017 664411020 664412012 665440004 665440007 665440015 665441003 665441010 665441014 665442005 665442008 665443002 665443004 665443008 665443017 665444004 665444010 665444011 665444012 665444020 666524011 666525007 666526002 666526009 666527001 666527006 666527011 666527016 666528001 666528004 666528008 666528013 666528014 666528019 666528020 666529003 666529005 666529006 666529007 666529008 666529013 666529014 666529017 666530001 666530003 666530006 666530007 666530010 666530014 666530016 666530018 666530019 667659007 667659008 667659009 667891013 667891014 667892004 667892010 667892019 667893006 667893009 667894002 667894004 667895001 667895007 667895008 667895010 667895015 667896003 667896005 667896011 667896016 667896017 669098002 669098010 669098013 669099001 669099003 669099004 669099013 669099017 669099018 669100004 669100006 669100007 669100008 669100009 669100010 669100012 669100013 669100016 669100017 669101002 669101003 669101004 669101006 669101007 669101008 669101010 669101011 669101012 669101013 669101014 669101015 669101019 669102001 669102002 669102005 669102008 669102011 669102016 669102017 669102019 669103005 670136003 670136004 670136005 670136013 670136016 670137001 670137007 670137009 670137015 670139002 670139006 670139009 670139015 670139016 671186004 671186019 671423008 671423019 671424001 671424007 671424009 671425007 671425008 671425010 671425013 671425015 671426004 671426013 671426014 671426016 671426017 671426018 671426020 671427008 671427010 671427012 672561015 672562002 672562011 672562012 672562013 672563004 672563019 672564004 672564007 672564009 672564015 672565001 672565002 672565009 672565012 672565014 672566002 672566004 672566005 672566010 672566017 672566018 672567002 672567003 672567004 672567009 672567010 672567012 672567014 672567016 672567017 672567018 672567019 672567020 672568012 672568014 672568016 673682005 673682019 673683011 673683015 673683016 673683017 673683020 673684010 673684016 673685002 673685005 673685006 673685009 673685010 673686003 673686004 673686005 673686006 673686012 673686013 673686014 673687006 673688010 673688019 673688020 674803015 674804003 674804004 674804010 674804013 674804016 674804018 674804019 674805004 674805005 674805014 674805017 674805019 674805020 674806004 674806010 674806014 674807002 674807007 674807010 674807011 674807017 674807018 674808001 674808002 674808003 674808008 674808014 674808018 674808020 674809004 674809017 675675002 675675013 675675014 675676005 675676006 675676007 675676008 675676018 675677013 675678010 675678011 675678013 675678014 675678016 675679009 675679013 675679014 675679015 675679017 675680009 675680015 675680018 675681010 675681011 675681012 675681013 675681014 675681016 675681018 675682006 675682009 675682010 675682013 675682014 675682015 676933006 676933014 676934001 676934003 676934005 676934015 676935002 676935008 676935009 676935010 676935014 676936004 676936008 676936009 676936010 676936012 676936014 676936015 676936019 676937001 676937002 676937004 676937005 676937006 676937010 676937012 676937013 676937016 676938001 676938002 676938004 676938007 676938009 676938011 676938014 676938018 676939007 676939011 676940002 676940003 676940008 676940012 676940014 676940017 676941001 676941002 676941016 676941018 676942004 676942006 676942008 676942009 676942010 676942012 676942013 676942014 677846001 677846002 677846003 677846015 677846019 677847001 677847002 677847003 677847009 677847012 677847013 677847015 677848011 677848018 677849003 677849004 677849006 677849008 677849009 677849010 677849014 677849015 677849016 677849018 677850003 677850015 677851002 677851003 677851007 677851008 677851010 677851018 677853001 677853007 677854008 677854010 677854014 677854015 678934011 678934012 678934018 678935001 678935010 678935014 678935015 678935016 678935017 678935019 678936002 678936005 678936010 678937007 678937008 678937010 678937011 678937012 678937013 678937014 678937015 678937016 678937017 678937018 678937020 678938008 678939001 678939005 678939011 678939014 678939019 678940001 678940002 678940007 678940013 678940014 678940015 678940016 678940017 678940018 678940019 678940020 678941008 678941009 678941012 678942001 678942002 678942004 678942005 678942007 678942008 678942010 678942011 678942015 678942016 678942017 678942018 678942019 678942020 678943001 678943005 678943006 678943007 678943009 678943010 678943011 678943013 678943019 680101003 680101004 680101005 680101006 680101007 680101008 680101009 680101010 680101011 680101012 680101013 680101014 680101015 680101017 680102001 680102011 680102019 680103003 680103015 680103017 680104008 680104011 680104012 680104017 680104020 680105010 680105012 680105017 680105020 680106006 680106007 680106008 680106019 680106020 681330002 681330004 681330009 681332004 681332010 681332011 681332012 681332017 681333005 681333010 681333012 681333015 681334007 681334010 681334014 681335004 681335007 681335010 681336001 681336002 681336003 681336012 681336015 681336016 681336017 681336019 681337011 682266002 682266005 682266010 682266017 682267003 682267012 682267019 682267020 682268004 682268006 682269001 682269002 682269004 682269011 682269012 682269015 682269017 682270002 682270004 682270007 682270010 682270011 682270017 682271001 682271002 682271004 682271007 682271009 682271010 682271011 682271013 682271015 682271016 682271017 682271018 682271019 682271020 682272002 682272003 682272008 682272019 682272020 683442001 683442002 683442005 683442011 683442012 683442015 683442018 683442019 683443001 683443002 683443005 683443007 683443014 683443015 683444004 683444007 683444010 683444011 683444014 683445004 683445011 683445013 683446001 683446002 683446003 683446008 683446010 683446012 683446013 683446014 683446016 683446018 683447003 683447004 683447006 683447008 683447011 683447014 683447015 683447016 683447017 683447018 683447019 683448002 683448003 683448004 683448012 683448013 683448016 683448017 683448018 683448019 683449017 683450003 683450004 683450005 683450009 683450010 683450012 683450014 683450015 683450016 683450018 684633004 684633005 684633007 684633009 684633011 684633014 684633016 684633017 684633018 684633020 684634003 684634009 684634019 684635002 684635008 684635014 684635016 684635017 684635018 685841011 685841012 685841018 685841019 685841020 685842002 685842003 685842004 685842006 685842007 685842011 685842012 685842013 685842014 685842020 685843005 685843006 685843007 685843009 685843011 685843013 685843017 685843018 687048002 687048007 687048009 687048010 687048011 687048018 687049004 687049005 687049006 687049007 687049008 687049009 687049011 687049012 687049013 687049014 687049015 687049016 687049017 687049018 687049019 687049020 687050011 687050015 687050017 687051001 687051002 687051012 687051017 687867011 687867012 687867013 687868003 687868005 687869009 687869015 687870001 687870002 687870004 687870006 687870007 687870010 687870011 687870012 687870018 688845004 688845005 688845007 688845008 688845017 688846001 688846004 688846005 688846006 688846007 688846008 688846011 688846013 688846016 688846018 688846019 688846020 688847001 688847002 688847003 688847006 688847007 688847008 688847010 688847012 688847013 688847014 688847015 688847016 688847017 688847018 688847019 688847020 688848004 688848005 688848007 688848010 688848012 688848013 688848014 688848015 688848016 688848017 688848018 688848019 688849001 688849007 688849014 688849015 688849016 690072002 690072003 690072006 690072008 690072012 690072013 690072015 690072017 690072018 690072019 690072020 690073001 690073003 690073004 690073005 690073007 690073008 690073010 690073011 690073012 690073013 690073016 690073017 690073018 690074006 690074007 690074011 690074012 690075003 690075006 690075010 690075013 690075015 690075017 690076002 690076016 690076018 690846005 690846008 690846013 690846014 690846019 690847001 690847013 690847016 690847018 690847020 690848001 690848002 690848003 690848007 690848011 690848013 690848014 690848015 690848016 690848019 690849001 690849002 690849003 690849011 690849014 690849015 690849018 690849020 691431005 691431009 691431010 691431013 691431018 691432006 691432009 691432010 691433002 691433008 691433010 691433012 691433013 691433014 691433015 691433017 691433018 691433019 691434006 691434010 691434015 691434019 691435001 691435006 691435009 691435010 691435011 691435012 691435013 691435015 691435016 691435018 692231004 692231010 692231014 692231017 692231018 692232004 692232005 692232006 692232010 692232012 692232013 692232015 692232017 692232018 692232019 692232020 692233003 692233005 692233006 692233007 692233013 692233014 692233015 692233016 692233017 692233018 692233019 692234002 692234003 692234004 692234005 692234006 692234007 692234008 692234009 692234011 692234013 692234018 692234019 692234020 692235004 692235008 692235009 692235017 692235019 692236001 692236003 692236008 692236009 692236012 692236013 692236018 692236019 692237004 692237005 692237008 692237010 692237013 692237014 692237015 692237016 692238001 692238012 692238017 693713001 693713002 693713003 693713006 693713013 693713014 693713015 693713016 693713019 693713020 693714001 693714008 693714010 693714016 693714017 693715002 693715008 693715010 693715011 693715012 693716003 693716004 693716009 693716013 693716019 693716020 693717001 693717004 693717008 693717010 693717011 693717016 693717018 693717019 694553001 694553005 694553008 694553011 694553013 694553017 694554001 694554003 694554005 694554006 694554007 694554012 694554020 694555001 694555003 694555005 694555008 694555009 694555016 694555018 694557002 694557003 694557006 694557013 694557017 694557018 694558003 694558004 694558009 694558015 694558016 694558018 694559006 694559007 694559008 694559009 694559011 694559016 694559019 694560001 694560002 694560004 694560005 694560006 694560008 694560009 694560012 694560018 695787010 695787012 695788002 695788006 695788008 695788011 695788018 695789011 695789012 695790005 695790011 695790013 695791003 695791005 695791010 695791011 695791015 695791018 695791020 695792001 695792006 695792010 695792016 696778001 696778002 696778004 696778007 696778010 696778011 696778013 696778014 696778016 696778018 696779010 696779018 696779019 696780005 696780006 696780009 696780011 696780017 696781002 696782001 696782003 696782008 696782010 696782013 696782018 696783001 696783007 696783008 696783013 696783016 696783020 696784001 696784002 696784003 696784006 696784008 696784009 696784013 696784014 696784016 696784020 696785001 696785006 696785010 696785011 696785012 696785013 696785015 696785016 698164002 698164005 698164006 698164009 698164010 698164011 698164013 698164017 698164018 698164020 698165002 698165004 698165005 698165008 698165010 698165013 698165014 698165015 698165019 698165020 698166006 698166007 698166011 698166012 698166015 698166018 698167001 698167002 698167003 698167004 698167009 698167010 698167012 698167015 698167016 698167017 699388004 699388012 699388014 699388018 699388019 699389001 699389004 699389005 699389007 699389015 699389018 699390001 699390009 699390013 699390016 699391001 699391002 699391004 699391008 699391011 699391012 699391013 699391014 699391015 699391016 699391017 699391018 699391019 699391020 699392002 699392003 699392004 699392005 699392006 699392007 699392008 699392009 699392010 699392012 699392013 699392017 699392018 699393004 699393005 699393008 699393011 699393012 699393016 699393018 699393019 699394002 699394004 699394005 699394007 699394008 699394009 699394010 699394012 699394013 699394014 699394015 699394016 699394017 699394018 699394019 699394020 699395003 699395004 699395008 699395009 699395012 699395014 699395015 699395019 699395020 699396001 699396002 699396003 699396004 699396009 699396012 699396013 699396014 699396017 699396018 699396019 700696001 700696003 700696004 700696005 700696006 700696007 700696009 700696010 700696014 700696015 700696017 700696018 700697002 700697004 700697005 700697006 700697007 700697013 700697014 700697015 700697016 700697018 700698001 700698002 700698008 700698009 700698015 700699003 700699005 700699006 700699007 700699010 700699012 700700004 700700008 700700009 700700013 700700014 700701003 700701006 700701011 700701013 700701015 700701016 700701017 700702003 700702004 700702005 700702006 700702007 700702011 700702014 700703007 700703008 700703009 700703010 700703017 700703018 700703019 700704003 700704005 700704017 700704018 700704019 701733008 701733010 701733015 701734001 701734003 701734006 701734009 701734010 701734013 701734018 701735003 701735004 701735005 701735006 701735016 701736002 701736018 701736019 701736020 701737005 701737006 701737007 701737008 701737010 701737011 701737013 701738010 701738017 701738020 701739007 701739010 701739011 701739016 701739020 703110004 703110011 703110013 703110017 703110018 703111005 703111007 703111009 703111015 703111019 703112005 703112007 703112017 703113004 703113007 703113009 703113012 703113017 703114001 703114002 703114003 703114006 703114011 703114013 703114014 703114018 703114020 703115001 703115007 703115009 703115011 703115013 703116001 703116002 703116003 703116004 703116010 703449001 703449006 703449007 703449008 703449012 703450001 703450002 703450005 703450006 703450008 703450015 703450018 703450019 704330004 704330006 704330007 704330009 704330011 704330012 704330014 704330016 704331003 704331017 704332001 704332002 704332003 704332004 704332008 704332009 704332012 704332014 704332016 704333001 704333002 704333005 704333007 704333010 704333011 704333013 704333016 704335005 704335013 705320001 705320002 705320003 705320004 705320005 705320006 705320007 705320011 705320012 705320014 705322001 705322003 705322005 705322006 705322007 705322009 705322011 705322012 705322013 705322014 705322020 705323002 705323003 705323005 705323007 705323011 705323013 705323017 706621002 706621004 706621005 706621006 706621009 706621010 706622020 706623015 706626009 707607006 707607010 707607015 707607020 707608018 Lot Serial Number Model number 720172-01 610651001 610651002 610651003 610651004 610651005 610651006 610651007 610651008 610651009 610651010 610651011 610651012 622966001 622966002 622966003 622966004 622966005 628617001 628617002 634914001 634914002 634914003 634914004 634914005 634914006 634914007 634914008 634914009 634914010 649686001 649686002 649686003 649686004 649686005 652568001 664382001 664382002 664382003 664382004 664382005 674801001 674801002 674801003 678805001 678805002 678805003 678805004 678805005 678805006 678805007 678805008 678806001 688224001 688224002 688224003 688224004 688224005 688224006 688224007 688224008 693882001 693882002 693882003 693882004 693882005 693882006 703145001 703145002 703145003 703145004
Other devices under the same FDA submission
31 devices cleared under P000053 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- AMS 800 Artificial Urinary Sphincter 72404130Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic1 death1,803 reports
- AMS 800 Artificial Urinary Sphincter 72404131Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic1,394 reports
- AMS 800 Artificial Urinary Sphincter 72400024Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic1 recall1 death999 reports
- AMS 800 Artificial Urinary Sphincter 720157-01Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic591 reports
- AMS 800 Artificial Urinary Sphincter 72404132Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic576 reports
- AMS 800 Artificial Urinary Sphincter 72400160Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic487 reports
- AMS 800 Artificial Urinary Sphincter 72404127Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic443 reports
- AMS 800 Artificial Urinary Sphincter 72400161Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic358 reports
- AMS 800 Artificial Urinary Sphincter 72404133Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic220 reports
- AMS 800 Artificial Urinary Sphincter 72400098Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic1 recall137 reports
- AMS 800 Artificial Urinary Sphincter 72400162Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic111 reports
- AMS 800 Artificial Urinary Sphincter 720133-01Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic99 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page