CAT 375-941-012***STRYKER***Arthroscopic Burr***4.0mm… recallZ-0961-2012 · Ascent Healthcare Solutions · initiated Jan 26, 2009
Terminated All recalls posted in February 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ascent Healthcare Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On 01/26/2009, Ascent Healthcare Solutions, Lakeland, FL initiated a recall of various models of Dyonics, Stryker, and Linvatech Reprocessed Arthroscopic Shaver Blades. The sterility of the devices may be compromised due to a potential breach in the seal of the Tyvek cover to the plastic tray.
FDA root cause category: Packaging process control
What the company is doing
Ascent Healthcare Solutions sent an "URGENT MEDICAL DEVICE VOLUNTARY RECALL" letter dated January 26, 2009 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to examine their inventory, remove and quarantine all affected devices. A Recall Effectiveness Check Form was attached for customers to complete and return via fax to 1-866-325-4760. Contact the firm at (888) 888-3433 ext. 5519 for questions regarding this recall.
Recall details
- Recall number
- Z-0961-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 26, 2009
- Posted by the FDA
- Feb 3, 2012
- Recalling firm
- Ascent Healthcare Solutions
- Quantity
- 40 devices
- Product code
- HRX
- FDA submissions
- K012605, K012635
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NH, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WI, and WY, and the country of Canada.
Product description
***CAT 375-941-012***STRYKER***Arthroscopic Burr***4.0mm BARREL BUR***Hollow 12 Flute***RED***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE SOLUTIONS***Lakeland, FL*** Lot number: 195182***Expiration Date: 11/2011. ***CAT 375-951-112***STRYKER***Arthroscopic Burr***5.0mm BARREL BUR***Hollow 12 Flute***BLUE***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE SOLUTIONS***Lakeland, FL*** Lot number: 194245***Expiration Date: 11/2011. Reprocessed arthroscopic shavers are indicated for use in orthopedic surgical procedures of the joints, jaw or sinuses where cutting and removal of soft and hard tissue or bone is needed in patients requiring orthopedic surgery.
Code information (models, lots, serial numbers)
Lot numbers: 195182 and 194245.
Other devices under the same FDA submission
7 devices cleared under K012605, K012635 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- N/AStryker Sustainability Solutions, Inc. · ARTHROSCOPE87 reports
- N/AStryker Sustainability Solutions, Inc. · ARTHROSCOPE41 reports
- N/AStryker Sustainability Solutions, Inc. · ARTHROSCOPE1 report
- N/AStryker Sustainability Solutions, Inc. · ARTHROSCOPE1 report
- N/AStryker Sustainability Solutions, Inc. · ARTHROSCOPE1 report
- N/AStryker Sustainability Solutions, Inc. · ARTHROSCOPE1 report
- N/AStryker Sustainability Solutions, Inc. · ARTHROSCOPE1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 58937).
- Z-0954-2012 · Terminated***CAT 275-542-000***STRYKER***Arthroscopic Blade***4.0mm RESECTOR***Red***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE SOLUTIONS***Lakeland, FL*** Lot nu…
- Z-0955-2012 · Terminated***CAT 275-941-000***STRYKER***Arthroscopic Burr***4.0mm BARREL BUR Hollow 6 Flute***Red***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE SOLUTIONS***Lakela…
- Z-0956-2012 · Terminated***CAT 375-545-000***STRYKER***Arthroscopic Blade***4.0mm TOMCAT CUTTER***Hub Style - Formula***Red/White***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE S…
- Z-0957-2012 · Terminated***CAT 375-544-000***STRYKER***Arthroscopic Blade***4.0mm AGGRESSIVE PLUS***Hub Style - Formula***Red/Lt Blue***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCA…
- Z-0958-2012 · Terminated***CAT 275-534-000***STRYKER***Arthroscopic Blade***3.5mm AGGRESSIVE PLUS***Yellow/Blue***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE SOLUTIONS***Lakelan…
- Z-0959-2012 · Terminated***CAT 375-532-000***STRYKER***Arthroscopic Blade***3.5mm RESECTOR***Hub Style - Formula***Yellow/Black***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE SOL…
- Z-0960-2012 · Terminated***CAT 375-941-000***STRYKER***Arthroscopic Bur***4.0mm BARREL BUR***Hollow 6 Flute***RED***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE SOLUTIONS***Lakel…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page