Bivona 60P050Recalls & reported problems
Icu Medical, Inc. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 15021312005776), received between Nov 17, 2017 and Dec 23, 2025 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
46 other recalls cover devices cleared under the same FDA submission (K083641, K912469); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 29, 2024Open, Classified
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Z-2595-2024 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn2
- Break1
- Device Markings/Labelling Problem1
- Fluid/Blood Leak1
- Fracture1
- Inflation Problem1
- Torn Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 23, 2025Fracture; Fracture
IT WAS STATED THAT THERE WAS FLANGE DAMAGE. THE EVENT OCCURRED DURING PATIENT USE, AND NO ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23877923 ↗MalfunctionJul 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT WHEN THE CANNULA WAS PLACED THE CANNULA WAS TORN AT THE "SHIELD". NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22564388 ↗MalfunctionJul 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT WHEN THE CANNULA WAS PLACED THE CANNULA WAS TORN AT THE "SHIELD". NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22564223 ↗MalfunctionOct 18, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE CUSTOMER STATED THE ITEM NUMBER ON THE BOX WAS DIFFERENT FROM THE LABEL INSIDE THE BOX. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15625446 ↗InjuryMay 11, 2021Break; Break
DEVICE EVALUATION COMPLETED AND SUMMARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11808881 ↗InjuryMay 20, 2020Fluid/Blood Leak; Inflation Problem; Fluid/Blood Leak; Inflation Problem
INFORMATION WAS RECEIVED INDICATING THAT THE CUFF TO A SMITHS MEDICAL PORTEX BIVONA UNCUFFED TRACHEOSTOMY TUBE WAS FOUND DEFLATED WHILE IN THE PATIENT. THE CUFF WAS ATTEMPTED TO BE RE-INFLATED, BUT WATER WAS OBSERVED TO BE LEAKING FROM THE PILOT BALLOON. SUBSEQUENTLY, A TUBE CHANGE OUT WAS PERFORMED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10075865 ↗MalfunctionNov 17, 2017Torn Material; Torn Material
IT WAS REPORTED THAT THE TUBE WAS TORN OFF WHEN THE PATIENT WAS MOVING ABOUT WHILE USING A BIVONA® UNCUFFED PEDIATRIC TRACHEOSTOMY TUBE. THERE WERE NO PATIENT ADVERSE EVENTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7042605 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083641, K912469. One submission often covers many models, so check the product and lot details in each record.
- Z-2609-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND STRAIGHT FLANGE , Product Code/List Number/Item Code 67NFPS25; b) NEONATA
- Z-2596-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 4.0MM FLEXTEND PLUS, Product Code/List Number/Item Code 60PFP40; b) PAED. TRACHEOSTOMY TUBE 4.0M
- Z-2606-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS NEONATAL TRACH TUBE , Product Code/List Number/Item Code 67N025; b) TTS NEONATAL TRACH TUBE , Product Code/List N
- Z-2600-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: 2.5MM PEDIATRIC UNCUFFED FLEXTEND PLUS TRACHEOSTOMY TUBE WITH NEONATE STRAIGHT NECKSTRAP, Product Code/List Number/Item Code
- Z-2597-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: 6.0MM FLEXTEND PEDIATRIC PLUS VEE TRACH TUBE , Product Code/List Number/Item Code 60PFP60
- Z-2594-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 3.0MM FLEXTEND NEO STRAIGHT TRACH TUBE , Product Code/List Number/Item Code 60NFPS30; b) 3.5MM FLEXTEND NEO STRAIGHT T
- Z-2612-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PEDIATRIC TRACHEOSTOMY TUBE 4.0mm TTS FLEXTEND PLUS STRAIGHT FLANGE , Product Code/List Number/Item Code 67PFPS40; b) PE
- Z-2604-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM AIRE-CUF , Product Code/List Number/Item Code 65P025; b) PAED. TRACHEOSTOMY TUBE 3.0MM AIR
- Z-2610-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS PEDIATRIC TRACH TUBE , Product Code/List Number/Item Code 67P025; b) TTS PEDIATRIC TRACH TUBE , Product Code/List Nu
- Z-2588-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS TRACHEOSTOMY TUBE 7.0MM TIGHT TO SHAFT CUFF, Product Code/List Number/Item Code 670170; b) TTS TRACHEOSTOMY TUBE 8.0
- Z-2608-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND V FLANGE , Product Code/List Number/Item Code 67NFP25; b) NEONATAL TRACHEO
- Z-2603-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 65N025; b) NEONATAL TRACHEOSTOMY TUBE , Product Code/Lis
- Z-2613-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PEDIATRIC TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND V NECK, Product Code/List Number/Item Code 67PFS25; b) PEDIATRIC TRACHEOS
- Z-2599-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM STANDARD FLEXTEND PLUS , Product Code/List Number/Item Code 60PFS25; b) PAED. TRACHEOSTOMY
- Z-2615-2024 · May 29, 2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 2.5mm TTS CUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SN025; b) 3.0mm TTS CUFFED NEONATAL T
…and 31 more.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.