Skip to content

SYSTEM 1E Liquid Chemical Sterilant Processing System… recallZ-3124-2011 · Steris Corporation · initiated Aug 11, 2011

Terminated All recalls posted in September 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

STERIS learned that SYSTEM 1E users may experience aborted cycles due to a failure within the UV light system.

FDA root cause category: Other

What the company is doing

STERIS sent an "Urgent Field Correction Notice" letter dated July 2011 to all affected customers. The letter describes the product, problem and actions to be taken. The letter states that STERIS Service Representative will contact customers to schedule the repair of the recalled product. For further information or questions contact STERIS Field Service Dispact at 1-800-333-8828.

Recall details

Recall number
Z-3124-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 11, 2011
Posted by the FDA
Sep 1, 2011
Recalling firm
Steris Corporation
Quantity
2,235 units
Product code
MLR
FDA submissions
K090036, K101409, K102462
Distribution
Worldwide Distribution - USA (nationwide) and the country of Bahrain.

Product description

SYSTEM 1E Liquid Chemical Sterilant Processing System, Model # P6500, S/N: 400001 through 405005; Mfr. By: STERIS Corporation, Mentor, OH 44060. Product Usage: The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Code information (models, lots, serial numbers)

Model # P6500, S/N: 400001 through 405005

Other devices under the same FDA submission

4 devices cleared under K090036, K101409, K102462 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page