Glucose HK 4657527190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
All models of Glucose HK (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04015630922338), received between Apr 11, 2024 and Apr 27, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (5 reports).
3 other recalls cover devices cleared under the same FDA submission (K071211); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results5
- High Test Results2
- Incorrect Measurement1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 27, 2026Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH GLUCOSE HK ON A COBAS C 111 ANALYZER. THE SAMPLE INITIALLY RESULTED IN A GLUCOSE VALUE OF 254.3 MG/DL. A PHYSICIAN REQUESTED A REPEAT OF THE SAMPLE. THE REPEAT RESULT WAS 124.7 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25001607 ↗MalfunctionMar 17, 2026Incorrect Measurement
THE INITIAL REPORTER COMPLAINED OF DISCREPANT HIGH RESULTS FOR 1 PATIENT SAMPLE TESTED FOR GLUCOSE HK ON A COBAS C 111 ANALYZER. THE INITIAL RESULT WAS "OVER" 200 MG/DL. THE SPECIFIC RESULT COULD NOT BE PROVIDED. THE REPEAT RESULT WAS 236.19 MG/DL. ON (B)(6) 2026, THE CUSTOMER RECEIVED A COMPLAINT THAT THE PATIENT¿S GLUCOSE RESULT FROM A DIFFERENT LABORATORY WAS "NORMAL." THE SPECIFIC RESULT WAS NOT PROVIDED. THE CUSTOMER REPEATED THE ORIGINAL SAMPLE, AND THE RESULT WAS 79.68 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615367 ↗MalfunctionFeb 19, 2026Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PEDIATRIC PATIENT SAMPLE TESTED WITH GLUCOSE HK ON A COBAS C 111 ANALYZER. THE SAMPLE INITIALLY RESULTED IN A GLUCOSE VALUE OF 7.7 MMOL/L. THE SAMPLE WAS REPEATED TWICE, RESULTING IN VALUES OF 5.5 MMOL/L AND 5.45 MMOL/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24388211 ↗MalfunctionFeb 6, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE GLUCOSE HK RESULT FROM THE COBAS C 111 ANALYZER. THE INITIAL RESULT WAS 184 MG/DL. THE REPEAT RESULT WAS 93 MG/DL. THIS RESULT WAS BELIEVED TO BE CORRECT. THE USER REPEATED THE SAMPLE 5 TIMES AND THE RESULTS WERE ALL 106-107 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21322452 ↗MalfunctionAug 22, 2024High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE GLUC3 GLUCOSE HK GEN 3 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 111 ANALYZER. THE CUSTOMER ALIQUOTED SERUM INTO AN EPPENDORF TUBE ON THE TOP OF THE PRIMARY SAMPLE TUBE AND THE INITIAL GLUCOSE RESULT WAS 666.8 MG/DL. ANOTHER SERUM SAMPLE COLLECTED AT THE SAME TIME WITH ALIQUOTED SERUM IN AN EPPENDORF TUBE ON TOP OF THE PRIMARY TUBE WAS TESTED AND THE RESULT WAS 167.3 MG/DL. A HEPARIN PLASMA SAMPLE COLLECTED AT THE SAME TIME AS THE INITIAL SERUM SAMPLE WAS TESTED AND THE GLUCOSE RESULT WAS 165.9 MG/DL. THE INITIAL PRIMARY SERUM SAMPLE WAS REPEATED AGAIN WITHOUT AN EPPENDORF TUBE AND THE RESULT WAS 168.8 MG/DL. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20033458 ↗MalfunctionJun 7, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE GLUC3 GLUCOSE HK GEN.3 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 111 ANALYZER. THE INITIAL GLUCOSE RESULT OF 536.54 MG/DL. THE SAMPLE WAS REPEATED TWICE AND THE RESULTS WERE 273.71 MG/DL AND 272.79 MG/DL. THE REPEAT RESULTS WERE DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19486657 ↗MalfunctionApr 11, 2024Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
THE CUSTOMER STATED THAT FOR TWO WEEKS, THEY HAVE BEEN HAVING ISSUES WITH QUESTIONABLE PATIENT RESULTS FOR GLUCOSE HK ON A COBAS C 111 ANALYZER. THE CUSTOMER PROVIDED TWO EXAMPLES OF PATIENT SAMPLES WITH DISCREPANT GLUCOSE RESULTS. THE FIRST SAMPLE INITIALLY RESULTED IN A GLUCOSE VALUE OF 0.3 MMOL/L AND IT REPEATED AS 0.4 MMOL/L. A NEW REAGENT BOTTLE OF THE SAME LOT WAS THEN USED TO REPEAT THE SAMPLE ON (B)(6) 2024 AND IT RECOVERED WITH A GLUCOSE VALUE OF 4.9 MMOL/L. THE SECOND SAMPLE INITIALLY RESULTED IN A GLUCOSE VALUE OF 0.6 MMOL/L. A NEW REAGENT BOTTLE OF THE SAME LOT WAS THEN USED TO REPEAT THE SAMPLE ON (B)(6) 2024 AND IT RECOVERED WITH A GLUCOSE VALUE OF 10.0 MMOL/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19087056 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071211. One submission often covers many models, so check the product and lot details in each record.
- Z-0387-2017 · Sep 27, 2016 · TerminatedRoche COBAS INTEGRA c111 Analyzer, Chemistry (Photometric, Discrete), for clinical use Product Usage: The Roche COBAS INTEGRA c111 analyzer is an in-vitro diagnostic analyzer capa
- Z-0311-2012 · Oct 18, 2011 · TerminatedRoche Cobas c 111 Analyzer The Roche Cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analy
- Z-1009-2009 · Nov 17, 2008 · TerminatedRoche COBAS c 111 Analyzer with ISE module option, GMMI No. 04777433001; Roche Diagnostics, Indianapolis, IN. Is an in-vitro diagnostic analyzer capable of performing clinical che
Other HEXOKINASE, GLUCOSE devices
- Glucose HK 4404483190Roche Diagnostics GmbH · HEXOKINASE, GLUCOSE30 reports
- Glucose HK 08057800190Roche Diagnostics GmbH · HEXOKINASE, GLUCOSE19 reports
- ARCHITECT 3L82-21Abbott Laboratories · HEXOKINASE, GLUCOSE14 reports
- Glucose HK Gen3 5168791190Roche Diagnostics GmbH · HEXOKINASE, GLUCOSE10 reports
- ARCHITECT 3L82-22Abbott GmbH · HEXOKINASE, GLUCOSE8 reports
- ARCHITECT 3L82-42Abbott GmbH · HEXOKINASE, GLUCOSE4 reports
- Alinity 07P5530Abbott GmbH · HEXOKINASE, GLUCOSE3 reports
- Alinity 07P5520Abbott GmbH · HEXOKINASE, GLUCOSE3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.