The SternaLock Blu Implant Tray includes plates, screws… recallZ-0843-2012 · Biomet Microfixation, Inc. · initiated Nov 23, 2011

Terminated All recalls posted in January 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet Microfixation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Biomet Microfixation, Inc., Jacksonville, FL recalled their SternaLock Blu Implant Tray on 11/23/11, model numbers 73-1300 through 73-1302 due to the possibility of an inaccurate reading on the Screw Measuring Area. If used incorrectly, the screw measuring area inside the SternaLock Blu Implant Tray, could lead to the improper selection of a screw 2mm longer than needed.

FDA root cause category: Device Design

What the company is doing

Biomet Microfixation, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provides a "Recommended Technique" to be followed to correct the problem until replacement trays are available. Contact the Quality Improvement Manager at 1-800-874-7711 for questions regarding this notice.

Recall details

Recall number
Z-0843-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 23, 2011
Posted by the FDA
Jan 24, 2012
Recalling firm
Biomet Microfixation, Inc.
Quantity
109 trays
Product code
HRS
FDA submissions
K110574
Distribution
Worldwide Distribution-- USA (nationwide) including the states of AL, AZ, CA, CO, DE, FL, GA, HI, IL, IN, KY, MD, MO, MS, NC, NE, NJ, NY, OK, OR, PA, SC, TN, TX, WA, and WI. and countries of Argentina, Australia, Canada, Denmark, Finland, Germany, South Africa, and Spain.

Product description

The SternaLock Blu Implant Tray includes plates, screws, screw measuring device, and blades. The SternaLock Blu Implant Tray would have been found within the following outer containers: Silver Instrument Tray, with a White rack, and there is no Power Driver warning. Labeled in part: "Manufactured for Biomet Microfixation***Biomet Microfixation 1520 Tradeport Drive Jacksonville, FL 32218 USA www.biometmicrofixation.com***#73-1301***STERNALOCK IMPLANT TRAY***256X 125.5 MM (10.080 x 4.940 ") (LXW)***Polypropylene, Aluminum, Stainless Steel***" A 3" tall container, with blue rubber handles, and a Power Driver warning on the container lid. Labeled in part: "Manufactured for Biomet Microfixation***Biomet Microfixation 1520 Tradeport Drive Jacksonville, FL 32218 USA www.biometmicrofixation.com***#73-1302***STERNALOCK BLU IMPLANT TRAY SEMIMODULAR***256X 125.5 MM (10.080 x 4.940 ") (LXW)Polypropylene, Aluminum, Stainless Steel***" A 5" tall container, with a wide metal latch to secure the lid, and there is no Power Driver warning. Labeled in part: "Manufactured for Biomet Microfixation***Biomet Microfixation 1520 Tradeport Drive Jacksonville, FL 32218 USA www.biometmicrofixation.com***#73-1300***STERNALOCK BLU IMPLANT SEMI MODULAR TRAY***228.6X 128.8MM (9.00 x 5.07 ") (LXW)Polypropylene, Aluminum, Radel***". The Biomet Microfixation SternaLock Blu System is intended for use in the stabilization and fixation of fractures of the anterior chest wall including sternal fixation following sternotomy and sternal reconstructive surgical procedures to aid in the alignment and stabilization of bone. There SternaLock instrument tray has been designed specifically to store the implants for this system.

Code information (models, lots, serial numbers)

Model Number 73-1300, Lot numbers: 032211, 042911, 100611. Model Number 73-1301, Lot numbers: 249640, 280130, 280131, 312560, 031411. Model Number 73-1302, Lot number: 011411.

Other devices under the same FDA submission

20 devices cleared under K110574 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.