VNS Therapy Aspire HC Generator and VNS, Model 105 recallZ-0210-2012 · Cyberonics, Inc · initiated Aug 12, 2011
Terminated All recalls posted in November 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cyberonics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The devices are being recalled because the output current delivered to the vagus nerve is less than the design intent and there is a potential charge imbalance at the lead cathode and generator-can during stimulation.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Cyberonics sent a Field Safety Alert on August 15, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The firm decided to recall and initiated contact with implanting facilities in the United States on 08/15/2011. A Field Safety Alert was delivered on 09/19/2011 to implanting facilities. Customers were instructed not to prescribe AspireHC pulse generators for implant until Cyberonics resumes shipment of AspireHC pulse generators that have the issue remedied. Cyberonics recommends that physicians continue their current practice of titrating programming parameters for patients implanted with an AspireHC pulse generator to achieve appropriate clinical outcomes, paying particular attention to report of: Lack or loss of efficacy Pain, inflammation, swelling/edema in the neck area; and/or Patient perception of stimulation or muscle twitching in the area surrounding the implant. Cyberonics recommended routine systems diagnostic testing to aid in device and patient monitoring. Physicians may contact Clinical Technical Support at 866-882-8804 or via email at [email protected]. For questions regarding this recall call 281-228-7323.
Recall details
- Recall number
- Z-0210-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 12, 2011
- Posted by the FDA
- Nov 18, 2011
- Recalling firm
- Cyberonics, Inc
- Quantity
- 340 units
- Product code
- LYJ
- FDA submissions
- P970003
- Distribution
- Worldwide Distribution - USA including TN, TX, IL, OH, FL, CA, CO, OR, AZ, NY, PA, MO, NJ, WI, OK, MI, ID, KY, IA, MA, NE, GA, LA, NY, and Norway.
Product description
VNS Therapy Aspire HC Generator and VNS, Model 105 Cyberonics, Inc. 100 Cyberonics Blvd., Houston, TX 77058 For implantation in humans to treat Epilepsy and Depression.
Code information (models, lots, serial numbers)
Model 105, VNS Therapy AspireHC Generator
Other devices under the same FDA submission
23 devices cleared under P970003 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VNS Therapy® SenTiva™ Model 1000 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted For Epilepsy3 recalls188 deaths3,378 reports
- VNS Therapy® AspireSR® Model 106 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted For Epilepsy1 recall151 deaths2,934 reports
- VNS Therapy® PerenniaFLEX® Model 304 Lead 304-20Livanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)1 death1,614 reports
- VNS Therapy® Demipulse® Model 103 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)24 deaths371 reports
- VNS Therapy® AspireHC® Model 105 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)1 recall17 deaths307 reports
- VNS Therapy® Demipulse Duo® Model 104 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)20 deaths266 reports
- VNS Therapy® PerenniaDURA® Model 303 Lead 303-20Livanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)1 recall181 reports
- VNS Therapy® Symmetry™ Model 8103 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)8 deaths223 reports
- VNS Therapy® Model 302 Lead 302-20Livanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)161 reports
- VNS Therapy® SenTiva Duo™ Model 1000-D GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)2 recalls4 deaths28 reports
- VNS Therapy® PerenniaFLEX® Model 304 Lead 304-30Livanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)108 reports
- VNS Therapy® Pulse Model 102 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted (Depression)6 deaths81 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page