STERIS SYSTEM 1E Liquid Chemical Sterilant Processing… recallZ-0869-2012 · Steris Corporation Hopkins Facility · initiated Dec 7, 2011
Terminated All recalls posted in January 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Steris Corporation Hopkins Facility or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software deficiency. STERIS identified that water specifications were not being met at some of their customer sites. Critical parameters for initiating cycles are: water quality, water pressure and water temperature. Failure to meet these criteriae may lead to cycle aborts by the SYSTEM 1E Liquid Chemical Sterilant Processing System. The firm intends to perform a software field correction which
FDA root cause category: Software design
What the company is doing
The firm, STERIS Corp. prepared an "URGENT VOLUNTARY FIELD CORRECTION NOTICE" letter dated December 8, 2011. The firm intends to send the letter to its customers upon STERIS' receipt of 510(k) clearance from FDA. The letter includes a description of the Product-SYSTEM Liquid Chemical Sterilant Processing System, problem and actions taken. The customers were instructed to in case of an incomplete or cancelled cycle, reprocess devices in the cycle following the directions in the Operator manual. The letter also states that STERIS Corporation will install updated software on the affected system, and will be providing the customers with a new Operator Manual reflecting the revised wording resulting from the installation of the new software. For further information or if they have any questions regarding the firms visit to their facility, contact STERIS Field Service Dispatch at 1-800-333-8828. If you have questions regarding this matter, contact Director Low Temperature Sterilization, at 440-392-7455.
Recall details
- Recall number
- Z-0869-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Dec 7, 2011
- Posted by the FDA
- Jan 25, 2012
- Recalling firm
- Steris Corporation Hopkins Facility
- Quantity
- 3,882 units were distributed
- Product code
- MED
- FDA submissions
- K090036, K102462
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Bahrain, Hong Kong, Japan and South Korea.
Product description
STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Cleared 4/05/2010; K101409 cleared 8/2/2010; K102462, cleared 9/2/2010; add to file 2/24/2011. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reuseable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.
Code information (models, lots, serial numbers)
MED. Serial Numbers: 400000 through 405499 are subject to correction.
Other devices under the same FDA submission
4 devices cleared under K090036, K102462 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA Quick ConnectSteris Corporation · Sterilant, medical devices2 reports
- S40 S4000Steris Corporation · Sterilant, medical devices1 report
- SYSTEM 1E Liquid Chemical Sterilant Processing SystemCorporation Steris Canada · Sterilant, medical devices1 report
- NA Quick ConnectSteris Corporation · Sterilant, medical devices1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page