STA - System Control N + P recallZ-1044-2012 · Diagnostica Stago, Inc. · initiated Aug 23, 2011

Terminated All recalls posted in February 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Factor VIII quality control results are being recovered below the assigned ranges for STA System Control N+P.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Stago sent a recall letters via email dated August 23, 2011 to all affected customers . The letter identified the product, description of problem, and corrective actions to be taken. Customers were asked to immediately examine their inventory and follow the corrective actions provided. Customers were instructed to complete the response form attached to the letter. Because of incorrect manufacturer's instructions included in the initial letter another letter was emailed by the firm on August 30, 2011. For questions or technical assistance for updating the parameters please call the Diagnostica Stago 24 hour hotline at 1-800-725-0607.

Recall details

Recall number
Z-1044-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 23, 2011
Posted by the FDA
Feb 16, 2012
Recalling firm
Diagnostica Stago, Inc.
Quantity
1259 units
Product code
GGN
FDA submissions
K943518
Distribution
Worldwide Distribution and USA (Nationwide) and including Canada

Product description

STA - System Control N + P Product Usage: The STA -System Control N+P kit provides a normal plasma and an abnormal plasma intended for use as two control levels for the following assays performed on analyzers of the STAbrand name suitable to these reagents: -the Reagent 1 (STA-System Control N)is for the following tests: prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen (Clauss method), thrombin time (TT), Reptilase time, factors II, V, VII, VIII, IX, X, XI and XII, antithrombin (AT), protein C, protein S, plasminogen and antiplasmin. -the Reagent 2 (STA -System Control P) is for the tests as described above except thrombin time (TT) and Reptilase time. Do not use the STA -System Control N+P kit on STA Satellite for the control of the APTT performed with the STA-rTT kit.

Code information (models, lots, serial numbers)

K9483518 D029396 Lot #106775

Other devices under the same FDA submission

1 device cleared under K943518 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.