Dimension(R) Vista(R) V-LYTE(R) Standard A (K820) recallZ-0644-2012 · Siemens Healthcare Diagnostics, Inc. · initiated Oct 31, 2011
Terminated All recalls posted in January 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm has confirmed that specific lots of V-LYTE(R) Fluids may contribute to reduced V-LYTE(R) Integrated Multisensor on-board use life as demonstrated by increases in: Urine sodium QC results within 48 hours after installation of a new IMT sensor. The results can be elevated approximately 10-20% which may cause QC recovery being out of acceptable ranges. Patient Urine sodium results may also b
FDA root cause category: Material/Component Contamination
What the company is doing
Siemens sent an Urgent Field Safety Notice letter dated October, 2011 to all consignees. The letter identified the product, description of problem, and actions to be taken. Consignees were instructed to replace the affected lots of V-LYTE(R) Fluids with alternate lots if available and recalibrate. If no alternate lots were available they were recommended to replace the V-LYTE(R) Integrated Multisensor every 24 hours or when indicated by QC until the impacted lots were no longer in use. For replacement product consignees were instructed to contact the Siemens Customer Service at 888-588-3916 or your local Siemens Service Center. Consignees were advised to complete and return the enclosed Customer Response Sheet via fax to the Technical Solutions Center at 302-631-8467. For questions or concerns contact Siemens Technical Solutions Center at 800-441-9250 for further assistance.
Recall details
- Recall number
- Z-0644-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 31, 2011
- Posted by the FDA
- Jan 11, 2012
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 6,000
- Product code
- JJE
- FDA submissions
- K051087
- Distribution
- Worldwide Distribution - USA (nationwide) iin the states of : AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MS, MO, MT, NE, NH, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, VT, UT, VA, WA, WV, WI, WY including Puerto Rico and Washington, DC and the countries of: Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Italy, Netherlands, Norway, Portugal, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Switzerland, Canada, Australia, Japan, New Zealand, Malaysia, and South Korea.
Product description
Dimension(R) Vista(R) V-LYTE(R) Standard A (K820) Product Usage: V-LYTE(R) Standard A is for the calibration of NA, K, and CL on the Dimension Vista(R) system
Code information (models, lots, serial numbers)
Lot number KA1G01, KA1H01, and KA1J01
Other devices under the same FDA submission
4 devices cleared under K051087 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Dimension Vista® 500 10284473Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall524 reports
- Dimension Vista® 500 Intelligent Lab System 10488217Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE13 reports
- Dimension Vista® V-LYTE® Integrated Multisensor Na+ / K+ / Cl- 10484429Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, POTASSIUM10 reports
- Dimension Vista® 1500 Intelligent Lab System 10459469Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE9 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 60644).
- Z-0645-2012 · TerminatedDimension(R) Vista(R) V-LYTE(R) Standard B/Salt Bridge (K825) Product Usage: V-LYTE(R) Standard B/Salt Bridge is for the calibration and measurement of NA, K, and CL on the Dimen…
- Z-0646-2012 · TerminatedDimension(R) Vista(R) V-LYTE(R) Diluent (K835) Product Usage: V-LYTE(R) Diluent is for the automatic dilution of samples processed in the V-LYTE(R) Integrated Multisensor on the …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page