Skip to content

Bivona 60P050Recalls & reported problems

Icu Medical, Inc. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 15021312005776), received between Nov 17, 2017 and Dec 23, 2025 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

46 other recalls cover devices cleared under the same FDA submission (K083641, K912469); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2595-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Break1
  • Device Markings/Labelling Problem1
  • Fluid/Blood Leak1
  • Fracture1
  • Inflation Problem1
  • Torn Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 23, 2025Fracture; Fracture

    IT WAS STATED THAT THERE WAS FLANGE DAMAGE. THE EVENT OCCURRED DURING PATIENT USE, AND NO ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23877923 ↗
  • MalfunctionJul 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT WHEN THE CANNULA WAS PLACED THE CANNULA WAS TORN AT THE "SHIELD". NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22564388 ↗
  • MalfunctionJul 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT WHEN THE CANNULA WAS PLACED THE CANNULA WAS TORN AT THE "SHIELD". NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22564223 ↗
  • MalfunctionOct 18, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT THE CUSTOMER STATED THE ITEM NUMBER ON THE BOX WAS DIFFERENT FROM THE LABEL INSIDE THE BOX. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15625446 ↗
  • InjuryMay 11, 2021Break; Break

    DEVICE EVALUATION COMPLETED AND SUMMARY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11808881 ↗
  • InjuryMay 20, 2020Fluid/Blood Leak; Inflation Problem; Fluid/Blood Leak; Inflation Problem

    INFORMATION WAS RECEIVED INDICATING THAT THE CUFF TO A SMITHS MEDICAL PORTEX BIVONA UNCUFFED TRACHEOSTOMY TUBE WAS FOUND DEFLATED WHILE IN THE PATIENT. THE CUFF WAS ATTEMPTED TO BE RE-INFLATED, BUT WATER WAS OBSERVED TO BE LEAKING FROM THE PILOT BALLOON. SUBSEQUENTLY, A TUBE CHANGE OUT WAS PERFORMED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10075865 ↗
  • MalfunctionNov 17, 2017Torn Material; Torn Material

    IT WAS REPORTED THAT THE TUBE WAS TORN OFF WHEN THE PATIENT WAS MOVING ABOUT WHILE USING A BIVONA® UNCUFFED PEDIATRIC TRACHEOSTOMY TUBE. THERE WERE NO PATIENT ADVERSE EVENTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7042605 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083641, K912469. One submission often covers many models, so check the product and lot details in each record.

  • Z-2609-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND STRAIGHT FLANGE , Product Code/List Number/Item Code 67NFPS25; b) NEONATA
  • Z-2596-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 4.0MM FLEXTEND PLUS, Product Code/List Number/Item Code 60PFP40; b) PAED. TRACHEOSTOMY TUBE 4.0M
  • Z-2606-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS NEONATAL TRACH TUBE , Product Code/List Number/Item Code 67N025; b) TTS NEONATAL TRACH TUBE , Product Code/List N
  • Z-2600-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: 2.5MM PEDIATRIC UNCUFFED FLEXTEND PLUS TRACHEOSTOMY TUBE WITH NEONATE STRAIGHT NECKSTRAP, Product Code/List Number/Item Code
  • Z-2597-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: 6.0MM FLEXTEND PEDIATRIC PLUS VEE TRACH TUBE , Product Code/List Number/Item Code 60PFP60
  • Z-2594-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 3.0MM FLEXTEND NEO STRAIGHT TRACH TUBE , Product Code/List Number/Item Code 60NFPS30; b) 3.5MM FLEXTEND NEO STRAIGHT T
  • Z-2612-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PEDIATRIC TRACHEOSTOMY TUBE 4.0mm TTS FLEXTEND PLUS STRAIGHT FLANGE , Product Code/List Number/Item Code 67PFPS40; b) PE
  • Z-2604-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM AIRE-CUF , Product Code/List Number/Item Code 65P025; b) PAED. TRACHEOSTOMY TUBE 3.0MM AIR
  • Z-2610-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS PEDIATRIC TRACH TUBE , Product Code/List Number/Item Code 67P025; b) TTS PEDIATRIC TRACH TUBE , Product Code/List Nu
  • Z-2588-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS TRACHEOSTOMY TUBE 7.0MM TIGHT TO SHAFT CUFF, Product Code/List Number/Item Code 670170; b) TTS TRACHEOSTOMY TUBE 8.0
  • Z-2608-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND V FLANGE , Product Code/List Number/Item Code 67NFP25; b) NEONATAL TRACHEO
  • Z-2603-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 65N025; b) NEONATAL TRACHEOSTOMY TUBE , Product Code/Lis
  • Z-2613-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PEDIATRIC TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND V NECK, Product Code/List Number/Item Code 67PFS25; b) PEDIATRIC TRACHEOS
  • Z-2599-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM STANDARD FLEXTEND PLUS , Product Code/List Number/Item Code 60PFS25; b) PAED. TRACHEOSTOMY
  • Z-2615-2024 · May 29, 2024 · Open, Classified
    PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 2.5mm TTS CUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SN025; b) 3.0mm TTS CUFFED NEONATAL T

…and 31 more.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →