Draeger Infinity(R) Delta, Delta XL, Gamma XXL and Kappa… recallZ-0281-2012 · Draeger Medical, Inc. · initiated Oct 31, 2011
Terminated All recalls posted in November 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm received 2 complaints reporting Ventricular Fibrillation (Vfib) arrhythmias were not recognized by the monitor and did not alarm when the Masimo SET pod is used for SpO2 measurements.
FDA root cause category: Software design
What the company is doing
Firm initiated their recall of this device on October 25, 2011 by sending an "Urgent Medical Device Recall" letter to consignees. A Technical Service Bulletin will be generated and distributed globally that instructs the affected International Draeger subsidiaries to inform customers in their countries. The software anomaly has been corrected in software version 7.3. Draeger Service Representatives will schedule time with customers to upgrade the software in affected monitors. The software upgrade will be provided free of charge.
Recall details
- Recall number
- Z-0281-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 31, 2011
- Posted by the FDA
- Nov 23, 2011
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 14,894
- Product code
- MHX
- FDA submissions
- K070566
- Distribution
- Worldwide distribution- USA including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SD, TN, TX, UT, VA, WI and WV and International distribution including the countries of Algeria, Angola, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brasilien, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Deutschland, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Finland, France, Frankreich, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indien, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Jordanien, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Madagascar, Malaysia, Malta, Mauritius, Mayotte, Mexico, Moldavia, Moldova, Morocco, Netherlands, Norway, Norway, Peru, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Rumania, Russian Fed., Saudi Arabia, Serbia, Seychelles, Singapore, Slovenia, Slowenien, South Africa, South Korea, Spain, Sri Lanka, S¿dafrika, Sweden, Switzerland, Syria, Taiwan, Thailand, Tschechische Republic, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and White Russia.
Product description
Draeger Infinity(R) Delta, Delta XL, Gamma XXL and Kappa Monitors are used to monitor, physiologically, the vital signs of patients (with Arrhythmia Detection or Alarms).
Code information (models, lots, serial numbers)
Catalog Number(s): MS18597 and MS18852, Software versions: VF6.2, VF6.3, VF6.4, VF7.1 and VF7.2
Other devices under the same FDA submission
5 devices cleared under K070566 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Delta Monitor MS18597Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2 deaths11 reports
- DeltaXL Monitor MS18596Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 reports
- Infinity Vista XL Monitor MS18986Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 death4 reports
- Infinity Kappa Monitor MS18853Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)4 reports
- Infinity Gamma XXL MS18852Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page