Temporary Pacing Wire, BIPOLAR, Models BM, V, VE and VF, Rx… recallZ-1038-2012 · European Custom Manufacturing B.V. · initiated Apr 18, 2011
Terminated All recalls posted in February 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from European Custom Manufacturing B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The crimping operation during product production was not performed properly. As such, the pacing wire may separate from the connection pin, which causes non-functionality.
FDA root cause category: Process design
What the company is doing
ECM sent an "ECMbv RECALL OF MCE TEMPORARY PACING WIRES" letter dated May 2, 2011 to all affected customers. Customers were also informed of a potential problem via e-mail and/or phone between April 18 and 21, 2011. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to place the affected product on hold and to return the product to the firm. For questions on this matter, contact ECM or MCE Customer Service at 1-877-886-1985.
Recall details
- Recall number
- Z-1038-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 18, 2011
- Posted by the FDA
- Feb 15, 2012
- Recalling firm
- European Custom Manufacturing B.V.
- Quantity
- 2,120 individual units (106 boxes)
- Product code
- LDF
- FDA submissions
- K033858
- Distribution
- Nationwide Distribution-including the states of AL, AR, CA, FL, GA, KY, NC, NV, NY, TX, VA, and WI.
Product description
Temporary Pacing Wire, BIPOLAR, Models BM, V, VE and VF, Rx Only, STERILE/EO, Single use only -- Mce (MEDICAL CONCEPTS EUROPE) --- Manufacturer ECM B.V. Oost-Om 54, P.O. Box 53, NL 5420 AB GEMERT --- Shipping or Unit Package: 20 individual units packaged per box with 6 boxes per case. --- Device Classification Name: Electrode, pacemaker, temporary. ---The device is supplied sterile and intended for single use. Used to pace the heart during or after open heart surgery. The electrodes are placed on the heart; the other end of the pacing wire is inserted through the skin of the patient with a thoracic needle. The thoracic needle can snap-off at the end where the pacing wire is attached, after snap-off the remaining pins are connected to an external pacemaker. For temporary atrial and ventrical pacing and sensing for contemplated implant duration of 7 days or less.
Code information (models, lots, serial numbers)
Model BM Lot numbers 1101044-10, 1012035-10, 1012036-10, 1011128-10, 1101023-10, 1101134-10, 1102035-10, 1102061-10, 1009089-10; -- Model V lot number: 1103066-10; -- Model VE lot numbers: 1101055-10, 1101008-10; -- Model VF lot numbers: 1011092-10, 1011148-20, 1012038-10, 1101021-10, 1101136-10.
Other devices under the same FDA submission
5 devices cleared under K033858 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MYO/Wire® Ultra-Thin Temporary Pacing Wires QEF2008ABW2DA & E Medical Corporation · ELECTRODE, PACEMAKER, TEMPORARY3 reports
- MYO/Wire® Ultra-Thin Temporary Pacing Wires VE60WWA & E Medical Corporation · ELECTRODE, PACEMAKER, TEMPORARY2 reports
- MYO/Wire® Ultra-Thin Temporary Pacing Wires BMZF605ACBBGA & E Medical Corporation · ELECTRODE, PACEMAKER, TEMPORARY2 reports
- MYO/Wire® Ultra-Thin Temporary Pacing Wires V605PBBA & E Medical Corporation · ELECTRODE, PACEMAKER, TEMPORARY1 report
- MYO/Wire® Ultra-Thin Temporary Pacing Wires VF608ABBA & E Medical Corporation · ELECTRODE, PACEMAKER, TEMPORARY1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page