VenaCure EVLT NeverTouch - FRS PROCEDURE KIT Featuring… recallZ-0402-2012 · Angiodynamics Worldwide Headquarters · initiated Nov 10, 2011
Terminated All recalls posted in January 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Angiodynamics Worldwide Headquarters or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an in-process inspection the hemostasis sidearm assemblies were identified to have the potential for flash to exist on the housing at the base of the threads. This flash has the potential to break off and relocate within the fluid path.
FDA root cause category: Process control
What the company is doing
AngioDynamics sent an URGENT: Medical Device Recall letters and response form (dated November 10, 2011) to all affected customers. The letter identified the affected products, the problem, and the actions to be taken by customer/user. The customers were instructed to do the following: 1) Identify, segregate and return the recalled products to ANGIODYNAMICS and 2) Complete the enclosed Recall Reply Form even if they do not have any products remaining in their possession, noting zero quantity to be returned and fax to 518-798-1360. The letter states that a returned authorization number (RA#) will be assigned to each customer in the recall notification letter, an UPS account will be provided for return freight and a replacement product will be shipped upon receipt and confirmation of the returned product. If you have any additional questions or concerns or need more detailed instruction on how to comply with this notice, please do not hesitate to contact your local sales representative or call the Customer Service Manager, at ANGIODYNAMICS Customer Service at 1-800-772-6446. You may also e-mail your questions to [email protected].
Recall details
- Recall number
- Z-0402-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 10, 2011
- Posted by the FDA
- Jan 11, 2012
- Recalling firm
- Angiodynamics Worldwide Headquarters
- Quantity
- 3165 units
- Product code
- GEX
- FDA submissions
- K102796
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Canada, Great Britain, France, India, New Zealand, Netherlands, Saudi Arabia, Turkey, and Singapore
Product description
VenaCure EVLT NeverTouch - FRS PROCEDURE KIT Featuring Gold-Tipped Fiber, 45 cm, NeverTouch 45cm Kit w/Gripper and RFID T, Catalog No./REF 11403001, STERILE -- AngioDynamics, Inc. 603 Queensbury, NY USA 12804. Product Usage: The AngioDynamics, Inc. VenaCure Procedure Kit is indicated for endovascular coagulation of the Great Saphenous Vein (GSV) in patients with superficial vein reflux, for the treatment of varicose veins and varicosities associated with superficial reflux of the Great Saphenous Vein (GSV), and for the treatment of incompetence and reflux of superficial veins of the lower extremity.
Code information (models, lots, serial numbers)
Batch/Lot Numbers: 549604, 549605, 549607, 549806, 549991, 550848, 551098, 551099, 551100, 551101, 551102, 551503, 551970, 551971, 551972, 551973, 551974, 552939, 552960, 553299, 553300, 553301, 553302, 553583, 553605, 553937, 553938, 553939, 554040, 554041, 554267, 554268, 554269, 554270, 554624
Other devices under the same FDA submission
1 device cleared under K102796 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VenaCure EVLT 51403002Angiodynamics, Inc. · Powered laser surgical instrument6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 60467).
- Z-0400-2012 · TerminatedVenaCure EVLT NeverTouch PROCEDURE KIT Featuring Gold-Tipped Fiber, 45 cm, NeverTouch Kit 45CM Internal Sheath, Catalog No./REF 11402001, STERILE -- AngioDynamics, Inc. 603 Queensb…
- Z-0401-2012 · TerminatedVenaCure EVLT NeverTouch PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, NeverTouch Kit 65CM Internal Sheath, Catalog No./REF 11402002, STERILE -- AngioDynamics, Inc. 603 Queensb…
- Z-0403-2012 · TerminatedVenaCure EVLT NeverTouch - FRS PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, NeverTouch 65cm Kit w/Gripper and RFID T, Catalog No./REF 11403002, STERILE -- AngioDynamics, Inc. …
- Z-0404-2012 · TerminatedVenaCure EVLT NeverTouch - FRS .018" System PROCEDURE KIT Featuring Gold-Tipped Fiber, 25 cm, VC EVLT NeverTouch-FRS 0.18 25 cm Kit, Catalog No./REF 11403003, STERILE -- AngioDynam…
- Z-0405-2012 · TerminatedVenaCure EVLT NeverTouch - FRS PROCEDURE KIT Featuring Gold-Tipped Fiber, 90 cm, VC EVLT NeverTouch-FRS 90cm Kit, Catalog No./REF 11403004, STERILE -- AngioDynamics, Inc. 603 Queen…
- Z-0406-2012 · TerminatedVenaCure EVLT NeverTouch - FRS w/19 ga Needle PROCEDURE KIT Featuring Gold-Tipped Fiber, 45 cm, NT FRS 45cm Kit w/19Ga Needle, Catalog No./REF 11403005, STERILE -- AngioDynamics, I…
- Z-0407-2012 · TerminatedVenaCure EVLT NeverTouch - FRS w/19 ga Needle PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, NT FRS 65cm Kit w/19Ga Needle, Catalog No./REF 11403006, STERILE -- AngioDynamics, I…
- Z-0408-2012 · TerminatedVenaCure EVLT NeverTouch - FRS .018 " System PROCEDURE KIT Featuring Gold-Tipped Fiber, 45 cm, 45cm NeverTouch FRS .018 Procedure Kit, Catalog No./REF 11403012, STERILE -- AngioDyn…
- Z-0409-2012 · TerminatedVenaCure EVLT NeverTouch - FRS .018 " System PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, 65cm NeverTouch FRS .018 Procedure Kit, Catalog No./REF 11403013, STERILE -- AngioDyn…
- Z-0410-2012 · TerminatedVenaCure EVLT NeverTouch PROCEDURE KIT Featuring Gold-Tipped Fiber, 45 cm, NeverTouch Procedure Kit 45cm, Catalog No./REF 51402007, STERILE -- AngioDynamics, Inc. 603 Queensbury, N…
- Z-0411-2012 · TerminatedVenaCure EVLT NeverTouch PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, NeverTouch Procedure Kit 65cm, Catalog No./REF 51402008, STERILE -- AngioDynamics, Inc. 603 Queensbury, N…
- Z-0412-2012 · TerminatedVenaCure EVLT NeverTouch-FRS PROCEDURE KIT Featuring Gold-Tipped Fiber, 45 cm, VenaCure NeverTouch FRS 45cm Clear Needle, Catalog No./REF 51403001, STERILE -- AngioDynamics, Inc. 6…
- Z-0413-2012 · TerminatedVenaCure EVLT NeverTouch-FRS PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, VenaCure NeverTouch FRS 65cm Clear Needle, Catalog No./REF 51403002, STERILE -- AngioDynamics, Inc. …
- Z-0414-2012 · TerminatedVenaCure EVLT NeverTouch-FRS PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, VC NT FRS 65cm Clear Non-Echo Needle, Catalog No./REF 51403005, STERILE -- AngioDynamics, Inc. 603 Qu…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page