Baxter Interlink System Extension Set, 13" (32 cm), Check… recallZ-0475-2012 · Baxter Healthcare Corp. · initiated Dec 6, 2011
Terminated All recalls posted in January 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection.
FDA root cause category: Employee error
What the company is doing
The firm, Baxter, sent an "Urgent Product Recall" letter dated December 6, 2011, via first class mail to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to inspect all locations where their facility's inventory is located for products with the affected lot numbers; return the affected product; request a credit or replacement by calling Baxter Healthcare Center for Service at 1-888-229-0001 between 7:00am to 6:00pm Central time, Monday through Friday; complete and return the attached Customer Reply Form via fax to: 1-847-270-5457, or scan and email to [email protected]; and if they are a dealer, wholesaler, or distributor/reseller that distributed any affected product to other facilities, please notify their customers of this recall. If you have any questions regarding the communication, please call The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.
Recall details
- Recall number
- Z-0475-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Dec 6, 2011
- Posted by the FDA
- Jan 11, 2012
- Recalling firm
- Baxter Healthcare Corp.
- Quantity
- 4,560 units
- Product code
- FPA
- FDA submissions
- K961225
- Distribution
- Worldwide distribution: USA (nationwide) including states of: Alaska, California, Colorado, Florida, Georgia, Hawaii, Illinois, Kansas, Kentucky, Massachusetts, Maine, Michigan, Minnesota, Missouri, North Carolina, New Hampshire, New Jersey, New York, Ohio, Pennsylvania, Rhode Island, South Carolina, Tennessee, Virginia, Washington, Wisconsin and West Virginia; and country of: New Zealand.
Product description
Baxter Interlink System Extension Set, 13" (32 cm), Check Valve, Injection Site, Male Luer Lock Adapter; an Rx sterile, nonpyrogenic fluid pathway for the administration of sterile I.V. fluids to the patient; Baxter Healthcare Corporation, Deerfield, IL 60015 USA; product code 2C6605 Intended use: for the administration of drug and solutions to patient's vascular system through a vascular access device
Code information (models, lots, serial numbers)
product code 2C6605, lot number GR295725
Other devices under the same FDA submission
24 devices cleared under K961225 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CLEARLINK/CONTINU-FLO/DUO-VENT 2C8541Baxter Healthcare Corporation · Set, administration, intravascular2 recalls797 reports
- CLEARLINK/CONTINU-FLO 2C8537Baxter International Inc. · Set, administration, intravascular2 recalls490 reports
- CLEARLINK/CONTINU-FLO 2C8519Baxter Healthcare Corporation · Set, administration, intravascular3 recalls1 death221 reports
- CLEARLINK/CONTINU-FLO 2R8537Baxter International Inc. · Set, administration, intravascular1 recall136 reports
- CLEARLINK/CONTINU-FLO/DUO-VENT UC8519Baxter International Inc. · Set, administration, intravascular1 recall66 reports
- CLEARLINK/CONTINU-FLO 2H8519Baxter Healthcare Corporation · Set, administration, intravascular93 reports
- CLEARLINK/CONTINU-FLO 2R8519Baxter International Inc. · Set, administration, intravascular1 recall37 reports
- CLEARLINK/CONTINU-FLO 2R8546Baxter International Inc. · Set, administration, intravascular59 reports
- CLEARLINK/DUO-VENT 2C7462Baxter Healthcare Corporation · Set, administration, intravascular57 reports
- DUO-VENT 1C8507Baxter Healthcare Corporation · Set, administration, intravascular1 recall4 reports
- CLEARLINK 2C7461Baxter Healthcare Corporation · Set, administration, intravascular49 reports
- CLEARLINK/DUO-VENT 2H7462Baxter International Inc. · Set, administration, intravascular47 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 60608).
- Z-0471-2012 · TerminatedBaxter Interlink System Extension Set, 20" (51 cm), Vol. 2.9 mL, Injection Site, Male Luer Lock Adapter; an Rx sterile, nonpyrogenic fluid pathway for the administration of sterile…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page