REF: 91-6704***THINFLAP SYSTEM CROSS-DRIVE, SELF-DRILLING… recallZ-2873-2011 · Biomet Microfixation, Inc. · initiated May 17, 2010
Terminated All recalls posted in July 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet Microfixation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On May 17, 2010, Biomet Microfixation issued a Safety Alert on their Thinflap System Cross Drive, Self Drilling Screw, 1.5x4mm, Model #: 91-6704, Lot#: 173770. The product is labeled incorrectly. Some packages of 91-6704 lot 173770 are labeled as cross drive screws, but contain center drive screws. All affected customers were notified of the recall on May 17, 2010.
FDA root cause category: Labeling mix-ups
What the company is doing
Biomet Microfixation notified distributors and customers by letter, email and telephone calls. Distributors and customers were instructed to fill out and return the Inventory Reconciliation sheet, via Fax to 904-741-9425. Distributors were also instructed to notify their customers if the product was distributed further. For questions regarding this recall customers were instructed to call 1-800-874-7711 or 904-741-4400, ext. 468.
Recall details
- Recall number
- Z-2873-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 17, 2010
- Posted by the FDA
- Jul 22, 2011
- Recalling firm
- Biomet Microfixation, Inc.
- Quantity
- 927
- Product code
- HWC
- FDA submissions
- K953385
- Distribution
- Worldwide Distribution - USA including AL, CA, GA, HI, IL, MO, NC, NY, TX and the country of Japan.
Product description
"***REF: 91-6704***THINFLAP SYSTEM CROSS-DRIVE, SELF-DRILLING SCREW 1.5X4MM (.059X.159") (WXL)***TITANIUM***NON-STERILE PRODUCT***BIOMET MICROFIXATION 1520 TRADEPORT DRIVE JACKSONVILLE, FL 32218 USA*** www.biometmicrofixation.com***QTY 00001***". Internal fixation screws intended to aid in the alignment and stabilization of bone in the oral cranio-maxillofacial skeletal system.
Code information (models, lots, serial numbers)
Lot 173770
Other devices under the same FDA submission
41 devices cleared under K953385 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- "1.5MM" SYSTEM 95-6104Biomet Microfixation, Inc · FASTENER, PLATE, CRANIOPLASTY30 reports
- "1.5MM" SYSTEM 95-1505Biomet Microfixation, Inc · COVER, BURR HOLE23 reports
- "1.5MM" SYSTEM 91-6105Biomet Microfixation, Inc · COVER, BURR HOLE20 reports
- THINFLAP SYSTEM 91-6704Biomet Microfixation, Inc · Plate, cranioplasty, preformed, alterable19 reports
- "1.5MM" SYSTEM 95-6105Biomet Microfixation, Inc · FASTENER, PLATE, CRANIOPLASTY16 reports
- "1.5MM" SYSTEM 91-6104Biomet Microfixation, Inc · Plate, cranioplasty, preformed, alterable9 reports
- "1.5MM" SYSTEM 91-6106Biomet Microfixation, Inc · COVER, BURR HOLE7 reports
- "1.5MM" SYSTEM 01-6482Biomet Microfixation, Inc · PLATE, BONE6 reports
- TMS SYSTEM 25-2005Biomet Microfixation, Inc · PLATE, BONE6 reports
- 1.5MM SYSTEM 91-6105XBiomet Microfixation, Inc · FASTENER, PLATE, CRANIOPLASTY5 reports
- 1.5MM SYSTEM 95-6105XBiomet Microfixation, Inc · Plate, cranioplasty, preformed, alterable4 reports
- "1.5MM" SYSTEM 01-6480Biomet Microfixation, Inc · PLATE, BONE3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page