Baxter Solution Set with Duo-Vent Spike recallZ-2920-2011 · Baxter Healthcare Corp. · initiated Jul 21, 2011
Terminated All recalls posted in August 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Solution Sets are being recalled because the position of the slide clamp and roller clamp may have been reversed during assembly. The reversed orientation of the slide clamp and roller clamp will cause the set to be incorrectly loaded into an infusion pump, resulting in fluid flow from the patient to the pump, rather than from the pump to the patient.
FDA root cause category: Employee error
What the company is doing
Baxter Healthcare Corporation sent a "Urgent Product Recall" letter dated July 21, 2011 to all affected customers. The letter included a description of the problem and requested they contact Baxter at 1-888-229-0001 to return all affected product. They were asked to complete and return the attached customer reply form, and to contact and notify their customers. For questions contact Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.
Recall details
- Recall number
- Z-2920-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jul 21, 2011
- Posted by the FDA
- Aug 3, 2011
- Recalling firm
- Baxter Healthcare Corp.
- Quantity
- 4,752 units
- Product code
- FPA
- FDA submissions
- K961225
- Distribution
- (USA) Nationwide Distribution including the states of Alabama, California, Iowa, Illinois, Indiana, Massachusetts, Maryland, Michigan, North Carolina, New Hampshire, Ohio, Pennsylvania, Tennessee and Virginia.
Product description
Baxter Solution Set with Duo-Vent Spike; 109" (2.8 m), Male Luer Lock Adapter 10 drops/mL. Baxter Healthcare Corporation. A sterile, single use, nonpyrogenic fluid pathway for use with rigid nonvented and flexible containers.
Code information (models, lots, serial numbers)
Product Code 1C8507s, lot numbers GR291542, GR292409, GR292797 and GR294454.
Other devices under the same FDA submission
24 devices cleared under K961225 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CLEARLINK/CONTINU-FLO/DUO-VENT 2C8541Baxter Healthcare Corporation · Set, administration, intravascular2 recalls797 reports
- CLEARLINK/CONTINU-FLO 2C8537Baxter International Inc. · Set, administration, intravascular2 recalls490 reports
- CLEARLINK/CONTINU-FLO 2C8519Baxter Healthcare Corporation · Set, administration, intravascular3 recalls1 death221 reports
- CLEARLINK/CONTINU-FLO 2R8537Baxter International Inc. · Set, administration, intravascular1 recall136 reports
- CLEARLINK/CONTINU-FLO/DUO-VENT UC8519Baxter International Inc. · Set, administration, intravascular1 recall66 reports
- CLEARLINK/CONTINU-FLO 2H8519Baxter Healthcare Corporation · Set, administration, intravascular93 reports
- CLEARLINK/CONTINU-FLO 2R8519Baxter International Inc. · Set, administration, intravascular1 recall37 reports
- CLEARLINK/CONTINU-FLO 2R8546Baxter International Inc. · Set, administration, intravascular59 reports
- CLEARLINK/DUO-VENT 2C7462Baxter Healthcare Corporation · Set, administration, intravascular57 reports
- DUO-VENT 1C8507Baxter Healthcare Corporation · Set, administration, intravascular1 recall4 reports
- CLEARLINK 2C7461Baxter Healthcare Corporation · Set, administration, intravascular49 reports
- CLEARLINK/DUO-VENT 2H7462Baxter International Inc. · Set, administration, intravascular47 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.