Skip to content

VenaCure EVLT NeverTouch-FRS PROCEDURE KIT Featuring… recallZ-0413-2012 · Angiodynamics Worldwide Headquarters · initiated Nov 10, 2011

Terminated All recalls posted in January 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Angiodynamics Worldwide Headquarters or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During an in-process inspection the hemostasis sidearm assemblies were identified to have the potential for flash to exist on the housing at the base of the threads. This flash has the potential to break off and relocate within the fluid path.

FDA root cause category: Process control

What the company is doing

AngioDynamics sent an URGENT: Medical Device Recall letters and response form (dated November 10, 2011) to all affected customers. The letter identified the affected products, the problem, and the actions to be taken by customer/user. The customers were instructed to do the following: 1) Identify, segregate and return the recalled products to ANGIODYNAMICS and 2) Complete the enclosed Recall Reply Form even if they do not have any products remaining in their possession, noting zero quantity to be returned and fax to 518-798-1360. The letter states that a returned authorization number (RA#) will be assigned to each customer in the recall notification letter, an UPS account will be provided for return freight and a replacement product will be shipped upon receipt and confirmation of the returned product. If you have any additional questions or concerns or need more detailed instruction on how to comply with this notice, please do not hesitate to contact your local sales representative or call the Customer Service Manager, at ANGIODYNAMICS Customer Service at 1-800-772-6446. You may also e-mail your questions to [email protected].

Recall details

Recall number
Z-0413-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 10, 2011
Posted by the FDA
Jan 11, 2012
Recalling firm
Angiodynamics Worldwide Headquarters
Quantity
186 units
Product code
GEX
FDA submissions
K102796
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Canada, Great Britain, France, India, New Zealand, Netherlands, Saudi Arabia, Turkey, and Singapore

Product description

VenaCure EVLT NeverTouch-FRS PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, VenaCure NeverTouch FRS 65cm Clear Needle, Catalog No./REF 51403002, STERILE -- AngioDynamics, Inc. 603 Queensbury, NY USA 12804. This is a Special Order. Product Usage: The AngioDynamics, Inc. VenaCure Procedure Kit is indicated for endovascular coagulation of the Great Saphenous Vein (GSV) in patients with superficial vein reflux, for the treatment of varicose veins and varicosities associated with superficial reflux of the Great Saphenous Vein (GSV), and for the treatment of incompetence and reflux of superficial veins of the lower extremity.

Code information (models, lots, serial numbers)

Batch/Lot Numbers: 551253, 553312, 554789

Other devices under the same FDA submission

1 device cleared under K102796 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 60467).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page