Infinity Gamma X XL Portable Patient Monitor recallZ-3048-2011 · Draeger Medical Systems, Inc. · initiated Jul 29, 2011
Terminated All recalls posted in August 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Monitor keys may spontaneously become inoperative or active. This may cause a membrane switch panel malfunction, which could result in the monitor discharging a patient automatically.
FDA root cause category: Device Design
What the company is doing
Draeger Medical sent an "Urgent Medical Device Recall Notification" letters dated July 2011 to all affected customers. The letter included product and problem affected. The letter informed the customer a Draeger representative will contact them to upgrade the system. Until the solution is implemented, Draeger recommends the Delta and/or Gamma XXL may be used. If the affected monitors become inoperative or activate without user interaction, Draeger advises removing the monitor from service and contacting a Draeger representative. For questions or for information on this recall please call Draeger Service at (800) 543-5047 (press 1 at the prompt, then 2, then 32349).
Recall details
- Recall number
- Z-3048-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jul 29, 2011
- Posted by the FDA
- Aug 22, 2011
- Recalling firm
- Draeger Medical Systems, Inc.
- Quantity
- 29
- Product code
- MHX
- FDA submissions
- K070566
- Distribution
- Worldwide Distribution - (USA) Nationwide distribution including the states of AK, FL, LA, MN, MS, MO, NJ, NC, OH, OK, PA, TN, TX, NC, NH, NJ, NY, WV, and WI; and the countries of Australia, Austria, Argentina, Bahrain, Brazil, Bulgaria, Canada, China, Colombia, Czech Republic, France, Germany, Georgia, Greece, Hungary, Italy, Ireland, India, Kazakhstan, Latvia, Macedonia, Malaysia, Mexico, Netherlands, Norway, Russian Fed., Romania, Saudi Arabia, Slovakia, Spain, South Africa, Switzerland, United Arab Emirates, United Kingdom, Uruguay, Vietnam, and Yemen.
Product description
Infinity Gamma X XL Portable Patient Monitor. Draeger Medical Systems, Inc. Portable monitor used to monitor vital signs of neonatal, pediatric, and adult patients for bedside and transport applications.
Code information (models, lots, serial numbers)
Catalog Number(s): MS18597/MS18852 with serial numbers (US) 6001246076, 6001246575, 6001247173, 6001365874, 6001067474, 6001073681, 6001073779, 6001073877, 6001095176, 6001098966, 6001099466, 6001069472, 6001077080, 6001198466, 6001216679, 6001229674, 6001208082, 6001211782, 6001218178, 6001220086, 6001243676, 6001246771, 6001245968, 6001357179, 6001214877, 6001075679, 6001077178, 6001083180, and 6001095773.
Other devices under the same FDA submission
5 devices cleared under K070566 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Delta Monitor MS18597Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2 deaths11 reports
- DeltaXL Monitor MS18596Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 reports
- Infinity Vista XL Monitor MS18986Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 death4 reports
- Infinity Kappa Monitor MS18853Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)4 reports
- Infinity Gamma XXL MS18852Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 59568).
- Z-3047-2011 · TerminatedInfinity Delta Portable Patient Monitor. Draeger Medical Systems, Inc. Portable monitor used to monitor vital signs of neonatal, pediatric, and adult patients for bedside and tr…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page