Skip to content

Trilogy Trilogy100 Ventilator, AustraliaRecalls & reported problems

Respironics, Inc. · Trilogy 100 Ventilator, Australia · Class 2

All models of Philips Trilogy Ventilator (90) · Trilogy (46)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Respironics Trilogy ventilator has 97 adverse event reports describing malfunctions, with 1 injury and no deaths. Most describe operational or battery and charging issues found during servicing, usually with the device not in patient use. 1 report mentions low blood pressure and loss of consciousness. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
97
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 97 adverse event reports that list this exact device (DI 00606959021662), received between Oct 14, 2015 and Dec 21, 2021 — classified as 1 injury, 96 malfunctions.

The most frequently reported specific device problem is “Circuit Failure” (15 reports).

8 other recalls cover devices cleared under the same FDA submission (K083526); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available15
  • Circuit Failure15
  • Device Operational Issue14
  • Failure to Charge13
  • Device Displays Incorrect Message10
  • Degraded9
  • Failure to Recalibrate9
  • Battery Problem8

Patient problems

  • No Patient Involvement62
  • No Clinical Signs, Symptoms or Conditions26
  • No Consequences Or Impact To Patient13
  • No Known Impact Or Consequence To Patient5
  • Loss of consciousness1
  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 25 most recent of 97 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 21, 2021Failure to Recalibrate; Failure to Recalibrate

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A TEST STEP DURING TESTING. THE DEVICE'S POWER MANAGEMENT BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13051052 ↗
  • MalfunctionSep 28, 2021Circuit Failure; Circuit Failure

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WOULD NOT CYCLE PROPERLY. THE DEVICE WAS NOT IN PATIENT USE. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S POWER MANAGEMENT BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12542387 ↗
  • MalfunctionSep 28, 2021Failure to Charge; Circuit Failure; Failure to Charge; Circuit Failure

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ERROR CODE RELATED TO THE CHARGING OF THE INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S INTERNAL BATTERY AND POWER MANAGEMENT BOARD WERE REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12540873 ↗
  • MalfunctionAug 27, 2021Failure to Recalibrate; Failure to Recalibrate

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12377662 ↗
  • MalfunctionAug 24, 2021Failure to Charge; Battery Problem; Failure to Charge; Battery Problem

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ERROR CODE RELATED TO THE CHARGING OF THE INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12357846 ↗
  • MalfunctionAug 24, 2021Circuit Failure; Degraded; Circuit Failure; Degraded

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12357845 ↗
  • MalfunctionAug 11, 2021Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A VENTILATOR'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12307028 ↗
  • MalfunctionJun 30, 2021Failure to Charge; Circuit Failure; Failure to Charge; Circuit Failure

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR'S BATTERY WAS DEPLETED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER BUT THE EVALUATION IS NOT YET COMPLETE. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12093102 ↗
  • MalfunctionApr 21, 2021Circuit Failure; Failure to Recalibrate; Circuit Failure; Failure to Recalibrate

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG AND THE DEVICE FAILED STEPS DURING TESTING. THE DEVICE'S SENSOR BOARD WAS REPLACED, AND THE DEVICE'S SOFTWARE WAS RELOADED TO ADDRESS THE ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11701604 ↗
  • MalfunctionApr 12, 2021Battery Problem; Battery Problem

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A FAILURE OF THE DEVICE TO CHARGE ITS INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11649701 ↗
  • MalfunctionApr 8, 2021Failure to Recalibrate; Failure to Recalibrate

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A TEST STEP DURING TESTING. THE DEVICE'S POWER MANAGEMENT BOARD AND SYSTEM BOARD WERE REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11633322 ↗
  • MalfunctionApr 8, 2021Failure to Recalibrate; Failure to Recalibrate

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A TEST STEP DURING TESTING. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11633320 ↗
  • MalfunctionApr 6, 2021Failure to Charge; Failure to Charge

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTATIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ERROR CODE RELATED TO THE CHARGING OF THE INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11619050 ↗
  • MalfunctionApr 6, 2021Circuit Failure; Circuit Failure

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "VENTILATOR INOPERATIVE" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11617871 ↗
  • MalfunctionDec 11, 2020Circuit Failure; Circuit Failure

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTATIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10991639 ↗
  • MalfunctionDec 7, 2020Failure to Charge; Circuit Failure; Mechanical Problem; Battery Problem; Failure to Charge; Circuit Failure; Mechanical Problem; Battery Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WOULD NOT POWER ON. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S POWER SUPPLY WAS REPLACED TO ADDRESS THE ISSUE. ERROR CODES RELATED TO THE CHARGING OF THE INTERNAL BATTERY WERE OBSERVED. THE DEVICE'S INTERNAL BATTERY AND POWER MANAGEMENT BOARD WERE REPLACED TO ADDRESS THE ISSUES. THE TECHNICIAN OBSERVED THE DEVICE'S FLOW SENSOR ASSEMBLY WAS LOOSE DURING EVALUATION. THE FLOW SENSOR ASSEMBLY WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10960691 ↗
  • MalfunctionDec 7, 2020Failure to Power Up; Failure to Power Up

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ISSUE RELATED TO THE DEVICE FAILING TO POWER ON WAS OBSERVED. THE DEVICE'S POWER MANAGEMENT BOARD, SYSTEM BOARD AND FRONT PANEL BOARD WERE REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10960491 ↗
  • MalfunctionDec 4, 2020No Display/Image; No Display/Image

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE'S DISPLAY WAS BLANK. THE DEVICE'S LCD SCREEN WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10949249 ↗
  • MalfunctionNov 23, 2020Circuit Failure; Degraded; Circuit Failure; Degraded

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO PATIENT HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10886283 ↗
  • MalfunctionOct 12, 2020Circuit Failure; Circuit Failure

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S POWER MANAGEMENT BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10662447 ↗
  • InjurySep 29, 2020Failure to Cycle; Failure to Cycle

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING THAT ON (B)(6) 2019 A PATIENT WAS FOUND UNRESPONSIVE AND HYPOTENSIVE ON THE FLOOR WHILE THE VENTILATOR WAS ALARMING FOR LOW MINUTE VENTILATION AND CIRCUIT DISCONNECT. THE PATIENT WAS NOT DISCONNECTED FROM THE CIRCUIT. THE PATIENT RECEIVED MANUAL RESUSCITATION WITH A MANUAL RESUSCITATION BAG, WAS SUCTIONED AND THEN REGAINED CONSCIOUSNESS. ON (B)(6) 2019, THE CUSTOMER ALLEGED THAT WHEN THE PATIENT WAS CONNECTED TO THE SAME VENTILATOR, THE VENTILATOR FAILED TO IMMEDIATELY DELIVER BREATHS TO THE PATIENT. THE PATIENT WAS THEN GIVEN BREATHS WITH A MANUAL RESUSCITATION BAG UNTIL THE VENTILATOR STARTED CYCLING. THE PATIENT WAS THEN PLACED BACK ON THE VENTILATOR WITHOUT INCIDENT. THE DEVICE WAS RETURNED TO THE MANUFACTURER'S PRODUCT INVESTIGATION LABORATORY FOR INVESTIGATION. THE MANUFACTURER WAS UNABLE TO DUPLICATE THE REPORTED EVENTS. THE DEVICE MAINTAINS EVENT LOGS IN NON-VOLATILE MEMORY. AN EVENT, AS LOGGED BY THE DEVICE, INCLUDES EVERY KEYPRESS, ALARM OR FAILURE OF THE DEVICE TO REMAIN WITHIN SPECIFICATIONS. THERE WERE NO EVENTS LOGGED RELATED TO DEVICE PERFORMANCE OR DEVICE MALFUNCTIONS. IF A DEVICE MALFUNCTION OCCURRED, THE EVENT LOGS WO

    Patient problems: Loss of consciousness; Low Blood Pressure/ Hypotension

    MAUDE report 10600250 ↗
  • MalfunctionApr 23, 2019Circuit Failure; Circuit Failure

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "VENTILATOR INOPERATIVE" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Patient Involvement

    MAUDE report 8539585 ↗
  • MalfunctionApr 23, 2019Circuit Failure; Degraded; Circuit Failure; Degraded

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 8539584 ↗
  • MalfunctionApr 2, 2019Failure to Recalibrate; Failure to Recalibrate

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A TEST STEP DURING TESTING. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Patient Involvement

    MAUDE report 8473504 ↗
  • MalfunctionApr 2, 2019Failure to Recalibrate; Failure to Recalibrate

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A TEST STEP DURING TESTING. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Patient Involvement

    MAUDE report 8473496 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083526. One submission often covers many models, so check the product and lot details in each record.

  • Z-0884-2023 · Dec 7, 2022 · Open, Classified
    Philips Garbin Ventilator, Model Number 1058180B
  • Z-0882-2023 · Dec 7, 2022 · Open, Classified
    Philips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B, R1054260, RSL1054260, CA1054096, R1054260B, R1054655
  • Z-1956-2021 · Jun 14, 2021 · Open, Classified
    Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators.
  • Z-1065-2016 · Jan 18, 2016 · Terminated
    Trilogy 100, Trilogy 200, Trilogy O2, Trilogy 202, Trilogy EC, Garbin, Garbin Plus Trilogy 100, Trilogy 200, Garbin, Garbin plus provides continuous or intermittent ventilatory su
  • Z-1143-2014 · Feb 11, 2014 · Terminated
    Trilogy Ventilators The Respironics Trilogy 100 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilog
  • Z-2024-2012 · Apr 27, 2012 · Terminated
    Respironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous or intermittent ventilatory support for the care of individuals who require m
  • Z-0296-2012 · Oct 11, 2011 · Terminated
    Trilogy 100, 200 and 202 Ventilators The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used fo
  • Z-1896-2010 · Jun 10, 2010 · Terminated
    Trilogy 100 Ventilator, Model/Catalog Number 1054260, U1054260 (Domestic). Philips Respironics, Monroeville, PA 15146. Provides continuous or intermittent ventilatory support for

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page