MEDRAD Avanta Fluid Management Injection System, Catalog… recallZ-2866-2011 · Medrad Inc · initiated Jun 2, 2011
Terminated All recalls posted in July 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medrad Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Event was initially intended to be a product enhancement of the MEDRAD Avanta Fluid Management Injection System. Upgrades were initiated in October, 2007. Observance of a possible increase in complaints and adverse event reporting resulted in the firm issuing "Important Product Use Reminder" to remind users of proper procedures for using the system and the need for diligence in reducing air embo
FDA root cause category: Device Design
What the company is doing
Medrad Inc., sent two types of "MEDICAL DEVICE FIELD CORRECTION" notifications on June 2, 2011 to all affected customers. 1. To those customers who Avanta Systems have already undergone market withdrawal and thus have been upgraded, the notification provided customer awareness that prior action should be considered a field correction. The letter stated that NO action is required since their system(s) has already been upgraded. If you have any questions, concerning these activities contact the Director of Quality Systems at (724) 940-6905. 2. To those customers whose Avanta Systems have not yet been upgraded through market withdrawal, the notification informed customers that the recalling firm is undertaking a Field Correction to upgrade their system(s). The letter included a step by step guide that will allow customers to verify that their system(s) has NOT been upgraded. Customers are instructed to complete and fax this form to MEDRAD at (412) 406-0941. If you have any questions, concerning these activities contact the Product Manager at (412) 767-2400, extension 8401.
Recall details
- Recall number
- Z-2866-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jun 2, 2011
- Posted by the FDA
- Jul 22, 2011
- Recalling firm
- Medrad Inc
- Quantity
- 842
- Product code
- DXT
- FDA submissions
- K050456
- Distribution
- Worldwide Distribution -- (USA) (nationwide) including the state of Puerto Ricoand the countries of: AE, AL, AT, BE, BG, BR, CA, CH, CO, CY, DE, DK, DO, ES, FI, FR, GB, HK, HR, HU, IE, IL, IS, IT, JO, JP, KR, KW, LY, MA, MT, MX, MY, NL, PA, PH, PK, PT, QA, RS, RU, SA, SE, SG, SI, TR, VI, and ZA.
Product description
MEDRAD Avanta Fluid Management Injection System, Catalog numbers: AVA 500 TABL, AVA 500 PEDL Product Usage: Administration of intravascular radio-opaque contrast media and common flushing solutions during angiographic procedures.
Code information (models, lots, serial numbers)
Catalog numbers: AVA 500 TABL, AVA 500 PEDL
Other devices under the same FDA submission
3 devices cleared under K050456 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MEDRAD® Avanta Fluid Management Injection SystemBayer Medical Care Inc. · Injector And Syringe, Angiographic1 death5 reports
- MEDRAD® Avanta Fluid Management Injection SystemBayer Medical Care Inc. · Injector And Syringe, Angiographic1 death3 reports
- MEDRAD® Avanta Single Patient Disposable Set – Transducer Compatible AVA 500 SPAT LBayer Medical Care Inc. · Injector And Syringe, Angiographic1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page