Trilogy 100, 200 and 202 Ventilators recallZ-0296-2012 · Respironics, Inc. · initiated Oct 11, 2011

Terminated All recalls posted in January 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Respironics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Trilogy 100, 200 and 202 units are being recalled due to low impeller press-force used during the manufacturing process. If the press force were low enough, theoretically the impeller could move up the shaft and come into contact with the inside of the blower housing/impeller enclosure. This may impact therapy.

FDA root cause category: Process design

What the company is doing

Philips Healthcare, began notifying customers via telephone on October 11, 2011. The telephone notification discussed the product, problem and actions to be taken. The customers were instructed to not remove a ventilator from a patient without providing a replacement unit; provide a fax number or email address Philips can use to send the Business Reply Form to complete and return via fax to: 800-733-9962 or email: [email protected]; locate their affected devices (if the device is with a patient, replace it with a non-affected unit-Philips will provide a loaner overnight if needed)-(if the device is in their inventory, quarantine it and return it using the RA number provided in the email/fax that Philips sends them); and If loaner units are needed, they are to call 1-877-387-3311 to arrange return of your affected device and to obtain a loaner unit. If you have any questions or concerns, contact Philips Respironics at 877-387-3311.

Recall details

Recall number
Z-0296-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 11, 2011
Posted by the FDA
Jan 11, 2012
Recalling firm
Respironics, Inc.
Quantity
127
Product code
CBK
FDA submissions
K083526
Distribution
Worldwide distribution: USA (nationwide) including states of: AZ, CA, CO, GA, ID, IL, KS, MI, MO, NE, NV, NJ, OH, PA, TN, and TX; and countries of: Australia, Austria, Canada, China, Denmark, Germany, Great Britain, India, Italy, Japan, Korea (Republic of [South] Korea), Netherlands, Philippines, Saudi Arabia, Spain (Espana), and Turkey.

Product description

Trilogy 100, 200 and 202 Ventilators The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Serial Numbers: GV111030238, GV111040533, GV111040627, GV111040632, GV111040640, GV111040651, GV111040652, GV111040653, TV011031510, TV011040511, TV011040516, TV011040518, TV011040524, TV011040529, TV011040530, TV011040601, TV011040602, TV011040609, TV011040615, TV109081805, TV109081806, TV109081807, TV111021835, TV111022551, TV0111022832, TV111030422, TV111030441, TV111030445, TV111030453, TV111030774, TV111030790, TV111030853, TV111030932, TV111030947, TV111030976, TV111030978, TV111031001, TV111031011, TV111031037, TV111031045, TV111031057, TV111031064, TV111031067, TV111031113, TV111031403, TV111031413, TV111031420, TV111031462, TV111031554, TV111032252, TV111032254, TV111032306, TV111032316, TV111032562, TV111032809, TV111032811, TV111032812, TV111032822, TV111032827, TV111032829, TV111032830, TV111032832, TV111032833, TV111032901, TV111032905, TV111032906, TV111032909, TV111032912, TV111032913, TV111032914, TV111032920, TV111032923, TV111032924, TV111032925, TV111032926, TV111032929, TV111032934, TV111032935, TV111032938, TV111032939, TV111032940, TV111033001, TV111033004, TV111033005, TV111033006, TV111033009, TV111033010, TV111033016, TV111033018, TV111033019, TV111040110, TV111040113, TV111040117, TV111040404, TV111040405, TV111040411, TV111040415, TV111040416, TV111040418, TV111040419, TV111040420, TV111040423, TV111040424, TV111040428, TV111040432, TV111040438, TV111040703, TV111040729, TV111040735, TV111040803, TV111040808, TV111040830, TV111041118, TV111041304, TV111041306, TV111041435, TV111041436, TV211032405, TV211032409, TV211032412, TV211032413, TV211032414, TV211032415, TV211032419, TV211032420, TV211032430, TV211032431