Bivona TTS Cuffed Neonatal Tracheostomy Tube Sizes 2.5 -… recallZ-0657-2012 · Smiths Medical ASD, Inc. · initiated Nov 28, 2011
Terminated All recalls posted in January 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Difficulty disconnecting accessories from the Bivona Uncuffed Neonatal, Pediatric and Flextend Tracheostomy Tubes TR3 connector, which may result in a delay in initiating therapy or patient discomfort. If excessive force is used to detach the accessory, this may result in decannulation of the tube which could result in serious injury or death; especially if no replacement tracheostomy tube is imm
FDA root cause category: Device Design
What the company is doing
Smith Medical sent an Urgent Field Safety Notice dated November 28, 2011, via U.S. Mail to all affected consignees. The notice identified the product, description of problem, and actions to be taken. This notice included a Customer Information Bulletin and Urgent Field Safety Notice Confirmation Form. Customers were instructed to 1) identify all affected unused product in inventory and segregate it to a quarantine location; 2) complete and return the attached Confirmation Form by fax to 219-989-7259 or email to [email protected]; and 3) distributors were instructed to stop distributing and provide Smith Medical with a list of their customers, including contact details, so that Smith Medical can notify the customers of this notice. For questions contact Smiths Medical Customer Service at 800-258-5361, Option 1.
Recall details
- Recall number
- Z-0657-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 28, 2011
- Posted by the FDA
- Jan 19, 2012
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- US distribution 153,033 devices ; Foreign 21,490 devices
- Product code
- BTO
- FDA submissions
- K912469
- Distribution
- Worldwide Distribution - USA Nationwide: AZ, CA, CO, CT,DE, FL, GA, HI, IL, IN, KS, KY, LA, SC, TX, OR, NJ, MO, UT, MI, NY, NM, MA, MN, MS, NC, ND, NE, NH, NV, OH, OK, OR, RI, SD, TN, VA, VT, WA, WI, WV including Washinton DC, Australia, Austria, Barbados, Belgium, Brazil, Canada, Chile, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Gibraltar, Great Britain (UK), Greece, Hong Kong, Ireland, Israel, Italy, Kuwait, Libya, Malaysia, Netherlands, New Zeland, Norway, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates, United States, Vatican City State
Product description
Bivona TTS Cuffed Neonatal Tracheostomy Tube Sizes 2.5 - 4.0 (With a V-Shaped Neck Flange).
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Lot numbers: 1644864, 1671968, 1713941, 1754786, 1835043, 1856208, 1875201, 1890748, 1913138, 1636740, 1655827, 1658988, 1667772, 1675682, 1687749, 1701032, 1704636, 1729117, 1745607, 1778507, 1809051, 1809052, 1821983, 1848186, 1851435, 1875202, 1896186, 1900039, 1914915, 1631504, 1633571, 1648150, 1651810, 1663885, 1680985, 1667773, 1679964, 1683713, 1683714, 1704637, 1713942, 1704638, 1749575, 1749576, 1778395, 1778396, 1782715, 1793502, 1804358, 1809053, 1821984, 1835044, 1856209, 1864910, 1883542, 1883543, 1887779, 1908912, 1914916, 1631505, 1631506, 1636040, 1644869, 1658989, 1680981, 1671972, 1679965, 1683715, 1692758, 1713943, 1704639, 1721313, 1725330, 1725329, 1745609, 1793503, 1804359, 1814087, 1821985, 1825903, 1821986, 1838655, 1851436, 1878007, 1891571, 1890749, 1904090, 1914917, and 1919156. Reorder numbers: 67N025, 67N030, 67N035, and 67N040.
Other devices under the same FDA submission
26 devices cleared under K912469 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Bivona 60P045Icu Medical, Inc. · Tube tracheostomy and tube cuff2 recalls12 reports
- Bivona 60SP040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall20 reports
- Bivona 60SP035Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall3 deaths19 reports
- Bivona 60P055Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall16 reports
- Bivona 60SP050Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall13 reports
- Bivona 60P040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall12 reports
- Bivona 60SP045Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall8 reports
- Bivona 60P050Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall7 reports
- Bivona 60SN040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall5 reports
- Bivona 60SP055Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall5 reports
- Bivona 60SN035Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall4 reports
- Bivona 60SP030Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 60449).
- Z-0656-2012 · TerminatedBivona TTS Cuffed Neonatal Tracheostomy Tube Sizes 2.5 - 4.0 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheosto…
- Z-0658-2012 · TerminatedBivona Aire-Cuf Pediatric Tracheostomy Tube Sizes 2.5 - 5.5 Product Usage: This tube is intended to provide direct airway access for a tracheostomized patient for up to 29 days. …
- Z-0659-2012 · TerminatedBivona Aire-Cuf Pediatric Tracheostomy Tube Sizes 2.5 - 5.5 Product Usage: This tube is intended to provide direct airway access for a tracheostomized patient for up to 29 days. …
- Z-0660-2012 · TerminatedBivona Uncuffed Neonatal Flex Tend Plus Tracheostomy Tube Sizes 2.5 - 4.0 (With a V-Shaped Neck Flange) Product Usage: This tube is intended to provide direct airway access for…
- Z-0661-2012 · TerminatedBivona Uncuffed Neonatal Flex Tend Plus Tracheostomy Tube Sizes 2.5 - 4.0 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway access for …
- Z-0662-2012 · TerminatedBivona Uncuffed Pediatric Flex Tend Plus Standard Tracheostomy Tube Sizes 2.5 - 5.5 (With a V-Shaped Neck Flange) Product Usage: This tube is intended to provide direct airway ac…
- Z-0663-2012 · TerminatedBivona Uncuffed Pediatric Flex Tend Plus Standard Tracheostomy Tube Sizes 2.5 - 5.5 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway a…
- Z-0664-2012 · TerminatedBivona Uncuffed Pediatric Flex Tend Plus Extra Length Tracheostomy Tube Sizes 4.0 - 6.0 (With a V-Shaped Neck Flange) Product Usage: This tube is intended to provide direct airw…
- Z-0665-2012 · TerminatedBivona Uncuffed Pediatric Flex Tend Plus Extra Length Tracheostomy Tube Sizes 4.0 - 6.0 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airwa…
- Z-0666-2012 · TerminatedBivona TTS Cuffed Pediatric Tracheostomy Tube Sizes 2.5 - 5.5 (With a V-Shaped Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheostom…
- Z-0667-2012 · TerminatedBivona TTS Cuffed Pediatric Tracheostomy Tube Sizes 2.5 - 5.5 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheosto…
- Z-0668-2012 · TerminatedBivona Uncuffed Pediatric Tracheostomy Tube Sizes 2.5 - 5.5 (With a V-Shaped Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheostomi…
- Z-0669-2012 · TerminatedBivona Uncuffed Pediatric Tracheostomy Tube Sizes 2.5 - 5.5 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheostomiz…
- Z-0670-2012 · TerminatedBivona Uncuffed Neonatal Tracheostomy Tube Sizes 2.5 - 4.0 (With a V-Shaped Neck Flange)
- Z-0671-2012 · TerminatedBivona Uncuffed Neonatal Tracheostomy Tube Sizes 2.5 - 4.0 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheostomi…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page