The Trabecular Metal Modular Acetabular System is a modular… recallZ-3279-2011 · Zimmer Inc. · initiated Sep 1, 2011
Terminated All recalls posted in September 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lot # 61132698 and Lot 001122467 (added 9/ 18/2012) may have a manufacturing non conformance that could prevent the devices from functioning as intended. Specifically, devices in the affected lot may have a locking ring groove depth that prevents the corresponding liner from snapping into the shell. The groove provides the locking ring room to displace when inserting a liner into the shell. Th
FDA root cause category: Nonconforming Material/Component
What the company is doing
Zimmer, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated July 27, 2011 to all affected customers of lot 61132698 and 9/18/2012 to only affected customer of lot 001122467. The letters describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to stop using and to quarantine the affected product. An Inventory Return Certification Form was included with the letter for customers to complete and return to Zimmer, Inc. via fax at (574) 372-4265. Replacement product will be provided for all returned product. Contact Zimmer, Inc. at 1-800-613-6131 for questions regarding this recall.
Recall details
- Recall number
- Z-3279-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 1, 2011
- Posted by the FDA
- Sep 27, 2011
- Recalling firm
- Zimmer Inc.
- Quantity
- 24 devices lot 61132689; 1 device lot 001122467
- Product code
- JDI
- FDA submissions
- K021891
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of CA, NC,NY, PA, WA, and WI and countries of Australia, Germany, The Netherlands, and Switzerland.
Product description
The Trabecular Metal Modular Acetabular System is a modular acetabular cup system consisting of a shell with a preassembled locking ring, a liner, and optional screws. The Shells are available in a variety of sizes and are designed to be used with Trilogy Acetabular System liners. The shell implants are used to replace the hip socket during total hip arthroplasty. The device is indicated for primary or revision surgery for rehabilitating hips damaged as a result on noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis,protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis and diastrophic variant. The device is intended for either cemented or noncemented use.
Code information (models, lots, serial numbers)
Part 00-6202-052-21, Lot 61132698 and lot 001122467
Other devices under the same FDA submission
10 devices cleared under K021891 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trabecular Metal® 6202-54-22Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented5 reports
- Trabecular Metal® 6202-52-20Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- Trabecular Metal® 6202-58-20Zimmer, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer4 reports
- Trabecular Metal® 6202-50-20Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented3 reports
- Trabecular Metal® 6202-52-22Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented3 reports
- Trabecular Metal® 6202-56-20Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- Trabecular Metal® 6202-50-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trabecular Metal® 6202-60-20Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trabecular Metal® 6202-62-20Zimmer, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 report
- Trabecular Metal® 6202-62-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗