STERNALOCK® BLU SYSTEM 73-2414 (UDI 00841036141158)Recalls & reported problems

Biomet Microfixation, Inc · Class 2 · Implant

All models of STERNALOCK BLU SYSTEM (31)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00841036141158), received between Apr 4, 2019 and Nov 4, 2020 — classified as 13 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

1 other recall cover devices cleared under the same FDA submission (K110574, K121302, K161896); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Fracture5
  • Migration or Expulsion of Device2
  • Device Slipped1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Material Deformation1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding4
  • No Consequences Or Impact To Patient2
  • No Patient Involvement2
  • Bone Fracture(s)1
  • Foreign Body In Patient1
  • Injury1
  • No Code Available1
  • No Information1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2020Material Deformation; Material Deformation

    IT WAS REPORTED THE STERNAL SCREW STRIPPED EASILY UPON INSERTION DURING A PRIMARY STERNOTOMY. WHEN THE SCREW WAS BEING TIGHTENED WITH A MANUAL DRIVER, THE SCREW STRIPPED EASILY AND DEFORMED THE SCREW HEAD. THE SCREW HEAD WAS REMOVED USING A CHISEL AND MALLET WHILE THE SCREW BODY WAS REMOVED USING PICK-UPS. THE SCREW WAS REPLACED WITH A DIFFERENT SCREW ONE SIZE SMALLER. THE MALFUNCTION RESULTED IN A FIVE (5) MINUTE SURGICAL DELAY. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 10785882 ↗
  • InjuryOct 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

    Patient problems: Unspecified Infection

    MAUDE report 10645020 ↗
  • InjuryAug 11, 2020Fracture; Migration or Expulsion of Device; Fracture; Migration or Expulsion of Device

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Wound Dehiscence

    MAUDE report 10391410 ↗
  • InjuryAug 3, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 10359256 ↗
  • InjuryJun 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 10212710 ↗
  • InjuryMar 18, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

    Patient problems: Foreign Body In Patient

    MAUDE report 9849000 ↗
  • InjuryJan 3, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT UNDERWENT AN EMERGENT RE-ENTRY DUE TO AN UNKNOWN REASON. STERNAL PLATES AND SCREWS WERE REMOVED AND THE PATIENT WAS CLOSED WITH NEW STERNAL PLATES AND SCREWS. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Information

    MAUDE report 9547024 ↗
  • InjuryJan 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PATIENT UNDERWENT AN EMERGENT RE-ENTRY DUE TO BLEEDING. THE PATIENT WAS CLOSED WITH NEW STERNAL PLATES AND SCREWS. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 9540128 ↗
  • InjuryDec 23, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED THAT INDICATED THE RE-OPERATION WAS PERFORMED DUE TO BLEEDING AND WAS NOT IN ANY WAY RELATED TO A DEVICE PROBLEM. THIS IS NOT A REPORTABLE EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 9513376 ↗
  • InjuryNov 27, 2019Fracture; Fracture

    THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

    Patient problems: No Code Available

    MAUDE report 9386761 ↗
  • InjurySep 5, 2019Fracture; Fracture

    THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

    Patient problems: Injury; No Patient Involvement

    MAUDE report 8968050 ↗
  • InjuryJul 18, 2019Fracture; Fracture

    THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

    Patient problems: No Patient Involvement

    MAUDE report 8805778 ↗
  • InjuryJun 21, 2019Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT BYPASS SURGERY IN 2016. THE STERNLOCK PRODUCT WAS IMPLANTED IN SPRING/SUMMER 2017 AFTER A STERNAL WOUND HEALING DISORDER. AFTER APPROXIMATELY TWO YEARS, IMAGING SHOWED TRANSVERSE FRACTURES AND DEHISCENCE AT THE MANUBRIUM/XIPHOID. UPON REMOVAL OF THE PRODUCT, THREE SCREWS WERE FRACTURED AT THE LOWER THREADS. THE LOWER PART OF THE THREADS WERE NOT REMOVED FROM THE PATIENT'S BONE.

    Patient problems: Bone Fracture(s); Organ Dehiscence

    MAUDE report 8723529 ↗
  • MalfunctionMay 10, 2019Device Slipped; Device Slipped

    IT WAS REPORTED THAT THE SCREW HEAD DID NOT RETAIN ON THE BLADE ADEQUATELY. THEREFORE, THE SCREW COULD NOT BE INSERTED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8601327 ↗
  • InjuryApr 4, 2019Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A REVISION WAS PERFORMED DUE TO THE SCREWS PULLING OUT OF THE BONE RESULTING IN STERNAL SEPARATION. THE PATIENT'S STERNUM WAS ORIGINALLY CLOSED WITH WIRE AFTER A CARDIAC SURGERY APPROXIMATELY TWO MONTHS PRIOR TO IMPLANTATION OF STERNALOCK. THE WIRES FAILED AND STERNALOCK WAS IMPLANTED AS A RESULT. THE PATIENT REPORTED THAT HE FELT PAIN AND INSTABILITY WHILE AT A GUN RANGE AND IT WAS DETERMINED VIA X-RAY THAT THE SCREWS HAD PULLED AWAY FROM THE STERNUM ON ONE SIDE. SUBSEQUENTLY, ALL STERNLOCK PLATES AND SCREWS WERE REMOVED. THE SURGEON ADVISED THAT HE DOES NOT BELIEVE THAT THIS ISSUE WAS RELATED TO THE DEVICE. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Pain

    MAUDE report 8483815 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K110574, K121302, K161896. One submission often covers many models, so check the product and lot details in each record.

  • Z-0843-2012 · Nov 23, 2011 · Terminated
    The SternaLock Blu Implant Tray includes plates, screws, screw measuring device, and blades. The SternaLock Blu Implant Tray would have been found within the following outer cont

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page