Roche Cobas c 111 Analyzer recallZ-0311-2012 · Roche Diagnostics Operations, Inc. · initiated Oct 18, 2011

Terminated All recalls posted in November 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Internal investigations have revealed a potential printer malfunction on the Cobas c 111 analyzer, where one or more numbers or characters can be missing on the hard copy result printout, or the printer may skip leaving blanks which may result in an incorrect print layout. The missing number of characters can be in any position, therefore it could be in the middle of a result, (e.g. 48.53 mmol in

FDA root cause category: Software design

What the company is doing

Roche Diagnostics Operations, Inc sent a Urgent Medical Device Correction dated October 18, 2011, via UPS Ground (receipt required) to all affected customers and one distribution center, with instructions to disable the Auto Print Result feature until new software is available. All printed results will need to be confirmed against the results indicated on the display screen. Questions are to be directed to Roche Diagnostics Technical Support at 1-800 428-2336,

Recall details

Recall number
Z-0311-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 18, 2011
Posted by the FDA
Nov 30, 2011
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
135
Product code
JJE
FDA submissions
K071211
Distribution
Worldwide Distribution - USA including West Virginia and Texas.

Product description

Roche Cobas c 111 Analyzer The Roche Cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Serial Numbers 50865, 50626, 50798, 50731, 50654, 50541, 51023, 50946, 50989, 50634, 50506, 50251, 50887, 50948, 50453, 50688, 51111, 50997, 51122, 50244, 50790, 50475, 50597, 50740, 50656, 50793, 51086, 50999, 51146, 50635, 50744, 50586, 50632, 50846, 1167, 50693, 50787, 50670, 50672, 51128, 50941, 50562, 50788, 50547, 51003, 50624, 50677, 50678, 50814, 51015, 50687, 50680, 50894, 50560, 51004, 50247, 50538, 51118, 50926, 50847, 50841, 51127, 50741, 50764, 50633, 51106, 51116, 50201, 51124, 50612, 51088, 50576, 51016, 50689, 50681, 50424, 51126, 50959, 50878, 50784, 50661, 50761, 50452, 50450, 50290, 50947, 51327, 50630, 50985, 50763, 50319, 50984, 50250, 50549, 50925, 50844, 50660, 50845, 50683, 50735, 51056, 50463, 50775, 50840, 50734, 50655, 50843, 50758, 50852, 50539, 50651, 50653, 50657, 50964, 50563, 50742, 50738, 50727, 50934, 50561, 50803, 51000, 51107, 50686, 50815, 50649, 50470, 50759, 50644, 51053, 50739, 50595, 51087, 50575, and 50679.

Other devices under the same FDA submission

2 devices cleared under K071211 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page